The HygroLog HL-NT humidity, dew point and temperature data logger can be used with any of the HygroClip 2 probes. The HL-NT can monitor up to 7 probes as well as up to two external on/off contacts, such a a door contact.
When used together with the validated ROTRONIC HW4 software the HygroLog HL-NT data logger complies with FDA 21 CFR part 11 regarding electronic data records and electronic signatures (ERES).
The HW4 software can display the data from several data loggers on multiple PCs and can centrally record on one of the PCs the measurement data from several loggers. As a protection against power outages and network failures, HW4 can be used to automatically download the locally recorded data at regular intervals of time.
Relative humidity, temperature, dew point or other calculated parameter
Measuring range: 0 to 100%RH and -100..200°C /-148..392°F (with external probe)
Interchangeable humidity-temperature HygroClip 2 probe for fast and easy maintenance
Wide selection of probes to satisfy every application
Optional internal probe, protected against unauthorized removal
Large recording capacity (depends on the size of removable flash memory card)
Power supply: A/C adapter, backed up by a standard or rechargeable 9 VDC battery
Full featured, Windows based ROTRONIC HW4 networking software
HW4 software has optional OPC tags to export data to any third-party OPC client application
Which HL-NT model is right for you?
HL-NT2 Data Logger Key Features:
Measures and records humidity, temperature and calculated parameter (see list below)
Internal humidity-temperature probe designed to prevent unauthorized removal
Measuring range: 0..100%RH and -100..200°C
(-148..392°F) when probe is separated from logger with a cable (depends on probe model)
Large recording capacity with removable flash memory card
Full probe adjustment functions directly from keypad
Interchangeable humidity-temperature HygroClip 2 probe for fast and easy maintenance
Complies with FDA 21 CFR Part II when used together with the ROTRONIC HW4 software
The optional docking station adds the following features (depends on docking station model):
Up to 4 additional external probes: HygroClip 2 probes, third party analog probes or 4-wire RTD
Monitoring of up to two external contacts (door, relay contact, etc.)
Power provided by an AC adapter
Data download without interruption of the measurements
Networking (RS485 or Ethernet)
HL-NT3 Key Features
All the features of the HL-NT2 + 2 additional probe inputs (HygroClip 2 probes)
The optional docking station adds the following features (depends on docking station model):
The optional docking station adds the following features (depends on docking station model)
Up to 4 additional external probes: HygroClip probes, third party analog probes or 4-wire RTD
Monitoring of up to two external contacts (door, relay contact, etc.)
Power provided by an AC adapter
Data download without interruption of the measurements
Networking (RS485 or Ethernet)
Model Numbers:
HL-NT2-D: data logger only (without internal HC2-S probe)
HL-NT2-DP: includes internal HC2-S probe
HL-NT3-D: data logger only (without internal HC2-S probe)
HL-NT3-DP: includes internal HC2-S probe
Calculated Parameter:
Each probe input of the HL-NT or docking station can be individually configured to calculate and record one of the following parameters:
Dew point (Dp)
Frost point (Fp)
Wet bulb temperature (Tw)
Enthalpy (H)
Vapor concentration (Dv)
Specific humidity (Q)
Mixing ratio by weight (R)
Vapor concentration at saturation (Dvs)
Vapor partial pressure (E)
Vapor saturation pressure (Ew)
Docking Stations
The HL-NT data logger requires a docking station for any of the following:
Mount the HL-NT data logger to a wall
(logger can be secured to docking station with a padlock)
Place the HL-NT on the optional desk-top stand
Use an external AC adapter to power the HL-NT
Connect the HL-NT to a PC via a serial port (RS232), a USB port, or Ethernet - wired or wireless
Connect up to 4 additional probes to the HL-NT
Monitor up to two external contacts (relay, door, etc.)
Connect up to 64 HL-NT data loggers together in a multi-dropped arrangement (RS-485). Unlimited number of such networks
Control up to 2 relay outputs
Access the HL-NT from a web browser
HL-NT data logger with docking station
Models
Primary
Interface
Description / Function
HL-DS-NT1
None
This model simply allows mounting the HL-NT data logger to a wall and powering it with the AC adapter
HL-DS-NT2
RS-232
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher).
HL-DS-NT3
USB
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher)
HL-DS-NT4
Ethernet
(RJ45 wired)
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher)
HL-DS-NT4-WEB
Ethernet
(RJ45 wired)
Internal web server can be accessed from a web browser without requiring the HW4 software.
HL-DS-NT4-WL
Ethernet
(wireless)
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher)
HL-DS-U-1
RS-232
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher)
Connects up to four HygroClip 2 probes or single channel analog probe with voltage output, and up to two digital input signals (external contacts).
