` 753 Documenting Process Calibrator

Fluke

753 Documenting Process Calibrator

Sources, simulates and measures pressure, temperature, and electrical signals


$11,118.00 CAD
Normally ships in 4-6 Weeks

 
  Request Quote

Can We Help You?

CALL US   EMAIL US   CHAT LIVE

Our product experts are here to help you find the right solution for your application!

The Fluke 753 is a powerful multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software, and upload data for printing, archiving, and analysis. You'll find it does the work of several tools, with simultaneous source and measurement capabilities for all common process parameters. It sources, simulates, and measures pressure, temperature, and electrical signals in one rugged, hand-held calibration tool. It also automates calibration procedures, captures data for documentation, and helps you meet rigorous standards like ISO 9000, FDA, EPA, and OSHA regulations.
 
The 753 includes a USB interface and a USB communication cable to enable two-way communication with DPCTrack 2 and other instrumentation management applications. To create a seamless/paperless calibration management system consider adding Fluke DPCTrack2 Calibration Management software for use with your Fluke 753 and 754 or even legacy Fluke 743 and 744 calibrators.

• Full-featured documenting process calibrator for calibrating and troubleshooting
• Measures volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
• Sources and simulate volts, mA, pressure, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
• Powers transmitters during test using loop supply with simultaneous mA measurement

Measurement Accuracy

Voltage DC

 

1 Year

2 Years

100.000 mV

0.02% + 0.005 mV

0.03% + 0.005 mV

3.00000 V

0.02% + 0.00005 V

0.03% + 0.00005 V

30.0000 V

0.02% + 0.0005 V

0.03% + 0.0005 V

300.00 V

0.05% + 0.05 V

0.07% + 0.05 V

Voltage AC

Range

40 Hz – 500 Hz

 

Resolution

1 Year

2 Years

3.000 V

0.001 V

0.5% + 0.002 V

1.0% + 0.004 V

30.00 V

0.01 V

0.5% + 0.02 V

1.0% + 0.04 V

300.0 V

0.1 V

0.5% + 0.2 V

1.0% + 0.2 V

Current DC

 

1 Year

2 Years

30.000 mA

0.01% + 5 µA

0.015% + 7 µA

110.00 mA

0.01% + 20 µA

0.015% + 30 µA

Resistance

 

1 Year

2 Years

10.000 Ω

0.05% + 50 mΩ

0.07% + 70 mΩ

100.00 Ω

0.05% + 50 mΩ

0.07% + 70 mΩ

1.0000 kΩ

0.05% + 500 mΩ

0.07% + 0.5 Ω

10.000 kΩ

0.1% + 10 Ω

0.15% + 15 Ω

Frequency

 

Resolution Accuracy (2 Years)

1.00 – 110.00 Hz

0.01 Hz

0.05 Hz

110.0 – 1100.0 Hz

0.1 Hz

0.5 Hz

1.101 – 11.000 kHz

0.001 kHz

0.005 kHz

11.01 – 50.00 kHz

0.01 kHz

0.05 kHz

           

 

Source Accuracy

 

   

1 Year

2 Years

 

Voltage DC

 

100.00 mV

0.01% + 0.005 mV

0.015% + 0.005 mV

 

1.00000 V

0.01% + 0.00005 mV

0.015% + 0.0005 V

 

15.0000 V

0.01% + 0.0005 mV

0.015% + 0.0005 V

 

Current DC

 

1 Year

2 Years

 

22.000 mA (Source)

0.01% + 0.003 mA

0.02% + 0.003 mA

 

Current Sink (Simulate)

0.02% + 0.007 mA

0.04% + 0.007 mA

 

Resistance

 

1 Year

2 Years

 

10.000 Ω

0.01% + 10 mΩ

0.015% + 15 mΩ

 

100.00 Ω

0.01% + 20 mΩ

0.15% + 30 mΩ

 

1.0000 kΩ

0.02% + 0.2 Ω

0.03% + 0.3 Ω

 

