Precision Industrial RTD's (Model 5627A)
Manufacturer: Hart Scientific
 

Features




- Vibration and shock resistant

- 3/4-inch bend radius for increased durability

- NIST-traceable calibration from –200°C to 500°C

Precision Industrial RTD's
 

Specifications

Resistance

Nominal 100 ohms at 0°C

Temperature Coefficient

0.00385 ohms/ohm/°C nominal

Temperature Range

–200°C to 500°C (transition and cable temperature 150°C maximum)

Resistance Stability

±0.13°C at 0°C after 1000 hours at 400°C

Sheath Material

316 Stainless Steel

Leads

Teflon™-insulated, nickel-plated stranded copper, 22 AWG

Termination

 Four gold-plated spade lugs are standard. Specify 1502 connector if needed. Other options available.

Hysteresis

0.04% maximum between –200°C and 500°C

Time Constant

Four seconds maximum for 63.2% response to step change in water moving at 3 fps.

Bending Radius

Sheath is bendable on a minimum radius of 3/4 inch except for 2-inch sensitive area of sheath near tip.

Calibration

Includes NIST-traceable table with R vs. T values in 1°C increments from –183°C to 500°C. ITS-90 coefficients included.

Probe Accuracy (includes calibration uncertainty and short-term stability)

–196°C ±0.030°C
0°C ±0.030°C
200°C ±0.031°C
420°C ±0.035°C

Size

5627A-12: 12" L x 1/4" Dia.
5627A-9: 9" L x 3/16" Dia.
5627A-6: 6" L x 3/16" Dia.

Looking for an RTD? Buying one can be a complex process. In our catalog we offer several great Burns probes to make the process of picking the right sensor easier for you.

When buying an RTD, performance isn’t the only criterion you need to look at. The issue really is price-to-accuracy and price-to-durability ratios.

The 5627A have a temperature range up to 500°C and an accuracy of ±0.03°C. They come in three different lengths and are calibrated at five points: –183°C, –38°C, 0°C, 200°C and 420°C. Each instrument is shipped with its ITS-90 coefficients and a calibration table from –200°C to 500°C in 1°C increments.

One of the best features of this sensor is that it conforms to the standard DIN 385 curve, letting you use your DIN RTD meters fully. Why use a probe that’s less accurate than your meter?

The Model 5627A are manufactured using the Burns Coil Suspension Element design for increased shock and vibration resistance. It has a mineral-insulated sheath with a minimum bend radius of 3/4-inch for flexibility and durability.

This probe is an excellent value. It has the price-to-accuracy and price-to-durability ratios you should demand in every RTD you buy!

 

 Resources
PDF File – Requires Adobe Reader 5627A Data Sheet
PDF File – Requires Adobe Reader 5627A User Guide
 Your browsing history
 Precision Industrial RTD's
 Fast Response PRT's
 The Black Stack
 Standard Thermometry Readouts
 Chub-E4 Standards Thermometer
Copyright © 2012 Alpha Controls & Instrumentation Inc..  All rights reserved.