Accuracies:
Positive Pressure: 0.05% of Full Scale (FS)
Vacuum: 0.25% of FS for ranges to 500 psi and 2% FS
for ranges > 500 psi
Accuracy includes all variations due to: non-linearity,
hysteresis, repeatability and temperature over the
compensated temperature range.
Ranges/Resolutions/Proof Pressure:
Range (Bar) Resolution Proof Pressure
-5-0-15 psi (1 Bar) .001 90 psi
-5-0-30 psi (2 Bar) .001 90 psi
-14.5-0-100 psi (7 Bar) .01 300 psi
-14.5-0-300 psi (20 Bar) .01 3000 psi
-14.5-0-500 psi (35 Bar) .01 3000 psi
-14.5-0-1,000 psi (70 Bar) .1 3000 psi
-14.5-0-3,000 psi (200 Bar) .1 10000 psi
-14.5-0-5,000 psi (350 Bar) .1 10000 psi
-14-0-10,000 psi (700 Bar) 1 15,000 psi
Environmental:
o o o o Operating range: -10 C-55 C (14 F-131 F)
o o o Compensated range: 0-50 C (32 F-122 F)
o o o o Storage range: -20 C-70 C (-4 F-158 F)
Agency Approvals: (75514 Battery Power versions only)
Built to comply with requirements for Class 1,
Division 1, Groups A, B, C & D operation
- CSA and ATEX Intrinsic Safety Approvals (pending)
- CE Mark |
Power:
3 x AA alkaline batteries. Delivers 1500
o hours service @ 3 samples/sec @ 72 F
without backlight or 2000 hours at the 1
o sample/2 secs w/o backlight @ 72 F
Display:
High contrast - LED backlight 5 ½ digit,
0.65" high; 20 segment analog bargraph
scaled to 0-100% of FS. LCD includes:
Engr. Unit indication, Min/Max icons, low
battery icon and backlight icon
Case:
Polished 304 Stainless Steel, NEMA-4/IP-
65 rated
Mounting flanges (optional):
Front or back 300 series SS welded to case
ANSI bolt pattern for 4 ½" gauge
RS-232 (optional):
Custom cable - ASCII data
Graphic Overlay:
Polycarbonate
Pressure Sensors:
316 SS - wetted parts are 316SS
Process Connection:
¼" NPT male 316SS
Dimensions (w/o flange):
4.375" x 5.0" x 1.5"
Weight (w/o flange):
1.09 lbs (495 gms) |
Standard Engineering Units :
2 o o o PSI, Bar, Kg/cm , inH O(4 C, 20 C or 60 F), 2
o o o o o ftH O(4 C, 20 C or 60 F), cmH O(4 C & 20 C), 2 2
o o mH O (4 C & 20 C), Kpa, mBar, inHg, mmHg, 2
Torr and custom Engineering Units
*Not all Engineering Units enabled for all ranges
Password Protection:
“Tamper-proof” password protection of Zero and
Calibration data is easily achieved via the keypad
LCD Bargraph:
20 segment bargraph is scaled at: 0-100% FS
Provides a convenient visual reference of pressure
readings relative to the pressure range of the gauge
Sample Rates:
User configurable 0.5/sec, 1/sec, 3/sec & 10/sec.
10/sec sampling good for QA/QC destructive tests
Power Management:
Adjustable auto power shut-off: 1-30 minutes -
shutoff feature when enabled extends battery life
Damping:
Damping can be enabled to integrate readings and
“smooth” LCD display: good for pulsation sources
Min/Max:
Capture & query Min/Max values: many
applications in process monitoring and control
Tare:
0-100% FS zero offset: when used with custom
Engr. Units, the gauge can be used in liquid level,
volume and weighing applications. |