The HygroLog is compact, accurate and calibration free
Measuring range: 0..100%RH and -50..200°C (-58..392°F) with remote probe
Up to 416 days of data recording
Communication with the serial port of PC via data cable (after disconnecting probe)
Full featured, windows based ROTRONIC HW3 software (not compatible with VISTA)
Interchangeable humidity-temperature probe for fast and easy maintenance
Continuous (Loop) Recording
Data recording begins as soon as the HygroClip probe is plugged in. When the memory is full, the oldest data is erased as new data is added.
Start / Stop Recording
Data recording begins at a specific time and ends when the memory is full.
Alarm Values
Alarm values can be programmed and shown on a chart.
The display of the HygroLog-D provides an indication when any alarm value has been exceeded.
HW3 is a software package designed for the HygroLog data logger to do the following:
Program / Read the HygroLog Data Logger
• Recording mode (start/stop or loop)
• Adjust date and time
• Set log interval
• Set alarms, etc.
Log Files
• Presentation of the data in tabular form
• Possibility of converting the relative humidity data
into another humidity parameter such as dew point, etc.
• Compatible with Microsoft's Excel
Graphs
• graph can be customized
• can display calculated values: dew-point, mixing, ratio etc.
after converting the log file
• zoom function
• cursor function (display of numerical values on screen)
• visualization of alarms, manual notes (events, etc.)
Probe Calibration
• Humidity: up to 4 points, Temperature: 2 points
• 1-point adjustment function (offset)
• Automatic temperature correction of
the ROTRONIC humidity standards
• certificate can be generated
System Requirements
PC
Pentium 233 MHz or better
Windows
95, 98, Me, NT, 2000, XP (not compatible with Vista)
Memory
32 MB or better
Disk Space
20 MB
Monitor
VGA or Super VGA
Resolution
1024 x 768 recommended (800 x 600 minimum)
High Color 16 Bit or better (256 color minimum)