Fluke Calibration

7900 Controller for Rosemount-Designed Baths

Easy-to-install controller designed for use with Rosemount baths

Starting At
$12,687.00 CAD
Normally ships in 1-2 Weeks

Configure   Request Quote

Can We Help You?

CALL US   EMAIL US   CHAT LIVE

Our product experts are here to help you find the right solution for your application!

Fluke Calibration’s bath controllers have long been recognized as being among the finest in the world. They’re the most popular retrofit controller in the industry, and now they’re available for Rosemount baths. The Model 7900 Controller installs easily and can replace the Rosemount Model 915 for all Rosemount bath models.

This controller uses the same circuitry as Fluke Calibration’s 2100 Controller to achieve long-lasting stabilities of +/- 0.003°C or better. Special noise-rejection techniques allow the 7900 to measure the very tiny resistance changes required for this level of stability. AC bridges are used within the controller to cancel thermal EMFs. Custom high-precision resistors contribute to short- and long-term stability and advanced filtering techniques force out troublesome line noise. The Model 7900 includes a special circuit that monitors the controller’s microprocessor and automatically resets it if its operations are interrupted. Two separate cutout systems are also included for keeping your bath’s temperature within its normal range.

A software cutout uses an adjustable high-temperature limit that can be easily accessed through the front panel and set to match the requirements of your bath fluid. Should the control sensor measure a temperature beyond this upper limit, heating is shut down. If the bath’s temperature falls below its normal operating range, the heaters are turned on and the LN2 cooling is shut off. A second, independent hardware cutout monitors the bath’s temperature with a thermocouple and shuts down all heating and LN2 cooling if the bath’s temperature rises above its range.

These cutout features, combined with the superior reliability and long-term stability performance of the 7900, allow you to run your system for as long as you like between shut-downs—365 days a year, if you wish. Your bath can be ready for you to take measurements the minute you walk into your lab each day.

• All the features of the Fluke Calibration 2100 Controller
• Installs easily
• Two independent over-temperature cutout circuits

Temperature Control Range -100C - 400C
Optional Ranges None
Stability +/- 0.003 (+/- 0.001 typical)
Stabilization Time 30 min
Display Accuracy +/- 1C
Cooling Control LN2 automatic
Heating Control 2-position, firmware or user controlled
Firmware High-Temp Cutout Yes, volatile, programmable (independent of the controller)
Hardware High-Temp Cutout Thermocouple controlled
Memory Non-volatile; 8 programmable set-points, each with ramp and soak features
Programmable Soak Time 1 to 500 min
Control Sensor 100 Ω PRT
α = 0.00385
Interface RS-232 and IEEE standard
Software Interface-it
Operating Temperature 5C - 50C
Operating Voltage 115 VAC (+/- 10%), 60 Hz
Current Rating 20 A maximum
Dimensions (W x H X D) 311 mm x 114 mm x 279 mm (12.25 in x 4.5 in x 11 in)
Weight 4 kg (9 lb)
Installation Freestanding or rack-mounted with optional hardware