Portable Relative Humidity Calibrator (RH CAL)
Manufacturer: EdgeTech
 
Unlike other comparative systems RH CAL is entirely self-sufficient, it does not need compressed air or water to operate which allows this system to be truly portable.

RH CAL is a microprocessor based, programmable humidity calibration system that is at home in the metrology lab or out in the field performing on site NIST traceable humidity calibrations. The system offers the highest accuracy available for both relative humidity and ambient temperature.
Portable Relative Humidity Calibrator
 

With RH CAL, temperature and relative humidity are controlled independently; therefore, the user is not limited to performing calibrations at the surrounding ambient temperature, which may not be appropriate for their calibration protocol. Using EdgeTech's Optical Chilled Mirror (OCM) primary measurement technique for traceability and feedback control, RH CAL is a standalone, portable humidity calibration system.

Contributing to RH CAL's ease of use are features such as parameter averaging, maintenance reducing Automatic Balance Cycle (ABC), and integral ambient temperature probe and D2 chilled mirror sensor which are both are located in the sample chamber providing not only superior accuracy but the fastest response.

Benefits:
EdgeTech field proven Optical Chilled Mirror (OCM) technology
Completely self-sufficient and portable humidity calibration system
Highest accuracy available for both RH and temperature (AT):

- RH Accuracy: ± 0.5%
- RH Range: 5% to 95%
- AT Accuracy: ± 0.1°C
- AT Range: 10°C to 50°C

Temperature and RH controlled independently
Automatic correction for mirror contaminant's
Certified measurements against NIST traceable standards
Standard features, where others charge extra

The RH CAL employs a unique control scheme for maintaining such precise RH control. The system incorporates a pair of finely matched volumetric proportional control valves. By independently modulating the “Dry” and “Wet” valves from full-open to full-close, and all points between, any desired humidity can be quickly generated and stabilized upon. Additionally, because the RH CAL is continuously monitoring the sample chamber conditions via its integral chilled mirror, it is able to instantaneously react to changes in humidity and maintain control stability.

Measurement:

Temperature and RH Ranges
RH: 5 to 95%
AT: 10 to 50 °C
Dew/Frost Point: –40 to 50°C, D2 Sensor

Accuracy
RH: ± 0.5%
AT: ± 0.1°C
Dew/Frost Point: ±0.1°C nominal

Dew/Frost Point and Temperature Sensors

3-wire Platinum Resistance Thermometer (PRT),
100 ohms at 0°C, nominal

Depression
60°C (113°F), nominal, D2 sensor

Slew Rate
1.0°C (1.8°F)/second max., above 0°C

Repeatability
±0.05°C

Hysteresis
Negligible


  Functionality:

Power Requirements
100 to 240 VAC, 50-60 Hz
75 Watts Maximum

Sample Flow Rate

1 liter/minute (2.0 SCFH)

Operating Temperature
Control Unit: 0 to +50°C (+32 to +122°F)

Outputs
Analog (0-5 VDC or 4-20 mA) and RS232C serial port

Optional Features:

Display
Multi-Line Graphical LED Display

Weight
14.55kg (32 pounds)

Dimensions
52W x 20.3H x 43.1D cm
(20.5W x 8H x 17D in)

Sensor Materials
Chromium, glass, epoxy, anodized aluminum

Enclosure
Plastic carrying case