Datum 2000™ Dual Channel Meter or Manometer (Datum 2000)
Manufacturer: Setra
 
Starting at: $439.00 (USD)
Stock Status: 3 to 4 weeks
Dual Channel Meter or Manometer - The Datum 2000™ Dual Channel Meter can be configured externally with one or two pressure transducers or transmitters. The Datum 2000™ Manometer is a fully self-contained unit with factory installed pressure transducer or transmitter.
Datum 2000™ Dual Channel Meter or Manometer
 

Features

  • Dual Channel
  • User Adjustable Hi/Lo Alarms
  • Easy Front Panel Setup
  • Pressure Display in Desired Engineering Units
  • Tabletop or Panel Mount
  • 12 VDC & 24 VDC Power Supply
  • Use with Voltage or Current Output Devices
  • High Accuracy
  • Internal Reference Voltage
  • UL Approved Power Supply
Datum 2000™ Specifications
Display
Type:
Light emitting diode
Color:
Red
Height of Digits:
0.56 inch
Maximum Readout:
+9999 to +31999
Reading Rate:
2 to 3 times per second depending on input voltage
Overload Signal (over range reading):
OVEr
Underload Signal (under range reading):
UndEr
Polarity Signal:
Negative only
Brightness:
Constant
Accuracy (Readout Only)
At 73°F (23°C):
±0.01%R ±1 digit
At 60°F to 95°F (16°C to 35°C):
±0.04%R ±1 digit
Environmental
Operating Temperature:
32°F to 130°F (0°C to 55°C)
Storage Temperature:
-40°F to 185°F (-40°C to 85°C)
Electrical Data
Power:
Domestic:
24 VDC (115 VAC 60 Hz, 12 watt adapter provided w/ -1 models)
European:
24 VDC (220 VAC 50 Hz, 14 watt adapter provided w/-2 models)
Power Cord Length:
6 feet (approx.)
Physical
Width:
3.85 inches (98mm)
Height:
Meter: 1.74 inches (44mm)
Manometer: 3.50 inches (89mm)
Depth (including connector):
7 inches: (178mm)
Cutout Dimensions
Cutout Dimensions:
Meter: 1/8 DIN: 1.77 in. x 3.62 in. (45mm x 92 mm)
Manometer: 1/4 DIN: 3.54 in. x 362 in. (90mm x 92 mm)
Font Panel:
Meter: 3.85 in x 1.99 in. (98mm x 50mm)
Manometer: 3.85 in x 3.74 in. (98mm x 95mm)
Analog Output
Transducer:
Same as transducer output voltage
Transmitters:
No analog output. The DATUM 2000™ terminates the current loop.
Adjustment Capabilities (accessible through the user set-up menus)
Voltage Inputs:
+11 to -11 volts
Current Inputs:
0 to 20 mA
Maximum Reading:
+31999
Minimum Reading:
-9999
Input Specifications
Transmitter Inputs Signal:
4-20 mA
Transmitter Inputs Impedance:
150 ohm
Transmitter Inputs Voltage Drop:
3 VDC (max.)
Transducer Inputs Signal:
+11 VDC to -11 VDC
Transducer Inputs Impedance:
45k ohm

* Specifications subject to change without notice.
If you don't see what you need here, please contact us.

 Resources
PDF File – Requires Adobe Reader Data Sheet
PDF File – Requires Adobe Reader Product Manual
 Product Details
 Part Number: View part numbers
 Accessories: View accessories