HygroGen HG2 (HG2)
Fabricant: Rotronic
 

The HygroGen HG2 humidity and temperature generator is designed as a calibrator for relative humidity probes. The HG2 uses a mixed flow method for generating the humidity (%RH) required by the user. A desiccant cartridge provides low humidity, a saturator high humidity, and a control system mixes these together to create a constant environment within a well insulated chamber. Temperature is controlled with a Peltier thermoelectric heat pump. The HG2 typically requires less than 5 minutes to equilibrate from one humidity condition to another and can accommodate up to five probes simultaneously.


The HG2 is our second generation of humidity and temperature generators and uses an embedded PC for the accurate control of both humidity and temperature conditions.


The HG2 features a user friendly touch screen graphical interface as well as a programmable ramp and soak function with multiple steps and profiles. The HG2 is pre-loaded with the ROTRONIC HW4-P data acquisition software which is used for calibrating and adjusting the HygroClip 2 probes and other AirChip 3000 humidity measurement devices.


The HG2 is a self-contained unit requiring no external resources except mains power. In addition, it can also be used together with a chilled mirror reference instrument.


 

HygroGen HG2
 

Key Features

Operation

 
  • Generates conditions within the range of 0 to 60°C and 5 to 95 %RH
  • Extremely stable humidity and temperature environment
  • Fast equilibrium: typically less than 5 min. from one condition to another
  • Accepts up to five humidity-temperature probes simultaneously
  •  

     

    Design

    • Mixed flow with desiccant dryer cell and piezoelectric humidifier
    • Peltier thermoelectric element with radial chamber mixing fan
    • Smart chamber design minimizes temperature and humidity gradients
    • Internal RTD for temperature control and reference
    • Function to correct RTD to match HygroClip 2 probe temperature data
    • Electronic monitoring of the water level and desiccant condition with on-screen indication
    • Integrated UV water sterilization system
    • Two standard door models: 15mm or 30mm adjustable ports to suit almost any probe diameter
    • Heated internal sample loop for connection with a chilled mirror reference instrument

    Embedded PC

    • Embedded PC for configuration, control and automated functions
    • Fully functional ROTRONIC HW4-P software with data acquisition, calibration and adjustment functions for HygroClip probes
    • Solid State Disk (SSD)
    • Integrated nine port USB hub for connecting Rotronic USB probe adapters and peripherals such as mouse and keyboard
    • RJ-45 Ethernet Port
    • Intuitive, easy-to-use touch screen graphical user interface
    • DVI - Monitor out for external display
    • User upgradable firmware

    Automated Calibration

    • User programmable ramp and soak function with up to 200 steps per profile
    • Up to 20 different profiles
    • Integrated ROTRONIC HW4 software for the calibration and adjustment of HygroClip probes

     

     

     

     Ressources
    Fichier PDF – requiert Adobe Reader Fiche technique
    Fichier PDF – requiert Adobe Reader Product Manual
     L'historique de votre navigation
     HygroGen HG2
     HygroGen
     Hygrostat
     EAxx-SCS Humidity Standard (Box of 5)
     Portable System in a Carrying Case
    Copyright © 2012 Alpha Contrôles et Instrumentation Inc..  Tous droits réservés.