10.000 kΩ

0.02% + 3 Ω

0.03% + 5 Ω

 

Frequency

 

2 Years

 

0.1 – 10.99 Hz

0.01 Hz

 

0.01 – 10.99 Hz

0.01 Hz

 

11.00 – 109.99 Hz

0.1 Hz

 

110.0 – 1099.9 Hz

0.1 Hz

 

1.100 – 21.999 kHz

0.002 kHz

 

22.000 – 50.000 kHz

0.005 kHz

 
         

 

 

Technical Data

Data Log Functions

Measure Functions

Voltage, current, resistance, frequency, temperature, pressure

Reading Rate

1, 2, 5, 10, 20, 30, or 60 readings/min

Maximum Record Length

8000 readings (7980 for 30 or 60 readings/min

Ramp Functions

Source Functions

Voltage, current, resistance, frequency, temperature

Rate

4 steps/s

Trip Detect

Continuity or voltage (continuity detection not available when sourcing current)

Loop Power Function

Voltage

Selectable, 26 V

Accuracy

10%, 18 V minimum at 22 mA

Maximum Current

25 mA, short circuit protected

Maximum Input Voltage

50 V DC

Step Functions

Source Functions

Voltage, current, resistance, frequency, temperature

Manual Step

Selectable step, change with arrow buttons

Auto Step

Fully programmable for function, start delay, stepvalue, time per step, repeat

 

Environmental Specifications

Temperature

Operating

-10°C - 50°C

Storage

-20°C - 60°C

Dust/Water Resistance

Meets IP52, IEC 529

Operating Altitude

3,000 m above mean sea level (9842 ft)

 

Safety Specifications

Agency Approvals

 

 CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

 

Mechanical & General Specifications

Size

 

136 mm x 245 mm x 63 mm (5.4 in x 9.6 in x 2.5 in)

Weight

 

1.2 kg (2.7 lb)

Batteries

Internal Battery Pack Lithium-ion:

 7.2 V, 4400 mAh, 30 Wh

Battery Life

 

 > 8 h typical

Battery Replacement

 

 Via snap-shut door without opening calibrator; no tools required

Side Port Connections

 

 Pressure module connector

 

 USB Connector to interface to your PC

 

 Connection for optional battery charger/eliminator

Data Storage Capacity

 

 1 week of calibration procedures results

90 Day Specifications

 

The standard specification interval for the 750 Series are 1 and 2 years.

 

Typical 90 day measurement and source accuracy can be estimated by

 

 dividing the one year "% of reading" or "%of output" specifications by 2.

 

Floor specifications, expressed as "% of full scale" or

 

 "counts" or "ohms" remain constant.

 

Temperature, Resistance Temperature Detectors

Degrees or % of Reading

Type (α)

Range (°C)

Measure °C1

Source Current

Source °C

Allowable Current2

 

 

1 year

2 year

1 year

2 year

100 Ω Pt (385)

-200 - 100
100 - 800

0.07°C
0.02 % + 0.05°C

0.14 °C
0.04 % + 0.10 °C

1 mA

0.05°C
0.0125 % + 0.04°C

0.10 °C
0.025 % + 0.08 °C

0.1 mA - 10 mA

200 Ω Pt (385)

-200 - 100
100 - 630

0.07°C
0.02 % + 0.05°C

0.14 °C
0.04 % + 0.10 °C

500 μA

0.06°C
0.017 % + 0.05°C

0.12 °C
0.034 % + 0.10 °C

0.1 mA - 1 mA

500 Ω Pt (385)

-200 - 100
100 - 630

0.07°C
0.02 % + 0.05°C

0.14 °C
0.04 % + 0.10 °C

250 μA

0.06°C
0.017% + 0.05°C

0.12 °C
0.034 % + 0.10 °C

0.1 mA - 1 mA

1000 Ω Pt (385)

