Resistor baths in three size options for any quantity of resistors
Regardless of the size and number of standard resistors you have to maintain, Fluke Calibration has a bath that will do the job for you. Choose one of the three models described here or call us for information on other sizes.
Like all Fluke Calibration baths, these resistor baths have unbeatable stability and uniformity. No other baths limit long-term and short-term drift—as well as gradients—better than these baths. Our proprietary controller senses temperature changes as small as 0.00001°C. This controller is the industry’s best-selling temperature calibration controller for bath retrofits because it improves the stability of almost every other poorly performing bath. So why not buy the best to begin with?
• Three size options for any quantity of resistors
• Stability to +/- 0.0007°C
• Independent high- and low-temperature cutout circuit
• Minimal long-term drift
7015 | 7009 | 7108 | |
Range | 0°C - 50°C† | 0°C - 50°C† | 20°C - 30°C |
Stability at 25°C | +/- 0.0007°C (water) | +/- 0.002°C (water) | |
Uniformity | +/- 0.003°C at 25°C (water) | +/- 0.005°C (water) | |
Temperature Setting | Digital display with push-button data entry | ||
Set-Point Resolution | 0.002°C; high-resolution mode, 0.00003°C | ||
Display Resolution | 0.01°C | ||
Digital Setting Accuracy | +/- 1°C | +/- 0.5°C | |
Digital Setting Repeatability | +/- 0.01°C | ||
Heaters | 500 W and 1000 W | Peltier Heating/Cooling | |
Cooling Capacity | 100 W - 200 W | 100 W in Ambient 23°C | |
Access Opening | 699 mm x 279 mm (27.5 in x 11 in) | 699 mm x 559 mm (27.5 in x 22 in) | 356 mm x 356 mm (14 in x 14 in) |
Bath Chamber Dimensions (H x W x D) (Unobstructed Space) | 699 mm x 279 mm x 330 mm | 669 mm x 559 mm x 330 mm | 356 mm x 203 mm x 355 mm |
Depth | 330 mm (13 in) | 203 mm (8 in) | |
Wetted Parts | 304 Stainless Steel | Tank: 304 Stainless Steel | |
Safety Cutout | Factory-set high temperature | N/A | |
Power | 115 VAC (+/- 10 %), 60 Hz, 15 A or 230 VAC, 50 or 60 Hz, 8 A, specify | 230 VAC (+/- 10 %), 50 or 60 Hz, 12 A (specify frequency) | 115 VAC (+/- 10 %), 50/60 Hz, 3 A or 230 VAC (+/- 10 %), 50/60 Hz, 1.6 A, specify |
Volume | 95 L (25 gal) | 167 L (44 gal) | 51 L (13.2 gal) |
Weight | 141 kg (310 lb.) | 150 kg (330 lb.) | 35 kg (75 lb.) |
Size (H x W x D) | 1219 mm x 1118 mm x 559 mm | 1092 mm x 1130 mm x 864 mm | 489 mm x 413 mm x 559 mm |
Automation Package | Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE. |
†Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with standard resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50°C to protect standard resistors placed in the bath.
Fluke Calibration 7009 / 7108 / 7015 Resistor Baths - 7009 Manual 411.9 KB |
Fluke Calibration 7009 / 7108 / 7015 Resistor Baths - 7108 Manual 1.6 MB |
Fluke Calibration 7009 / 7108 / 7015 Resistor Baths - 7015 Manual 2.7 MB |