Intrinsically safe loop calibrator (Model 707Ex)
Fabricant: Fluke
 
Prix: $1,103.00 (USD)
Stock Status: 2 à 3 semaines

The 707Ex is an intrinsically safe loop calibrator for use in explosion endangered areas. The Fluke 707Ex is certified in accordance with the ATEX directive (Ex II 2 G Eex ia IIC T4) in Zones 1 and 2.


Features


  • Large display and simple, quick-click push rotary button for easy one-handed operation

  • Simultaneous mA and % readout for quick, easy, interpretation of readings.

  • mA accuracy of 0.015%, superior to other loop calibrators

  • 1 µA resolution for mA source, simulate and measure

  • Push button with 25% steps for fast, easy linearity checks

  • 0-100% "span check" for fast confirmation of zero and span

  • Selectable slow ramp, fast ramp, and step ramp to provide smooth outputs for valve slewing and loop functional tests

  • Internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM

  • Measures Vdc to 28 V

  • Single 9 V battery that is easily changed

  • 0-20 mA or 4-20 mA default start up modes

  • Innovative output adjustment dial with 1 µA and 100 µA resolution

  • HART¢ç compatible resistance is connected in series with the loop supply for compatibility with HART communicators

  •  

    Intrinsically safe loop calibrator
     
    Specifications
    Current Measurement
    Range: 0 to 24 mA
    Resolution: 0.001 mA
    Accuracy: 0.015% Rdg + 2 LSD
    Current Sourcing
    Range: 0 to 20 mA or 4 to 20 mA
    Accuracy: 0.015% Rdg + 2 LSD
    Drive Capability: 700 ¥Ø at 20 mA
    Current Simulation
    Range: 0 to 20 mA or 4 to 20 mA
    Accuracy: 0.015% Rdg + 2 LSD
    Loop Power
    While measuring mA, 24 V
    Voltage measurement
    Range: 0 to 28 V DC
    Resolution: 1 mV
    Accuracy: 0.015% Rdg + 2 LSD

    Environmental Specifications
    Operating Temperature
    -10¨¬C to 55¨¬C
    Storage Temperature
    -40¨¬C to 60¨¬C
    Humidity (Without Condensation)
    0 to 95% (0¨¬C to 30¨¬C);
    0 to 75% (30¨¬C to 40¨¬C)
    0 to 45% (40¨¬C to 50¨¬C)
    Temperature Coefficient
    -10¨¬C to 18¨¬C, 28¨¬C to 55¨¬C, ¡¾.005% of range per ¨¬C

    Safety Specifications
    Safety
    CSA C22.2 No. 1010.1:1992
    EMC
    EN50082-1:1992 and EN55022:1994 Class B
    Maximum voltage
    30 V
    ATEX certification
    Ex II G Eex ia IIC T4 Zones 1 and 2, ZELM 02 ATEX 0120 X
    CE certification
    EN61010-1 and EN61326

    Mechanical & General Specifications
    Size
    152 x 70 x 41 mm
    Weight
    224 g (350 g including holster)
    Batteries
    One 9V battery Alkaline
    Warranty
    3 year
    Shock & Vibration
    per MIL-T-28800 for a Class 2 Instrument
    Display current and % of span
    Yes
    Auto step, auto ramp
    Yes
    Span Check
    Yes

     

     

     Ressources
    Fichier PDF – requiert Adobe Reader Livre d’instruction
    Fichier PDF – requiert Adobe Reader Safety Sheet