Skip to main content
Press Enter
Starting at $817.00CAD
Configure
In Stock *
Rotronic

HC2A-S3A/HC2-S3H Meteorolgy Probes

In Stock *

The most versatile probe from Rotronic and forms the basis of their product portfolio, measures RH, temperature and calculates dew/frost point

Starting at
$817.00CAD
Some configurations are in stock for immediate delivery!
Have questions? We're here to help!
Have questions? We're here to help!
Configure
The standard Rotronic Humidity and Temperature probes can be used in application with temperature from -50…100°C / 0…100 % RH. With an outstanding accuracy ± 0.8% RH ±0.1°C. For calibration the probes can be replaced without almost any downtime. Together with a Rotronic HF5, PF4/5 transmitter or HP32 and HP23 this probe is the first selection for many applications.
 
 
The HC2A-S3 / -S3H are the most sold versatile probes from Rotronic and form the basis of the product portfolio. They measure humidity and temperature and calculate the dew/frost point.
• Range of application -50…100 °C / 0…100 %RH
• Accuracy: ±0.8 %RH, ±0.1 K, at 23 °C ±5 K
• Digital interface (UART) and scalable analog outputs, 0…1 V
• GAMP compatible (HygroSoft Software)
Exploitation SoftwareROTRONIC HW4
Probe Adjustment by Software1 Point & multipoint %rh & °C
Probe TypeMeteorological probe
Operating Range-50…100 °C / 0…100 %rh
Accuracy at 23 °C±0.8 %RH / ±0.1 K
Accuracy w. Adjustment Profile "High Precision"±0.5 %RH / ±0.1 K, at 10...30 °C
at 23 °C and 10, 20, 30, 40, 50, 60, 70, 80, 90 %RH
Power Supply / Power Consumption3.3...5 VDC, Calibration @ 3.3 VDC / ~4.5 mA
Long-Term StabilityDrift <1%RH per year with clean air or similar. (see pollutant table in catalog etc.)
SensorsHumidity: Rotronic HT-1 / Temperature PT100 1/3 DIN Class B
Temperature SensorPT100 1/3 DIN Class B
Probe ProtectionPolycarbonate plastic cage, white
Filter CartridgePolyethylene dust filter, white, 40 um
Filter TypePolyethylene, 40 μm (SPA-PCB-WM)
Response Time T 63<15 sec.
Allowed Wind Speed20 m/s, with filter
Psychrometric CalculationsDew or frost point
Reproducibility Airchip<0.02 %rh / 0.01 K
Firmware UpdateWith internet and HW4 software
User InformationFrom Software HW4 (devicemanager)
Data Processing Via HW4With interface cables
Device Protection by PasswordYes
Type of Output Signals2 x 0…1 VDC
Analogue Outputs Scaleable by the UserYes
Analog Output Signal (Standard)0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Digital Output Signal (Standard)UART
Communication InterfacesUART
Service InterfaceUART
Reverse Voltage ProtectionMechanically in the connector
Max. Cable LengthUp to 5 m with passive cable. Up to 100 m with active cable
Enclosure MaterialPolycarbonate
FDA/GAMP CompatibilityFDA 21 CFR Part 11 and GAMP compatible
DimensionsØ15 x 108 mm
Weight17 g
Probe HousingPolycarbonate (white)
Dimension Measurement HeadØ 15 x 108 mm
Output SignalsDigital UART / 2x 0…1 VDC
CompatibleHF5, HP32, HP23A
• Pharmaceutical industries
• Incoming goods inspection
• Food Industry
  • Thumbnail of document Manual - HC2A-S3A/HC2-S3H Meteorolgy Probes
    Download

    Manual - HC2A-S3A/HC2-S3H Meteorolgy Probes

    Download
  • Thumbnail of document Data Sheet - HC2A-S3A/HC2-S3H Meteorolgy Probes
    Download

    Data Sheet - HC2A-S3A/HC2-S3H Meteorolgy Probes

    Download

Accessories

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more