Keeps your cells up and running reliably for weeks at a time - even during heavy usage
Maintain two triple point of water cells for weeks in a Fluke Calibration 7312 temperature bath or maintain Gallium cells for your fixed point calibrations. This temperature bath can be used as a calibration bath from -5°C - 110°C.
• Maintains TPW cells for up to six weeks
• Optional immersion freezer for simple cell freezing
• Independent cutout circuit protects cells from breaking
Range | -5°C - 110°C |
Stability | +/- 0.001°C at 0°C (alcohol-water mix) +/- 0.004°C at 30°C (alcohol-water mix) |
Uniformity | +/- 0.003°C at 0°C (alcohol-water mix) +/- 0.006°C at 30°C (alcohol-water mix) |
TPW Duration | Six weeks, typical (assumes correctly formed ice mantle) |
Set-Point Accuracy | +/- 0.05°C at 0°C |
Set-Point Repeatability | +/- 0.01°C |
Display Resolution | +/- 0.01°C |
Set-Point Resolution | +/- 0.002°C; 0.00003°C in high-resolution mode |
Access Opening | 121 mm x 97 mm (4.75 in x 3.8 in) |
Immersion Depth | 496 mm (19.5 in) |
Volume | 19 L (5 gal) |
Communications | RS-232 included |
Power | 115 VAC (+/- 10%), 60 Hz |
Size (H x W x D) | 819 mm x 305 mm x 622 mm (12 in x 24.5 in x 32.25 in) |
Weight | 34 kg (75 lb) |
Fluke Calibration 7312-16 Triple Point of Water Maintenance Bath - Manual 1.4 MB |
Fluke Calibration 7312-16 Triple Point of Water Maintenance Bath - Manual Supplement 327.5 KB |