The “DewK” Thermo-Hygrometer (Model 1620A)
Fabricant: Hart Scientific
 

Features

  • Superior accuracy 
  • NEW! Network enabled 
  • Powerful logging and analysis tools 
  • Two interchangeable calibrated sensors 
  • Huge memory 
  • NEW! Upgraded software 

Specifications

Operating Range

0 °C to 50 °C (32 °F to 122 °F); 0 %RH to 100 %RH 

Calibrated Temperature Accuracy (“H” Model)

± 0.125 °C from 16 °C to 24 °C
(± 0.225 °F from 60.8 °F to 75.2 °F) 

Calibrated temperature Accuracy (“S” Model)

± 0.25 °C from 15 °C to 35 °C
(± 0.45 °F from 59 °F to 95 °F) 

Calibrated RH Accuracy (“H” Model)

± 1.5 %RH from 20 %RH to 70 %RH 

Calibrated RH Accuracy (“S” Model)

± 2 %RH from 20 %RH to 70 %RH 

Expected Extrapolated Performance (Uncertified)

± 0.5 °C (± 0.9 °F) outside calibrated range 

± 3 %RH outside calibrated range 

Delta Temperature Accuracy

± 0.025 °C for ± 1 ° changes within 15 °C to 35 °C
(± 0.045 °F for ± 1 ° changes within 59 °F to 95 °F) 

Temperature Resolution

User selectable up to 0.001 °C/°F on front-panel display (0.01° recorded) 

Delta Humidity Accuracy

± 1.0 %RH for ± 5 % changes within 20 %RH to 70 %RH 

RH Resolution

User selectable up to 0.01 % on front-panel display (0.1 % recorded) 

Inputs

Up to two sensors, measure temperature and relative humidity, detachable, cable-extendable, interchangeable, self-contained calibrations, may be assigned unique 16-character identifications 

Display

240 x 128 graphics monochrome LCD, displays password-protected temperature/humidity data graphically, numerically, and statistically (one or both channels); 16 pre-defined, user-changeable screen setups 

Memory

400,000 typical individual date/time-stamped readings 

Alarms

Password-protected visual, audible, and external alarms for temperature, temperature rate, RH, RH rate, and fault conditions 

Alarm Port Output

2.5 mm two-conductor subminiature plug, 0 V normal,
11 to 12 V active, sources up to 20 mA 

Connectivity

Ethernet, RS-232, RF (optional) 

Ethernet

RJ45 jack, 10 Base-T or 100 Base-TX; static or dynamic (DHCP client) IP address assignment 

Web Page

Embedded web page interface features: instrument identification, measurements, password-protected terminal page; can be disabled 

Wireless Option

Requires wireless modem. 802.15.4 (underlies Zigbee), 2.4 GHz frequency, 1 mW transmit power, 30 m (100 ft) typical unobstructed range; can be disabled 

Mounting

Wall mounted (hardware included) or set on a bench top 

Power

12 V dc from external 100 to 240 V dc power supply 

Battery Backup

Standard 9 V battery enables continued measuring during power disruptions 

Size (DewK) (HxWxD)

125 mm x 211 mm x 51 mm (4.9 x 8.3 x 2.0 in) 

Size (Sensors) (LxDia)

79 mm x 19 mm (3.1 x 0.75 in) 

Weight

0.7 kg (1.5 lb) 

Calibration

Certificate of NIST-traceable NVLAP accredited temperature and humidity calibration included; As Found and As Left data supplied at three temperature points and three humidity points each at 20 °C (68 °F); Complies with NCSL/ISO/IEC 17025:2000 and ANSI/NCSL Z540-1-1994 

LogWare III
(Optional Software)

Requirements include: Microsoft® Windows® 2000 (SP4) or XP (SP2) operating system, IBM compatible Intel Pentium® IV 1 GHz PC processor or better, 512 Mb RAM (1Gb or more recommended), 200 Mb HDD space for installation (additional free space recommended for data storage), CD-ROM drive for installation 

Two years ago the 1620 DewK revolutionized environmental monitoring for calibration labs, but it just got better. Wait until you hear about the 1620A! 

Now you can easily monitor and record conditions throughout your entire facility with the DewK’s new Ethernet and wireless connections, and set your upgraded LogWare III software to notify you immediately of changing conditions. 

You’ll still have the superior convenience, dependability, and NIST-traceable accuracy of a 1620, but your 1620A will give you the added accessibility and peace of mind you’ve been looking for. 

Accuracy 

Two types of sensors are available from Hart, and the DewK may be originally purchased with either one. The high-accuracy sensor (“H” model) reads temperature to ± 0.125 °C over
a calibrated range of 16 °C to 24 °C.
Relative humidity readings are to ± 1.5 %RH from 20 %RH to 70 %RH. 

