Product Description
The AVS150 is a complete system used to perform autoclave validations. The AVS150 consists of a N.I.S.T. traceable HiTemp150 data logger, an IP68 extended range thermal shield, an IFC200 USB cable package and MadgeTechs FDA 21 CFR Part 11 Secure Software. For larger size autoclaves or mapping projects, the AVS150-5, includes five HiTemp150s and five thermal shields.
The HiTemp150 is completely submersible and can measure temperatures up to 150°C. The 2" rigid probe provides a fast response time and can record readings as fast as once per second. The HiTemp150 can be used in each autoclave cycle for verification and to provide data for validation. When used with the thermal shield the device can withstand up to 250°C (time dependent).
MadgeTech's Secure Software provides data in graphical and tabular form, and provides a summary and statistical report of the data. The software also features sterilization calculations to determine the F0 value. Furthermore the software allows for different user levels and logins, electronic signatures, an audit trail and a secure data file to stand up to auditors.
Please note, the AVS150 is not compatible with 64 bit operating systems.
Specifications
Temperature
| Temperature Sensor |
100 ohms Platinum RTD |
| Temperature Range |
-40 to +150°C |
| Temperature Resolution |
0.05°C |
| Temperature Calibrated Accuracy |
±0.5°C |
General
| Memory |
32,767 |
| Reading Rate |
1 second up to 12 hours |
| Baud Rate |
2,400 |
| Battery Life |
1 Year |
| Dimensions |
Body: 1.1" x 1.75" dia. (28mm x 45mm)
Probe: 2.0" x 3/16" dia. (50mm x 5mm) |
| Submersible |
Yes |
| Operating Environment |
-40~+150°C, 0~100%RH |
| Approvals |
CE |