-200 - 100
100 - 630

0.07°C
0.02 % + 0.05°C

0.14 °C
0.04% + 0.10°C

150 μA

0.06°C
0.017 % + 0.05°C

0.12 C
0.034 % + 0.10 °C

0.1 mA - 1 mA

100 Ω Pt (3916)

-200 - 100
100 - 630

0.07°C
0.02 % + 0.05°C

0.14 °C
0.04 % + 0.10 °C

1 mA

0.05°C
0.0125 % + 0.04°C

0.10 °C
0.025 % + 0.08 °C

0.1 mA - 10 mA

100 Ω Pt (3926)

-200 - 100
100 - 630

0.08°C
0.02 % + 0.06°C

0.16 °C
0.04 % + 0.12 °C

1 mA

0.05°C
0.0125 % + 0.04°C

0.10 °C
0.025 % + 0.08 °C

0.1 mA - 10 mA

10 Ω Cu (427)

-100 - 260

0.2°C

0.4 °C

3 mA

0.2°C

0.4 °C

0.1 mA - 10 mA

120 Ω Ni (672)

-80 - 260

0.1°C

0.2 °C

1 mA

0.04°C

0.08 °C

0.1 mA - 10 mA

1For two and three-wire RTD measurements, add 0.4°C to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms

 

Temperature, Thermocouples

Type

Source °C

Measure °C

Source °C

   

1 year

2 years

1 year

2 years

E

-250 to -200

1.3

2.0

0.6

0.9

-200 to -100

0.5

0.8

0.3

0.4

-100 to 600

0.3

0.4

0.3

0.4

600 to 1000

0.4

0.6

0.2

0.3

N

-200 to -100

1.0

1.5

0.6

0.9

-100 to 900

0.5

0.8

0.5

0.8

900 to 1300

0.6

0.9

0.3

0.4

J

-210 to -100

0.6

0.9

0.3

0.4

-100 to 800

0.3

0.4

0.2

0.3

800 to 1200

0.5

0.8

0.3

0.3

K

-200 to -100

0.7

1.0

0.4

0.6

-100 to 400

0.3

0.4

0.3

0.4

400 to 1200

0.5

0.8

0.3

0.4

1200 to 1372

0.7

1.0

0.3

0.4

T

-250 to -200

1.7

2.5

0.9

1.4

-200 to 0

0.6

0.9

0.4

0.6

0 to 400

0.3

0.4

0.3

0.4

B

600 to 800

1.3

2.0

1.0

1.5

800 to 1000

1.0

1.5

0.8

1.2

1000 to 1820

0.9

1.3

0.8

1.2

R

-20 to 0

2.3

2.8

1.2

1.8

0 to 100

1.5

2.2

1.1

1.7

100 to 1767

1.0

1.5

0.9

1.4

S

-20 to 0

2.3

2.8

1.2

1.8

0 to 200

1.5

2.1

1.1

1.7

200 to 1400

0.9

1.4

0.9

1.4

1400 to 1767

1.1

1.7

1.0

1.5

C

0 to 800

0.6

0.9

0.6

0.9

800 to 1200

0.8

1.2

0.7

1.0

1200 to 1800

1.1

1.6

0.9

1.4

1800 to 2316

2.0

3.0

1.3

2.0

L

-200 to -100

0.6

0.9

0.3

0.4

-100 to 800

0.3

0.4

0.2

0.3

800 to 900

0.5

0.8

0.2

0.3

U

-200 to 0

0.6

0.9

0.4

0.6

0 to 600

0.3

0.4

0.3

0.4

BP

0 to 1000

1.0

1.5

0.4

0.6

1000 to 2000

1.6

2.4

0.6

0.9

2000 to 2500

2.0

3.0

0.8

1.2

XK

-200 to 300

0.2

0.3

0.2

0.5

300 to 800

0.4

0.6

0.3

0.6

 

 

