Compact baths with the stability and uniformity required for precision thermometer testing
Find a high temperature bath (300°C) or refrigerated calibration bath for low temperature calibration (i.e. -80°C), in compact packages from Fluke Calibration. Fluke Calibration metrology baths are well known world-wide for excellent temperature control that maintains excellent temperature stability (+/- 0.005°C) and uniformity (+/- 0.005°C).
When you only need a circulator or utility bath to control a process within a few degrees or to maintain biological test samples, talk to a utility bath manufacturer. But when you’re doing precision thermometer testing, and stability and uniformity are critical to the success of your work, talk to us.
Fluke Calibration has been making the world’s best-performing temperature baths for almost two decades. With our proven heating/cooling designs and hybrid analog-digital controller, Fluke Calibration baths apply the most effective technologies that are commercially feasible. These four compact baths are no exception.
• Stability and uniformity each better than +/- 0.008°C
• Metrology-level performance in lab-friendly sizes
• Convenient use on benchtops or on matching carts
. | 6330 | 7320 | 7340 | 7380 |
Range | 35°C - 300°C | -20°C - 150°C | -40°C - 150°C | -80°C - 100°C |
Stability | +/- 0.005°C at 100°C (oil 5012) | +/- 0.005°C at -20°C (ethanol) | +/- 0.005°C at -40°C (ethanol) | +/- 0.006°C at -80°C (ethanol) |
Uniformity | +/- 0.007°C at 100°C (oil 5012) | +/- 0.005°C at –20°C (ethanol) | +/- 0.006°C at -40°C (ethanol) | +/- 0.008°C at -80°C (ethanol) |
Heating Time† | 250 min, from 35°C - 300°C (oil 5017) | 80 min, from 25°C - 150°C (oil 5012) | 60 min, from 25°C - 150°C (oil 5012) | 25 min, from 25°C - 100°C (oil 5010) |
Cooling Time | N/A | 100 min, from 25°C to -20°C (oil 5012) | 110 min, from 25°C to -40°C (ethanol) | 130 min, from 25 °C to -80 °C (ethanol) |
Stabilization Time | 15 -20 min | |||
Temperature Setting | Digital display with push-button data entry | |||
Set-Point Resolution | 0.01°; 0.00018° in high-resolution mode | 0.01° | ||
Display Resolution | 0.01 ° | |||
Digital Setting Accuracy | +/- 0.5 °C | |||
Digital Setting Repeatability | +/- 0.01 °C | |||
Access Opening | 94 mm x 172 mm (3.7 in x 6.8 in) | 86 mm x 114 mm (3.25 in x 4.5 in) | ||
Working Area | 81 mm x 133 mm (3.2 in x 5.25 in) | 86 mm x 114 mm (3 in x 4 in) | ||
Depth | 234 mm (9.25 in) | 178 mm (7 in) | ||
Wetted Parts | 304 Stainless Steel | |||
Power | 115 V AC (+/- 10%), 50/60 Hz, 7 A or 230 V AC (+/- 10 %), 50/60 Hz, 3.5 A, specify | 115 V AC (+/- 10 %), 60 Hz, 15 A or 230 V AC | 115 V AC (+/- 10 %) 60 Hz, 16 A or 230 V AC | |
Volume | 9.2 L (2.4 gal) | 4 L (1 gal) | ||
Size (W x D x H) | 305 mm x 546 mm x 470 mm (12 in x 21.5 in x 18.5 in) | 305 mm x 622 mm x 584 mm (12 in x 24.5 in x 23 in) off cart; 305 mm x 622 mm x 819 mm (12 in x 24.5 in x 32.25 in) on cart | 305 mm x 610 mm x 762 mm (12 in x 24 in x 30 in) | |
Weight | 19 kg (42 lb) | 35.4 kg (78 lb) | 52 kg (115 lb) | |
Automation Package | Interface-it software and RS-232 included (IEEE-488 optional) |
†Rated at nominal 115 V (or optional 230 V)
Fluke Calibration 7340 Compact Temperature Calibration Bath - 7340 - Manual 3.6 MB |