High-temperature secondary standards fill the gap between afforadable but temperature-limited PRTs and highly accurate SPRTs
Fluke Calibration's high-temp secondary standards fill the gap between affordable, but temperature-limited secondary PRTs (platinum resistance thermometers) and more expensive, highly accurate SPRTs (standard platinum resistance thermometers).
If you’re using block calibrators, furnaces, or temperature points above normal PRT (platinum resistance thermometer) temperatures (420°C), then this PRT is for you. The 5626 is nominally 100 Ω. The instrument has a temperature range of -200°C - 661°C. They make great working or check standards for calibration work up to the aluminum point.
Using a regular PRT at temperatures above 500°C exposes the platinum to contamination. If the PRT is used as a reference or calibration standard, contamination is a major problem. SPRTs (standard platinum resistance thermometers), which are more expensive and delicate, can handle the higher temperatures, but with greater risk to the instrument due to shock, contamination, or mishandling. The 5626 is designed to reduce the contamination risk through the use of internal protection while not impairing performance.
In addition to the right measurement performance and durability, a PRT for secondary applications should be priced affordably. Fluke Calibration’s new PRTs are inexpensive and come with an accredited calibration. The calibration comes complete with ITS-90 constants and a resistance-versus-temperature table.
• Range to 661°C
• Meets all ITS-90 requirements for resistance ratios
• Rtp drift < 20 mK after 500 hours at 661°C
• Calibrated accuracy of +/- 0.006°C at 0°C
Temperature Range | -200°C - 661°C |
Handle Temperature | 0°C - 80°C |
RTPW | 5626: 100 Ω (+/- 1 Ω) 5628: 25.5 Ω (+/- 0.5 Ω) |
W(Ga) | 31.11807 |
Calibrated Accuracy† (k = 2) | +/- 0.006°C at -200°C +/- 0.006°C at 0°C +/- 0.015°C at 420°C +/- 0.022°C at 661°C |
Stability | 5626: +/- 0.003°C 5628: +/- 0.002°C |
Long-Term Drift (k = 2) | 5626: < 0.006°C/100 h at 661°C 5628: < 0.004°C/100 h at 661°C |
Immersion | At least 12.7 cm (5 in) recommended |
Sheath | Inconel™ 600 |
Lead Wires | 4-wire Super-Flex PVC, 22 AGW |
Termination | Gold-plated spade lugs, or specify |
Size | 6.35 mm dia. x 305 mm, 381 mm, or 508 mm (0.25 in x 12 in / 15 in / 20 in) standard |
Calibration | Accredited calibrations from Fluke Calibration |
†Includes calibration and 100 h drift |
Fluke Calibration 5626 / 5628 Secondary Standard and High-Temp Platinum Resistance Thermometers - Manual 378.6 KB |