Small Diameter Industrial PRT's (Model 5618B)
Fabricant: Hart Scientific
 

Features

  • Small diameter sheath, 3.2mm (0.125 in) 
  • Excellent stability 
  • Includes ITS-90 coefficients 
  • NVLAP-accredited calibrated from –200°C to 500°C 

Specifications 

Resistance 

Nominal 100W at 0 °C 

Temperature Coefficient 

0.003923W/W/ °C nominal 

Temperature Range 

–200 °C to 500 °C
(–200 °C to 300 °C for 5618A-6-X) 

Drift Rate 

±0.1 °C when used periodically to 500 °C 

Sheath Material 

316 SST 

Leads 

22 AWG Teflon, 6’ 

Termination 

Specify 

Hysteresis 

Less than 0.01 °C at 0 °C when using –196 °C and 420 °C as the end points. 

Time Constant 

9 seconds max for 63.2 % 

Thermal EMF 

Less than 25 mV at 420 °C 

Calibration 

Includes NVLAP-accredited calibration w/ITS-90 coefficients, R vs. T values in 1 °C increments 

Size 

5618B-12: 305 mm L x 3.2 mm dia. (12 x 1/8 in)
5618B-9: 229 mm L x 3.2 mm dia. (9 x 1/8 in)
5618B-6: 152 mm L x 3.2 mm dia. (6 x 1/8 in) 

Probe Accuracy  (includes calibration uncertainty and short-term stability) 

±0.05 °C over entire range 

For secondary level performance with full ITS-90 calibration, Hart’s 5618B series PRTs are an excellent choice for critical temperature measurements. Featuring a 3.2 mm diameter (1/8 in) sheath, these industrial standards probes have reduced response time without compromising precision. This small diameter 5618A probe works well in many applications where immersion depth is limited. Larger diameter probes give more measurement error in short immersion depth applications because they conduct more heat between ambient and the sensor. 

With each probe you will receive a full NVLAP-accredited calibration report. On the report you’ll get the test data and the ITS-90 calibration coefficients that you can easily input into your Hart thermometer. If you are using a 1521 Handheld Thermometer readout, we’ll program the coefficients directly into your INFO-CON connector. 

The 5618B is also a great probe to use for calibrating your Hart 9132 or 9133 infrared calibrators. In fact, these IR black body heat sources were designed to be calibrated with this type of probe. Now you can calibrate these targets in your own lab! 

For use from –200 °C to 500 °C (the six-inch model goes to 300 °C), you won’t find a better industrial standard in this configuration than our 5618B. We recommend using the 5618B PRTs with the 1521, 1522, 1502A, 1529, or 1560 thermometer readouts. 

 

Small Diameter Industrial PRT's
 
 Ressources
Fichier PDF – requiert Adobe Reader 5618B Data Sheet
Fichier PDF – requiert Adobe Reader 5618 User's Guide
 L'historique de votre navigation
 Small Diameter Industrial PRT's
 Seconday Standard PRT's
 Secondary PRT with Calibration Options
 Precision Industrial RTD's
 Fast Response PRT's
Copyright © 2012 Alpha Contrôles et Instrumentation Inc..  Tous droits réservés.