One, two or three stage sensors available - integral or remote from controller
Measurement Performance: | |
Relative Humidity and ppmv Ranges: | 0 to 100% RH with all sensors, non-condensing conditions 0 to 999,999 ppmv with all sensors, non-condensing conditions |
Accuracy, All Sensors: | ± 0.2°C (0.36°F) dew/frost point standard ± 0.1°C (0.18°F) dew/frost point optional ± 0.5% of full scale pressure |
Remote Sensors: | Up to 250 feet (76 meters) from console with cable |
Display Resolution: | 0.1°C or °F 0.01 psia 0.1 ppmv below 1000 ppmv 1 ppmv above 1000 ppmv |
Air Temperature Sensor Range: | -100 to 250°C (-148 to 482°F) ± 0.1°C accuracy |
Packaging: | |
Weight: | Bench-Top: 6 lbs. (2.8 kg) Rack Mount: 7 lbs. (3.2 kg) NEMA 4: 19 lbs. (8.6 kg) |
Dimension: | Bench-Top with S3 Mounted: 11 in. W x 5.2 in. H x 17 in. D (27.9 cm W x 13.2 cm H x 43.2 cm D) Rack Mount with S3 Mounted: 19 in. W x 7 in. H x 17 in. D (48.3 cm W x 17.8 cm H x 43.2 cm D) Standard rack panel mounting dimensions NEMA 4: 11.6 in. W x 13.6 in. H x 6.2in D (29.5cm W x 34.6 cm H x 15.9 cm D) |
Mounting: | Bench-top: Standard Rack Mount: Optional Wall-mount NEMA 4: Optional |
Sensor Materials: | Chromium, glass, epoxy, and anodized aluminum or 316 stainless steel |
Functional: | |
Power Requirements: | 90-230 VAC ±10%, 50-400Hz |
Operating Temperature: | Control Unit: -10 to 60°C (15 to 140°F) Sensors: -40 to 100°C (-40 to 212°F) |
Auxiliary Coolant: | To augment temperature depression capability of sensor when necessary: Water (or other liquid): 0.5 gal/min (2 l/min) at 100psia maximum |
Sample Pressure Range: | Depends on sensor model, see chart above and consult Edgetech Instruments |
Sample Flow Rate: | 0.5/5.0 SCFH (0.25 to 2.4 l/min) |
Outputs (3 outputs available): | Outputs (3 outputs available): 4-20 mA, 0-5 VDC, RS232 |
Track or Hold: | Outputs, analog, digital, alarms can be set to track or hold while in Automatic Balance Control or programming mode |
Alarms: | Two Form C, SPDT alarm relays rated for 3 amps at 24 VDC, 120 VAC. Alarm mode (high or low) programmable from keyboard or RS232. Alarm set points programmable from -99.9 to 99.9°C (-148 to 212°F) from keyboard or via RS232, alarm can be latched or un |
Display: | 8 line LCD graphic data display, backlight 3 parameters display simultaneously |
Accessories/Options: | |
Remote Mounting Kit: | Remote sensor cable, standard length 10 ft.(3 m) Lengths up to 250 ft. (76 m) available |
Panel/Rack Mounting Kit: | Optional |
NEMA-4 Wall Mount Enclosure: | Optional |
Explosion-Proof Sensor Housing: | Optional |
In-Line Particulate/Coalescing Filter: | Optional |
Sampling Module: | Optional, consists of a vacuum pump, 10.7 - 19.2 psia (0.75 - 1.35 kg/cm2) variable area flowmeter assembled in a weather tight NEMA-4 enclosure, with bulkhead fittings. Specify 115 or 230 VAC, 50 or 60 Hz |
View All Part Numbers