HL-DS-U-2
USB
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher)
Connects up to four HygroClip 2 probes or single channel analog probe with voltage output, and up to two digital input signals (external contacts).
HL-DS-U-4
Ethernet
(RJ45 wired)
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher)
Connects up to four HygroClip 2 probes or single channel analog probe with voltage output, and up to two digital input signals (external contacts)..
HL-DS-U4-WEB
Ethernet
(RJ45 wired)
Internal web server can be accessed from a web browser without requiring the HW4 software.
Connects up to four HygroClip 2 probes or single channel analog probe with voltage output, and up to two digital input signals (external contacts).
HL-DS-U4-420
Ethernet
(RJ45 wired)
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher)
Connects up to four single channel analog probes with 4...20 mA signal, and up to two digital input signals (external contacts).
HL-DS-U4WEB-420
Ethernet
(RJ45 wired)
Internal web server can be accessed from a web browser without requiring the HW4 software.
Connects up to four single channel analog probes with 4...20 mA signal, and up to two digital input signals (external contacts).
HL-DS-PT-2
USB
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher).
Connects up to four RTD probes (direct 4-wire connection), and up to two digital input signals (external contacts).
HL-DS-PT-4
Ethernet
(RJ45 wired)
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher).
Connects up to two RTD probes (direct 4-wire connection), and up to two digital input signals (external contacts).
HL-DS-PT-4-WL
Ethernet
(wireless)
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher).
Connects up to two RTD probes (direct 4-wire connection), and up to two digital input signals (external contacts).
HL-DS-R1
USB
RS-485 interface allows connecting up to 63 additional ROTRONIC devices in a multi-drop arrangement (requires HW4 professional or higher).
Two independent relay contacts and connections for up to two digital input signals (external contacts)..
Notes:
All above models feature a receptacle for the AC1211 power adapter (not included).
Use of the AC adapter is required for all models with an Ethernet interface and is strongly recommended whenever several probes are connected to the data logger and docking station.
Docking station with Ethernet interface: separate the HygroClip 2 probe by a length of cable to prevent heat transfer.
Probe Options
HC2-S: measurement in ambient air with an indicator, transmitter or data logger
Humidity measuring range: 0…100 %RH
Temperature limits
-40…100 °C (-40…212 °F)
Diameter: 15mm (0.6”),
length: 85 mm (3.3”)
Housing and filter cap: polycarbonate, black
Standard dust filter:
NSP-PCB-PE (Polyethylene)
Weight: approx. 10 g (0.35 oz)
HC2-P05: hand-held probe for measuring in air ducts
Humidity measuring range: 0…100 %RH
Temperature limits
-40…85 °C (-40…185 °F)
Probe tip: diameter: 5 mm (0.2”),
length: 200 mm (7.9”)
Probe cable length: 2 m (6.5 ft)
Handgrip: POM, black - barrel: stainless steel
Weight: approx. 160 g (5.6 oz)
HC2-C04: measurement in tight spaces
Humidity measuring range: 0…100 %RH
Temperature limits:
-40…100 °C (-40…212 °F)
Probe tip diameter probe: 4 mm (0.16”)
Cable length (electronics – probe tip:
about 2 m (6.5 ft)
Electronics barrel: polycarbonate, black
Weight: approx. 150 g (5.3 oz)
HC2-C05: measurement in tight spaces
Same as HC2-C05 but with 5 mm diameter (0.2”)
Can be equipped with Teflon dust filter SP-T05
Weight: approx. 160 g (5.6 oz)
HC2-HK25 / HC2-HK40: hand-held air probe with wide temperature range
Humidity measuring range: 0..100 %RH
Temperature limits at sensor:
HK25: -100…150 °C (-148…302 °F)
HK40: -100…200 °C (-148…392 °F)
Barrel:
HK25: dia.: 15 mm (0.6”)
length: 250 mm (9.8”)
HK40: dia.: 15 mm (0.6”)
length: 400 mm (15.7”)
Probe cable length: 2 m (6.5 ft)
Handgrip: POM, black -, barrel: PPS, black
Filter base (NSP-ME) with wire mesh filter cartridge (SP-M15)
Weight:
HK25: approx. 210 g (7.4 oz)
HK40: approx. 240 g (8.5 oz)
HC2-HP28 / HC2-HP50: hand-held probe for measuring materials in bulk
Humidity measuring range: 0..100 %RH
Temperature limits at sensor:
-40…85 °C (-40…185 °F)
