Manufacturer: Veriteq
 

Veriteq VL-1000/1400 temperature data loggers are perfect for monitoring, alarming, and recording environmental conditions in FDA/GxP-regulated environments like Pharmaceutical Warehouses, Research and Development labs, Hospital & Clinical Environments and Ultra-low Freezers, fridges, cold rooms.


Used in conjunction with viewLinc or vLog reporting software, Veriteq VL-1000/1400 data loggers provide a simple, secure solution for temperature recording in FDA/GxP-regulated environments.



Our VL-1000/1400 temperature data loggers have on-board power and memory for autonomous recording, making your temperature data immune to any network or power failure.


Encrypted software provides tamper-proof files and reporting features that fully meet 21 CFR Part 11 requirements.


Browser-based viewLinc™ Monitoring & Alarming software allows you to monitor conditions for all VL-1000/1400 data loggers via wired or wireless ethernet while providing 24/7, multi-stage alarm notification and remote monitoring. PoE and Palm PDA connectivity are also available.


System Benefits


> Validation and Continuous Monitoring with the same logger


> Industry-leading precision and accuracy


> User-selectable recording intervals


> Multiple connectivity options - customized to your needs


> NIST-traceable, A2LA accredited calibration, done in our own laboratory, helping you meet cGMP, ISO 9000 and HACCP quality standards. Onsite calibration and rentals are also available.


> Presentation, audit-ready reports and graphs for any time period available using viewLinc or vLog. Reports from vLog can be exported to Excel.


> Two-year limited warranty


> Adjustable sample intervals with on-board recording capacity


> Secure, clip-in external inputs for secure probe connections, offering added application flexibility for difficult-to-access locations.


> Data can be viewed through a Palm PDA using Veriteq sGo.


 

Validatable Temperature Data Logger
 

 

  • Portable self-contained temperature data logger for critical pharmaceutical validation applications
  • Equipped with internal temperature sensor, memory and ten-year battery
  • Tamper-proof operation produces reliable, secure and CFR 21 Part 11 compliant documents
  • Password-protected setup, download and calibration functions ensure integrity of collected data
  • Records temperature without wires, power cords or connections; immune to power failures
  • Nist-traceable instrument-grade temperature measurement accuracy
  • Adjustable logging intervals with multi-year data recording capacity
  • High-stability temperature sensor offers exceptional long-term data logger accuracy and in-calibration performance
  • External input for remote temperature probe flexibility

 

Technical Specifications

Size 3.4 x 2.3 x 1" (85x59x26mm);
76g (2.7 oz)
Operating Range

-40 to 85°C (-40 to 185°F)
0 - 100% HR non-condensing

Interfaces RS-232 serial,USB, Ethernet,WiFi, PoE network interface available; USB interface available through USB Logger Cable. Contact Veriteq for more information.
Mounting Magnetic strips; velcro optional
PC Software Designed for use with viewLinc and vLog software 
Clock Accuracy +/- 1 min./month @ -25°C to +70°C (-13°F to +158°F)
Electromagnetic Interference Meets FCC Part 15 for digital devices; meets CE requirements for radiated emissions, electrostatic discharge, & radiated susceptibility.
Power Source Internal lithium battery with life of 10 years at 1 min. sampling rate.

Temperature Recorder Accuracy

Sensor Calibrated
Measuring Range
Resolution One year Accuracy
Internal -25°C to +70°C 0.02°C at +25°C

+/-0.15°C over +20°C to +30°C

+/-0.25°C over -25°C to +70°C

“N” Range External
Probe
-25°C to +70°C 0.02°C at +25°C +/-0.15°C over +20°C to +30°C

+/-0.25°C over -25°C to +70°C
“L” Range External Probe -50°C to +10°C 0.02°C at -20°C

+/-0.25°C over -50°C to +10°C

“V” Range External Probe -90°C to -40°C 0.02°C at -80°C +/-0.25°C over -90°C to -40°C

 

Internal Temperature Sensor

Type of Sensor Precision-tolerance epoxy-encapsulated NTC thermistor
Response Time 10 seconds in moving air (thermistor only)

Memory

Memory Type Non-volatile 32K x 8 EEROM
Data Sample Capacity 1000-2XX - 1000-LT 48,100 12-bit samples
1400-44X 85,300 12-bit samples
Memory Modes User-selectable: Wrap when memory full, or Stop when memory full.
Memory Protection Data retention > 20 years backup without power.
Sampling Rates User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day.
Recording Span Recording span depends on sample interval selected and number of channels enabled. The table below details typical sampling rates and length of time logger will retain data in memory before wrapping around or stopping (see Memory Modes, above).
1000 -2xx Sample Interval Recording Span
1 channel 2 channels
10 seconds 5.5 days 2.7 days
1 minute 1.1 months 16.7 days
5 minutes 5.5 months 2.7 months
15 minutes 1.3 years 8.3 months
1 hour 5.4 years 2.7 years
1400-44x   1 channel 2 channels 3 channels 4 channels
10 seconds 9.8 days 4.9 days 3.2 days 2.4 days
1 minute 1.9 months 29.6 days 19.7 days 14.8 days
5 minutes 9.8 months 4.9 months 3.2 months 2.4 months
15 minutes 2.4 years 1.2 years 9.8 months 7.4 months
1 hour 9.7 years 4.8 years 3.2 years 2.4 years


 Resources
PDF File – Requires Adobe Reader Data Sheet
PDF File – Requires Adobe Reader 21 CFR Part 11 White Sheet
 Your browsing history
 Validatable Temperature Data Logger
 Magnetic Level Gage for Clear, Safe Visual Level Indication
 Video Camera for Monitoring in Hazardous Locations
 Portable Transit Time Flow Meter w/Two Transducer Pairs & Data Logger
 Construction Weather Tracker
Copyright © 2012 Alpha Controls & Instrumentation Inc..  All rights reserved.