Item
Description
Availability
Price
Rubber overmolded surfaces and finger-hugging curves give the user a comfortable, reliable grip on the accessory so they can focus on making an accurate measurement
Normally ships in 4-6 Weeks
$85.00 CAD
Improve steadiness in slipper hands with comfortable rubber overmolded surfaces and finger-hugging curves to give a reliable grip so you can focus on making an accurate measurement
Normally ships in 4-6 Weeks
$70.00 CAD
SureGrip alligator clips
Normally ships in 4-6 Weeks
$58.00 CAD
One pair (red, black) with tips which include banana-style spring contacts with nickel-plated brass ends
Normally ships in 4-6 Weeks
$40.00 CAD
Set of push-on adapters for TL71 and TL75 test probes
Normally ships in 4-6 Weeks
$77.00 CAD
Hand operated pressure pump to provide pressures up to 150 PSI (1000 kPa)
Normally ships in 4-6 Weeks
$880.00 CAD
For use in connecting a Fluke-700HTP-1 or other hydraulic test pump to a unit under test
Normally ships in 4-6 Weeks
$1,051.00 CAD
Protects the most commonly used Fluke pressure modules from damage due to accidental over-pressurization
Normally ships in 4-6 Weeks
$1,230.00 CAD
For use with the 700 Series pressure modules or test gauges to supply pressures up to 600 PSI (40 bar)
Normally ships in 4-6 Weeks
$1,382.00 CAD
Hand operated pressure pump designed to generate calibration pressures to 10,000 PSI (690 bar)
Normally ships in 4-6 Weeks
$2,123.00 CAD
Fluke BC7240 Battery Charger/Eliminator for the 75x Documenting Process Calibrator
Normally ships in 4-6 Weeks
$141.00 CAD
Lithium ion battery pack for the 75X documenting process calibrator
Normally ships in 4-6 Weeks
$344.00 CAD
Hard carrying case with custom-cut foam liner to hold a documenting process calibrator, a pressure module, a battery charger and spare pack as well as test leads
Normally ships in 4-6 Weeks
$390.00 CAD
Zippered carrying case with detachable external pouch
Normally ships in 4-6 Weeks
$130.00 CAD
For use with the 753, 754 and 743, 744 calibrators
Normally ships in 4-6 Weeks
$6,958.00 CAD
Isothermal screw terminal for J wire
Normally ships in 4-6 Weeks
$58.00 CAD
Isothermal screw terminal for K wire
Normally ships in 4-6 Weeks
$58.00 CAD
Type-J thermocouple for general purpose applications
Normally ships in 4-6 Weeks
$58.00 CAD
Extending and repairing J-type thermocouple wires
Normally ships in 4-6 Weeks
$123.00 CAD
Type-K thermocouple for general purpose applications
Normally ships in 4-6 Weeks
$82.00 CAD
Type K thermocouple for use in air and non-caustic gas measurements
Normally ships in 4-6 Weeks
$219.00 CAD
Type K thermocouple suitable for food industry, liquids and gels
Normally ships in 4-6 Weeks
$219.00 CAD
Type K thermocouple with tapered tip for use in air, non-caustic gas and surface applications
Normally ships in 4-6 Weeks
$219.00 CAD
Type K thermocouple surface, air and non-caustic gases
Normally ships in 4-6 Weeks
$137.00 CAD
Extending and repairing K-type thermocouple wires
Normally ships in 4-6 Weeks
$123.00 CAD
Operates with T-type thermometers
Normally ships in 4-6 Weeks
$212.00 CAD
Immersion temperature probe designed for use in liquid and gels
Normally ships in 4-6 Weeks
$185.00 CAD
Designed to reliably measure temperatures on flat or slightly convex surface
Normally ships in 4-6 Weeks
$562.00 CAD
Type K thermocouple for fast temperature and superheat measurements of pipe surfaces
Normally ships in 4-6 Weeks
$315.00 CAD
1 pair (red, black) of test leads with multi-stacking 4 mm banana plugs and mini-hooks
Normally ships in 4-6 Weeks
$54.00 CAD
1 pair (red, black) of test leads with multi-stacking 4 mm banana plugs and mini-pincers
Normally ships in 4-6 Weeks
$86.00 CAD