Suitable for climatic and temperature chambers, dryers, air ducts
The handheld probe is especially suitable for mobile measurements of high temperatures. It measures humidity and temperature and calculates the dew/frost point.
Applications
Climatic and temperature chambers, dryers, air ducts
Probe type | High-Temperature Handheld Probe |
Operating range | -40…100 °C / 0…100 %rh |
Accuracy at 23 ±5 °C | ±0.8 %rh / ±0.1 K |
Accuracy w. adjustment profile "standard" | at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K |
Power supply / Power consumption | 3.2…5 VDC ±0 % / < 4.5 mA |
Long-term stability | <1 %rh / year |
Humidity sensor | Hygromer IN-1 |
Temperature sensor | PT100 1/3 DIN Class B |
Probe protection | Filter carrier, brass nickel plated |
Filter cartridge | Wire mesh, 20...25 um |
Response time t 63 | <15 sec. |
Allowed wind speed | 20 m/s, with filter |
Psychrometric calculations | Dew or frost point |
Reproducibility Airchip | <0.02 %rh / 0.01 K |
Exploitation software | ROTRONIC HW4 |
Probe adjustment by software | 1 Point & multipoint %rh & °C |
Firmware update | With internet and HW4 software |
User information | From Software HW4 (devicemanager) |
Data processing via HW4 | With interface cables |
Device protection by password | Yes |
Type of output signals | 2 x 0…1 VDC |
Analog outputs scaleable by user | Yes |
Analog output signal (standard) | 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C |
Digital ooutput signal (standard) | UART |
Service interface | UART |
Reverse voltage protection | Mechanically in the connector |
Max. cable length | 2 m |
Probe material | PEEK |
Enclosure material | POM handle / sensor tube: PPS, black |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |
Weight | 210 g / 240 g |
View All Part Numbers