| Technical data |
MP100H (voltage outputs)
|
MP400H (current outputs)
|
| Configurations |
- Separate temperature probe with active linear output signal (standard)
- Separate temperature probe with direct 2-, 3- or 4-wire RTD connection (optional)
- No separate temperature probe - temperature output provided by RTD from HygroClip (optional)
|
| Humidity Temperature Plug-in Probe |
HygroClip S3 (included with MPH) |
| Humidity sensor: ROTRONIC Hygromer® C94 (capacitive) |
Temperature sensor: RTD Pt100 Ohm, 1/3 DIN
|
Sensor protection: Wire mesh filter (standard)
Foam filter (optional) |
| Separate Temperature Probe |
Stainless steel rod 118 x 3 mm (4.64 x 0.12") |
Pt100 1/3 DIN RTD (standard)
Pt100 1/5 DIN RTD (optional)
Pt100 1/10 DIN RTD (optional) |
| Operating/Measuring range |
0...100%RH
-40...60°C / -40...140°F |
| Accuracy at 23°C / 73°F |
<±1.5%RH and ±0.2°C (valid for active outputs) |
| Repeatability |
better than 0.5%RH and 0.1°C |
Measuring ranges
(active outputs) |
0...1V=0...100%RH |
4...20mA=0...100%RH |
0...1V=-40...60°C (standard)
0...1V=-30...70°C (optional)
-0.4...0.6V=-40...60°C (optional) |
4...20mA=-40...60°C (standard)
4...20mA=-30...70°C (optional) |
| Active Outputs Load |
> 1000 Ohm |
< 500 Ohm |
| Supply Voltage |
3.6...35VDC, 10 mA |
5VDC + 0.02 x load, 45 mA max. |
| Cable Length Compensation |
effective for up to 100m/300ft |
N/A (current outputs) |
| Connecting Cable |
standard: 3m (9ft) PUR cable with tinned ends |
| Weight |
210 g / 0.47 lb |