At Fluke Calibration, we receive many requests for precision PRT accuracy at thermocouple temperatures. Until now metrologist have settled for expensive, high-temperature SPRTs or inaccurate thermocouples for high-temperature measurement. We’re pleased to introduce the world’s finest high-temperature secondary PRT, our Model 5624.

The 5624 is ideal for use as a reference thermometer in high-temperature furnaces, the 5624 can reach a temperature of 1000°C with long-term drift at 0°C of 10 mK and accuracy (including short-term stability and calibration uncertainty over the full range) of 55 mK. Due to Fluke Calibration's proprietary sensor design, this PRT has short-term stability of 5 mK, and an immersion requirement of less than 153 mm (6 in) at 700°C.

The 5624 is assembled in an alumina sheath that is 508 mm (20") long and 6.35 mm (0.25 in) in diameter. Several termination configurations can be selected to match different thermometer readouts. Each 5624 comes with a NIST-traceable, NVLAP-accredited fixed-point calibration from 0°C - 962°C. That’s a fixed-point calibration! The 5624 also comes in a protective carrying case.

• Temperature range of 0°C - 1000°C
• Accuracy (includes short-term stability and calibration uncertainty) of +/- 0.05°C to 962°C
• Long-term drift of 0.01°C at 0°C after 100 h at 1000°C
• Designed by Fluke Calibration’s primary standards design team
• Includes 2609 probe case

Range 0C - 1000C
Calibration Uncertainty +/- 0.004C at 0C
+/- 0.020C at 962C
Long-term Drift (RTPW) (k = 2) < 0.01C at 0C/100 h at 1000C
< 0.06C at 0C/1000 h at 1000C
Calibrated Accuracy (k=2) +/- 0.006C at 0C
+/- 0.031C at 660C
+/- 0.041C at 961C
Short-Term Stability +/- 0.005C
Immersion < 153 mm (6 in) at 700C
RTPW 10 Ω (+/- 0.1 Ω)
Hysteresis < 0.005C from 0C - 1000C
Thermocycling < 0.01C, 10 cycles from 0C - 1000C
Current 1 mA
Size 6.35 mm (0.25 in) O.D.
Length 508 mm (20 in)
Sheath Material Alumina
Calibration Included 1913-6 fixed-point calibration
Includes calibration and 100 h drift