Barrel:
HP28: dia.: 10 mm (0.4”)
length: 280 mm (11.0”)
HP50: dia.: 10 mm (0.4”)
length: 500 mm (19.7”)
Probe cable length: 2 m (6.5 ft)
Handgrip: POM, black - barrel: stainless steel
Filter base (NSP-ME) and sintered steel filter cartridge (SP-S15)
Weight:
HP28: approx. 200 g (7.1 oz)
HP50: approx. 250 g (8.8 oz)
HC2-HS28 / HC2-HS42: hand-held probe for measuring paper stacks and rolls
Humidity measuring range: 0…100 %RH
Temperature limits at sensor:
-40…85 °C (-40…185 °F)
Blade
HS28: thick.: 4 mm (0.16”)
width: 22 mm (0.87”),
length: 280 mm (11.0”)
HS42: thick.: 4 mm (0.16”)
width: 22 mm (0.87”),
length: 420 mm (16.5”)
Probe cable length: 2 m (6.5 ft)
Handgrip: POM, black - blade: aluminum
Weight
HS28:: approx. 240 g (8.5 oz)
HS42: approx. 300 g (10.6 oz)
HC2-IC1 / HC2-IC3: fixed through wall installation in industrial processes
Humidity measuring range: 0…100 %RH
Temperature limits at sensor:
-100…200 °C (-148…392 °F)
Barrel - PPS black :
IC1: diameter: 15 mm (0.6”)
length: 100 mm (3.9”)
See note below
IC3: diameter: 15 mm (0.6”)
length: 250 mm (9.8”)
Probe cable length:
IC102 / IC302: 2 m (6.5 ft)
IC105 / IC305: 5 m (16.4 ft)
Filter base (NSP-ME) with filter cartridge
(specify when ordering)
Weight with 2 m cable:
IC1: approx. 230 g (8.1 oz)
+ 80 g (2.8 oz) per additional meter
IC3: approx. 260 g (9.2 oz)
+ 80 g (2.8 oz) per additional meter
Note: HC2-IC1 (100 mm barrel length) – to avoid measurement errors due to temperature gradients fully immerse the probe barrel in the environment to be measured
HC2-IC3-A / HC2-IC4-A / HC2-IC5-A / HC2-IC7-A: fixed through wall installation in industrial processes
Humidity measuring range: 0…100 %RH
Temperature limits at sensor:
-100…200 °C (-148…392 °F)
Barrel - PPS, black :
IC3-A: diameter: 15 / 25 mm (0.6 / 1.0”)
length: 250 mm (9.8”)
IC4-A: diameter: 15 / 25 mm (0.6 / 1.0”)
length: 400 mm (15.7”)
IC5-A: diameter: 15 / 25 mm (0.6 / 1.0”)
length: 550 mm (21.6”)
IC7-A: diameter: 15 / 25 mm (0.6 / 1.0”)
length: 700 mm (27.5”)
Filter base (NSP-ME) with filter cartridge (specify when ordering)
Weight with 2 m cable:
IC3-A: approx. 290 g (10.2 oz)
+ 80 g (2.8 oz) per additional meter
IC4-A: approx. 320 g (11.3 oz)
+ 80 g (2.8 oz) per additional meter
IC5-A: approx. 350 g (12.3 oz)
+ 80 g (2.8 oz) per additional meter
IC7-A: approx. 380 g (13.4 oz)
+ 80 g (2.8 oz) per additional meter
HC2-IM1 / HC2-IM3 / HC2-IM4 / HC2-IM5: fixed through wall installation in industrial processes
Humidity measuring range: 0…100 %RH
Temperature limits at sensor:
-100…200 °C (-148…392 °F)
Barrel - Stainless steel:
IM1: diameter: 15 mm (0.6”)
length: 120 mm (4.7”). See note below
IM3: diameter: 15 mm (0.6")
length: 280 mm (11.0”)
IM4: diameter: 15 mm (0.6”)
length: 430 mm (16.9”)
IM5: diameter: 15 mm (0.6”)
length: 580 mm (22.8”)
Filter base (SP-MSB15)with filter cartridge (specify when ordering)
Weight with 2 m cable:
IM1: approx. 260 g (9.2 oz)
+ 80 g (2.8 oz) per additional meter
IM3: approx. 400 g (14.1 oz)
+ 80 g (2.8 oz) per additional meter
IM4: approx. 540 g (19.0 oz)
+ 80 g (2.8 oz) per additional meter
IM5: approx. 680 g (24.0 oz)
+ 80 g (2.8 oz) per additional meter
Note: HC2-IM1 (120 mm barrel length) – to avoid measurement errors due to temperature gradients fully immerse the probe barrel in the environment to be measured
HC2-IE3: fixed installation in pressurized conduits and vessels
Humidity measuring range: 0…100 %RH
Temperature limits at sensor:
-100…200 °C (-148…392 °F)
Barrel - Stainless steel
NPT ½” thread for maximum 400 bar / 5800 PSI)
Probe cable length:
IE302: 2 m (6.5 ft)
IE305: 5 m (16.4 ft)
Filter base (SP-MSB15)with filter cartridge (specify when ordering)
Weight with 2 m cable: approx. 290 g (10.2 oz) + 80 g (2.8 oz) per additional meter
Note: To avoid measurement errors, temperature should be the same on both sides of the mounting wall
HW4 Software
The ROTRONIC HW4 software (version 2.3.0 or higher) is available in the following versions:
HW4 Standard Edition: for single users
Unlimited number of instruments (depends on capabilities of PC and available ports)
On-line display of the measured and calculated values (dew point or other) - limited to a single instrument at a time (instrument may have several probes)
