Skip to main content
Press Enter
Buy now! $9,357.59CAD
Fluke

725EX Series Process Calibrator

Intrinsically safe multifunction process calibrator

Buy now!
$9,357.59CAD
Have questions? We're here to help!
Have questions? We're here to help!

 The 725Ex is an intrinsically safe multifunction process calibrator for use in explosion endangered areas. It is powerful, yet easy to use. Combine the 725EX with any of the eight Fluke 750PEx Pressure Modules (sold separately), and you’ll be able to calibrate almost any process instrument likely to need service in an area where explosive gasses may be present.

 
The 725EX sources and measures almost all process parameters including mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules. Its split display lets you view input and output values simultaneously to calibrate transmitters. For valve and I/P tests, you can source mA while measuring pressure. The 725Ex has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests.
 
From the moment you pick it up, the Fluke 725Ex is ready to perform. Its simple controls, without menus, make operation easy. Its memory functions make set up fast. And its rugged design lets it work as hard as you do.
 
Other useful features:
 
• Pressure switch test function to capture set, reset, and deadband values
• Small, streamlined shape, and rugged, reliable design that stands up to field use
• Simple, push-button user interface for easy operation
• Provides two-channel simultaneous source and measure capability to calibrate transmitters
• Powers transmitters during test using loop supply
• Stores frequently-used test setups for later use
• Measures pressure to 3,000 psi/200 bar using any of the eight intrinsically safe Fluke 700PEx Pressure Modules to calibrate pressure transmitters and I/P devices
• Intrinsically safe multifunction process calibrator
• Meets I.S. Class I, Division 1 Groups B-D, 171°C compliance standards
• Measures volts DC, mA, RTDs, thermocouples, frequency, and ohms
• Sources or simulates volts DC, mA, RTDs, thermocouples, frequency and ohms
Current DC24.000 mA 0.02% + 2 counts
Frequency2.0 to 1000.0 CPM 0.05% + 1 count
1.0 to 1100.0 Hz 0.05% + 1 count
1.00 to 10.00 kHz 0.05% + 1 count
Sensitivity 1 V peak-to-peak-minimum
PressureAccuracy from 0.025% of range using any of 8 intrinsic safe pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy100.00 mV 0.02% + 2 counts
10.000 V 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
Current DC24.000 mA (source) 0.02% + 2 counts
24.000 mA (simulate) 0.02% + 2 counts
Resistance
15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω 0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω 1 Ω (excitation current 0.05 to 0.4 mA)
Frequency
2.0 to 1000.0 CPM 0.05%
1.0 to 1100.0 Hz 0.05%
1.00 to 10.00 kHz 0.25%
Waveform 5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure accuracy
NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
J 0.7°C
K0.8°C
T0.8°C
E0.7°C
R1.8°C
S1.5°C
B1.4°C
L0.7°C
U0.75°C
N0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
XK 0.6°C
BP 1.2°C
Source accuracy
NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
Accuracy stated for 4-wire measurement.
J 0.7°C
K 0.8°C
T0.8°C
E0.7°C
R1.4°C
S1.5°C
B1.4°C
L0.7°C
U0.75°C
N0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
XK 0.6°C
BP 1.2°C
Specifications
Ramp functions
Source functions Voltage, current, resistance, frequency, temperature
RampsSlow ramp, Fast ramp, 25% step-ramp
Loop power function
Voltage 12 V
Accuracy 10%
Maximum current 22 mA, short circuit protected
Step Functions
Source functions Voltage, current, resistance, frequency, temperature
Steps 25% of range, 100% of range
Safety specifications
Agency approvals I.S. Class I, Division 1 Groups B-D
Mechanical and General Specifications
Size 130 x 236 x 61 mm (5.188 x 9.291 x 2.402 in)
Weight 0.85 kg (1.874 lbs.)
Batteries 4 AA alkaline batteries
Warranty Three year
Battery replacement Separate battery compartment, accessible without breaking calibration seal
Side port connections Pressure module connector
Measurement Accuracy
Voltage DC30.000 V 0.02% + 2 counts (upper display)
10.000 V 0.02% + 2 counts (lower display)
90.00 mV 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
Resistance0.0 to 400.0 Ω 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω 1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
  • Thumbnail of document Data Sheet - 725EX Series Process Calibrator
    Download

    Data Sheet - 725EX Series Process Calibrator

    Download
  • Thumbnail of document Manual - 725EX Series Process Calibrator
    Download

    Manual - 725EX Series Process Calibrator

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more