Versatile single and multi-channel thermocouple loggers accept inputs from type J, K, T, E, R and S thermocouple temperature probes. Wide measurement range from -240° C to 1760° C.
General
| Size |
2.8 x 2.1 x 0.7" (71x53x18mm);
60g (2.2 oz.) |
| Operating Range |
-40 to 85° C. (-40 to 185° F.) and
0 to 95% RH |
| Interfaces |
RS-232 serial port; Ethernet; PoE |
| Mounting |
Magnetic strips; Velcro optional |
| PC Software |
Compatible with the following software packages: Veriteq Spectrum, Veriteq viewLinc (for on-going monitoring & alarming), Veriteq sGo (for Palm OS), Spectrum Excel Add-in, and Spectrum API (for custom applications). |
| Clock |
Accuracy: +/- 1 min./month at 0 to 50° C. |
| Electromagnetic Interference |
Meets FCC Part 15 for digital devices; meets CE requirements for radiated emissions, electrostatic discharge, & radiated susceptibility. |
| Power Source |
Internal lithium battery with life of 10 years at 1 min. sampling rate.
|
System Operation
Series 1700 Thermocouple dataloggers record low-level millivolt thermocouple signals. The system automatically processes these signals and references them to the logger's on-board cold junction temperature sensor to produce accurate temperature readings.Two types of Series 1700 dataloggers are available: The narrow range "N" series provides high precision and exceptional low-temperature performance; the wide range "W" series provides a broader temperature measurement range ideal for high temperature applications. The following two charts detail the input specifications , measurement ranges, and rated accuracies of these two series of units.
Datalogger Inputs
|
|
|
|
|
|
|
|
| Narrow Range |
1700-21N |
2 |
1 |
1 |
| 1700-51N |
5 |
1 |
4 |
| Wide Range |
1700-21W |
2 |
1 |
1 |
| 1700-51W |
5 |
1 |
4 |
Note: One channel on each model provided for Cold Junction Temperature (CJT) reference using on-board precision-tolerance thermistor.