CM-3 Cut Monitor

Overview

The CM-3 Cut Monitor is a "Smart" Cut Monitor. It monitors the efficiency of oil/water separation vessels by measuring the percentage of water in oil. With its precise indication of water percentages from 0 to 50%, the CM-3 Cut Monitor can be used to indicate the effectiveness of wash tanks, freewater knockouts, heater treaters, and other separation vessels.

That information can be used on a "LACT" unit to divert "Wet" oil for additional processing/separation with ranges as low as 0 to 1%.

The CM-3 is unique in its ability to avoid false "Low-Water" readings when overloaded by conductive fluids, setting it apart from similar instruments.

Specifications - Two-Wire Transmitter

  • Power Requirement
    20 to 50 Vdc (two wire)
  • Output
    4-20 mAdc
  • Measurement Range
    0-1% water, 0-5% water, 0-10% water,
    0-30% water, 0-50% water
  • Ambient Temperature
    -30° to +140° F
  • Housing Seal Rating
    Nema 4X
  • Accuracy
    ± 0.1% water for 1% and 5% ranges.
    ± 5% of span for 10, 30, and 50% ranges
    (at standard conditions)
  • Digital Indicator (Remote electronics only)
    3+ digit liquid crystal display; 1/2" height (factory calibrated for %water.)
  • Step Response
    Less than 4 sec. to 90% of final output
  • Repeatability
    ± 0.5% span
  • Resolution
    0.2% span
  • Hazard Classification
    Sensing element: Intrinsically safe for all hazards.
  • Electronics: Intrinsically safe for Groups C & D (Div. 1 & 2) when supplied from an intrinsically safe power source, and explosion proof housing for Groups C & D (Div. 1&2) (Approvals worldwide)