The standard-accuracy sensor (“S” model) reads temperature to ± 0.25 °C over its calibrated range of 15 °C to 35 °C. Relative humidity readings are to ± 2 %RH from 20 %RH to 70 %RH. 

All DewK sensors come with NVLAP accredited certificates of calibration for both temperature and humidity, complete with data and NIST traceability. Hart provides exceptional uncertainties, including total test uncertainty ratios better than 3:1 for both temperature and relative humidity—even for the high-accuracy sensors! 

Both sensors can also measure temperature below their respective calibrated ranges to 0 °C and above their respective calibrated ranges to 50 °C with typical accuracy of ± 0.5 °C. And RH readings from 0 %RH to 20 %RH and from 70 %RH to 100 %RH are typically within ± 3 %. 

Ethernet and wireless capability 

The DewK gives you all the communications options you expect, and then some. With its built-in Ethernet RJ45 jack, multiple DewKs can be monitored from the same screen using our new LogWare III client-server software. Ethernet also gives you the possibility for remote connectivity over the internet, so you can monitor critical conditions while you’re away. 

Cables running along the floor can be a safety hazard, and cables hanging from the ceiling and walls are an eyesore. With the DewK, your wireless dreams will come true when you connect your computer through an RF modem up to 100 ft away, without the clutter of all the extra cables! 

Finally, if you need a printout, send data to a printer through the RS-232 interface in real time. 

Math and statistical functions 

In addition to temperature and humidity, the DewK calculates dew point, heat index, and rates of change for both temperature and humidity, without the need to buy additional software. Min, max, and a variety of other statistics are also calculated and can be shown on-screen.  Daily summary statistics, including min, max, and maximum rates of change are stored for the most recent sixty days. 

Calibrated sensors 

With the DewK you get two for the price of one. Having inputs for two sensors, each measuring both temperature and relative humidity, one DewK can monitor two locations at the same time. Both sensors can be run via extension cables to remote locations up to 100 feet away, or one sensor can be directly mounted to the top of the DewK. 

Each sensor is calibrated for both temperature and humidity at Fluke’s Hart Scientific Division. The calibration constants assigned to the sensors reside in a memory chip located inside the sensor housing, so sensors may be used interchangeably between different DewKs, and the recalibration of sensors doesn’t require an accompanying DewK. 

Sensors may also be assigned a unique identifier (up to 16 characters) to facilitate record keeping by matching the sensor identifier with the collected data. Each DewK ships with one sensor, with additional sensors available from Hart. Spare sensors may also be purchased as a kit, which includes a case for the sensor, a wall mounting bracket, and a 25-foot extension cable. 

Memory 

The DewK has enough on-board memory to store up to 400,000 date- and time-stamped data points. That’s two years’ worth of data for both measurements from two sensors if readings are taken every ten minutes! 

Alarms and battery backup 

Alarm settings can be set up quickly in the DewK based on temperature, the rate of change in temperature, RH, the rate of change in RH, and instrument fault conditions. Alarms can be both visual (flashing display) and audible (beeping). Likewise, alarm settings can be set up and events triggered in LogWare III. The DewK is also equipped with a 0 to 12 volt alarm output that can trigger a process control system. 

A backup battery shuts down the DewK’s display but maintains measurements for up to 16 hours in the event of a power failure. 

One very cool display 

Want to view data from across the room? Want to view data from two temperature and two humidity inputs simultaneously? Want to view data graphically, statistically, or both? At the same time?! The DewK does everything you could want—or at least everything we could think of. Up to sixteen different display setups can be stored and recalled at the touch of a single button. And all 16 can be easily modified, so you get exactly what you want.

  9932 Computer

The DewK lets you view data just about any way you like it. Both graphical data and statistical data can be shown for temperature and humidity from one or two inputs. Modifying any of the standard screens is easily done, so you see exactly what you want—no more, no less.

Confidence 

Fluke’s Hart Scientific Division supplies the world’s finest measurement laboratories with world-class temperature standards. We not only measure temperature and humidity better than anybody, we make temperature measurements functional and productive. 

Don’t compromise on your lab standards. Measure with confidence. Partner with Hart Scientific

The “DewK” Thermo-Hygrometer
 
 Ressources
Fichier PDF – requiert Adobe Reader Fiche technique
Fichier PDF – requiert Adobe Reader Livre d’instruction
Fichier PDF – requiert Adobe Reader Getting Started
 L'historique de votre navigation
 The “DewK” Thermo-Hygrometer
 Oxygen & Dew Point Analyzer
 Precision-Made Hygienic Clamps for Sanitary Connections
 Câble interface (série), Ordi à Logger, (IC-101)
 Two-Channel Pulse Data Logger
Copyright © 2012 Alpha Contrôles et Instrumentation Inc..  Tous droits réservés.