A self-contained system capable of producing atmospheres of known humidity using the fundamental “two-pressure” principle
The 1220 Humidity Generation System is a self-contained system capable of producing atmospheres of known humidity using the fundamental “two-pressure” principle. This system is capable of continuously supplying relative humidity, dew point, frost point, parts per million, or other calculated values for instrument calibration and evaluation as well as for precision environmental testing. This system will automatically generate manually entered humidity and temperature set points as well as user created multipoint profiles. All desired humidity’s, temperatures, flow rates, and time intervals may be programmed. The 1220 humidity generating system incorporates a test chamber, with internal dimensions of 8” x 8” x 8”. Test chamber pressure range is ambient. Access is available through one 1.875” diameter port located on the right side, for probes, cables, sample tubes. The test chamber accommodates various solid-state sensors, sensing systems, chilled-mirror hygrometers, as well as material samples for environmental testing. Virtually any humidity and temperature may be generated, for long periods of time, within the operational limits of the generator. The output or recording of the device under test may then be compared with the generator’s data for analysis.
Relative Humidity Range: | 10 to 95 %RH, 1 to 20 L/min |
Relative Humidity Range: | 95 to 98 %RH, 1 to 10 L/min |
Frost Point Temperature Range: | –22.0 to 0 °C |
Dew Point Temperature Range: | –24.5 to 59 °C |
Chamber Fluid Temperature Range: 2 | 5 to 60 °C |
Chamber Fluid Temperature Control Stability: 3 | 0.002 °C |
Chamber Temperature Uniformity: 4 | 0.05 °C |
Chamber Fluid Temperature Heating/Cooling Rate: | 0.3 °C per Minute (average) |
Gas Type: | Air or Nitrogen |
Gas Pressure Rating: (MAWP) | 175 psiG |
Gas Flow Rate Range: | 1 to 20 L/min |
Gas Flow Rate Specification: | 5% of full scale |
Saturation Pressure Range: | Ambient to 160 psiA |
Saturation Pressure Specification: | 0.02% of full scale |
Test Chamber Pressure Range: | Ambient |
Display Resolution: | 0.001 |
Test Chamber Dimensions: | 8" x 8" x 8" (203 mm x 203 mm x 203 mm) |
Access Port: | Ø1.875" (47.6 mm) located on right side |
Physical Dimensions: | 16.1" H x 27.9" W x 19.2" D (409 mm x 708 mm x 487 mm) |
Weight: (Generator Only) | 150 lbs (68 kg) |
UNCERTAINTY 5 & 6 | |
RH Uncertainty: 10 to 95 %RH, 5 to 60 °C, @ 20 L/min | 0.6% of reading |
RH Uncertainty: 95 to 98 %RH, ±10 °C of Ambient, @ 10 L/min | 0.6% of reading |
Dew/Frost Point Uncertainty: < 0 °C | 0.05 °C |
Dew Point Uncertainty: | 0.08 °C |
Temperature Uncertainty: 5 to 60 °C | 0.031 °C |
Test Chamber Pressure Uncertainty: Ambient | 0.007 psiA |
UTILITIES | |
Electrical Power: | 100-240 V~, 5 A, 50/60 Hz |
Gas Supply (External): | 175 psiG @ 20 L/min |
ENVIRONMENTAL | |
Operating Temperature: | 15 to 30 ºC |
Storage Temperature: | 10 to 50 ºC |
Humidity: | 5 to 95% Non-condensing |
Thunder 1220 Automated "Two-Pressure" Humidity Generation System - Data sheet 1.4 MB |