Automatic retention of the most recent data in a temporary on-line buffer
Logging of measured and calculated values to the PC (can be done simultaneously with local data logging)
Easy-to-read graphs and data tables
Statistical data analysis: mean, standard deviation, minimum and maximum (off-line data only)
Data print-out (table or graph)
Automatic device / instrument recognition and identification
Device/instrument configuration
Adjustment / calibration of the HygroClip and HygroClip 2 digital probes
Psychrometric Conversion Tool
Built-in security to protect against manipulation of recorded data
HW4 Professional Edition: meets the requirements of FDA 21CFR Part II / GAMP and allows multiple users, with either administrator or standard rights. On each single port (RS-232, USB or TCP/IP), HW4 Professional can monitor up to 64 instruments connected together in a multi-dropped RS-485 network. The number of such networks depends on the number of available ports.
All the functionality of HW4 Standard Edition
Password protected log-in
User Event Logging: automatically keeps track of user activity
On-line display in a single graph of the data from multiple instruments and probes
Overlaying of multiple log files into a single graph
Self Event Logging: automatically records eventual software problems to facilitate troubleshooting
Logger Event Logging: automatically records the data logger internal events and configuration changes
Automatic creation of protocols detailing instrument configuration changes, log function programming and probe adjustment
Extensive alarm functions and alarm reporting. Indication on the PC of an alarm condition, printing of a report or sending an e-mail (MAPI or SMTP)
Automatic downloading and archiving on the PC of the files recorded locally by a data logger
Automatic synchronization of the networked devices with the PC time
Device protection against unauthorized configuration / programming changes
HW4 Professional with OPC Server: this version includes all features of HW4 Professional while also providing OPC tags that can be enabled for each individual device communicating with HW4. This allows exporting and integrating the HW4 data with with practically any OPC client (requires configuration / programming of the OPC client by the user).
HW4 Validated: this version offers the same functionality as HW4 Professional with OPC server. Additionally, the “HW4 e-compliance Package” is available. This extensive collection of documents (including template of qualification documents) is designed to support the “regulated user” by qualifying/validating HW4-based solutions. The most current validated release of HW4 is v 2.3.0.
HW4 Professional Trial Version: this is a fully functional version of HW4 Professional, including the OPC server functionality. A compressed installation file (zip) can be downloaded free of charge from the ROTRONIC website (click on "DOWNLOADS" in this page header)
Activation requires a product key which can be requested by filling a form on the ROTRONIC web site. A trial product key will be sent to your e-mail address. This must be entered in the registration form which appears when starting HW4 for the first time. After 30 days, the trial product key expires and HW4 can no longer be started.
HW4 Requirements (version 2.3.0 or higher):
Computer/Operating System Requirements:
The following are the minimum values required to install and run HW4 on a computer. It is highly recommended to exceed these values.
Processor: Pentium II, 450 MHz
RAM: 128 MB
Available hard disk space: 50 MB
Monitor: SVGA, 1024 x 768, 256 colors
Ports: one free serial (COM) port or one free USB port or Network Interface Card (Ethernet)
Operating System Compatibility:
Windows XP, NT4 with SP 6a or higher, Vista
Windows 2000 with SP 2 or higher
Windows Server 2003
HW4 v 2.3.0 or higher was written for the Microsoft .NET framework v 2 and requires this framework to be installed on the computer. The .NET framework offers significant improvements in the areas of networking and user security. When new software is being installed, the .NET framework also eliminates the potential problem of conflicting dynamic library files (DLL).