Single and multi-channel loggers accept inputs from industry-standard sensors, transmitters and transducers. Units are available for 4-20mA, 1 volt, 5 volt and 10 volt ranges and offer feature precision measurements to 5.5 microamps and 250 microvolts.
General
| Size |
2.8 x 2.1 x 0.7" (71x53x18mm);
60g (2.2 oz) |
| Operating Range |
-40 to 85°C (-40 to 185°F) and
0 to100% RH |
| Interfaces |
RS-232 serial port; ethernet; PoE |
| Mounting |
Magnetic strips; velcro optional |
| PC Software |
Compatible with the following software packages: Veriteq Spectrum, Veriteq viewLinc (for on-going monitoring & alarming), Veriteq OPC Server (to allow your Veriteq data loggers to work with other systems), Spectrum Excel Add-in, Veriteq sGo (for Palm OS), and Spectrum API (for custom applications). |
| Clock |
Accuracy: +/- 1 min./month at 0 to 50°C |
| Electromagnetic Interference |
Meets FCC Part 15 for digital devices; meets CE requirements for radiated emissions, electrostatic discharge, & radiated susceptibility. |
| Power Source |
Internal lithium battery with life of 10 years at 1 min. sampling rate. |
Current Loop & Voltage Inputs
| Input Type |
Current Loop |
Analog Voltage |
| Available Range(s) |
0 to 22mA |
1, 5, 10 Vdc |
| Precision |
5.5 microamps |
0.025% F.S. |
| Accuracy |
0.15% F.S. at 25°C |
0.15% F.S. at 25°C |
| Input Impedances |
50 ohms |
>1 MOhm |
| Isolation |
One common per logger |
One common per logger |
| Overload Protection |
60mA max. (reverse-polarity protected) |
+/-30 volts max. (reverse-polarity protected) |
Memory
| Memory Type |
Non-volatile |
| Data Sample Capacity |
70,000 12-bit samples |
| Memory Modes |
User-selectable: Wrap when memory full, or Stop when memory full. |
| Memory Protection |
Data retention > 20 years backup without power. |
| Sampling Rates |
User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day. |
| Recording Span |
Recording span depends on sample interval selected and number of channels enabled. Chart below details typical sampling rates and length of time logger will retain data in memory before wrapping around or stopping (see Memory Modes). |
|
|
|
|
|
| 1 |
2 |
3 |
4 |
5 |
| 10 seconds |
8.1
days |
4.1
days |
2.7
days |
2.0
days |
1.6
days |
| 30 seconds |
24.3
days |
12.2
days |
8.1
days |
6.1
days |
4.9
days |
| 1 minute |
1.6
months |
24.3
days |
16.2
days |
12.1
days |
9.7
days |
| 5 minutes |
8.1
months |
4.1
months |
2.7
months |
2.0
months |
1.6
months |
| 15 minutes |
2.0
years |
1.0
year |
8.1
months |
6.1
months |
4.9
months |
| 1 hour |
8.0
years |
4.0
years |
2.6
years |
2.0
years |
1.6
years |