` HP125 Series Multi-Sense Humidity Probe

Edgetech Instruments

HP125 Series Multi-Sense Humidity Probe

Cost effective, easy to use solution to many industrial & laboratory measurement & control applications

Starting At
$1,559.00 CAD
Normally ships in 3-4 Weeks

Configure   Request Quote

Can We Help You?

CALL US   EMAIL US   CHAT LIVE

Our product experts are here to help you find the right solution for your application!

The HP125 Multi-Sense Humidity Probe is a cost effective, easy to use solution to many industrial and laboratory measurement and control applications. The HP125 probe exploits the latest advancements in polymer humidity sensing technology and compact electronic circuitry. It is designed to continuously measure up to 4 measurement parameters using a single intelligent sensor tip including RH%, calculated Dew Point, Temperature and either Absolute Pressure or Barometric Pressure!

The Smart Sensor Tip carries all configuration information and calibration data and automatically downloads to the HP probe once connected! Easily replaced, the Smart Sensor Tip can be replaced in seconds. Upon installation into the probe, it gets recognized and automatically synchronizes with the main software within the HP125 probe. This means that you have the option of quick turn-around for annual calibrations! No need to take the probe out of service for weeks for shipment and calibration off-site. Simply replace the old Sensor Tip with a new one! Each new tip comes with a NIST traceable calibration certificate for your records.

The HP125 probe provides 0-10VDC analog voltage outputs plus an RS-232 serial interface. The RS-232 offers reporting functions only. Each probe is configured with a smart sensor tip. The sensor tip is protected by a sintered metal filter cartridge. Both the filter and the sensor tip are replaceable. The probe may be directly inserted into process measurement points or may be coupled with a variety of application oriented sampling accessories for very flexible installation opportunities.

The probe may be used Stand-Alone or in combination with the HP125-DCM (Display Control Module) or the SENTRY series of Compressed Gas Alarm Modules. The probe comes with wire leads (labeled) for respective analog output and power inputs (25VDC Power).

Options:
• LDP option: measure dew point down to -60C
• Probe Configurations:
    • HP125 : RH / Temperature / Calculated Dew Point
    • HPP125 : Pressure / RH / Temperature / Calculated Dew Point
    • HPB125 : Barometric Pressure / RH / Temperature / Cal. Dew Point
• Display Control Module:
    • Provides local display of measured parameters and operating status
    • (2) adjustable relay alarms
    • (2) configurable 4-20 mA outputs
    • Bi-directional RS-232 Serial Interface
    • Standard power: 24 VDC, optional VAC power adapter

• Replaceable Smart Sensor Tip
• Analog Voltage Output (0-10VDC) for each measured parameter
• RS-232 Serial Interface for reporting function
• Replaceable Sintered Filter Element protects the sensor
• 24VDC Power Input

PROBE OPTIONS:
HP125 Series: RH / Temperature / Calculated Dew Point
HPP125 Series: Pressure / RH / Temperature / Calculated Dew Point
HPB125 Series: Barometric Pressure / RH / Temperature / Calculated Dew Point
MEASURING RANGE:
Relative Humidity: 0-99.9% Non-condensing
Temperature: -40°C to +125°C (-40°F to 257°F)
Dew Point: Dew Point Range - Standard: -40°C to +95°C (-40°F to 203°F )
Dew Point Range - Optional: -60°C to +95°C (-76°F to 203°F)
Barometric Pressure: 600 to 1200mbar (HPB125 Series)
Pressure: 0 to 400psia (0 to 27.6bar) (HPP125 Series)
ACCURACY:
Relative Humidity: ±1.8% @ 23°C (0 TO 90% RH)
Temperature: ±0.5°C
Pressure: ±1 mbar @ 25°C (600 to 1200 mbar) / ±5mbar (-40°C to +125°C)
Dew Point: ±1°C for –10°C to 95°C DP and ±2°C for –60°C to –10°C
OPERATING RANGE:
Temperature: -40 °C to +125 °C (-40°F to 257°F )
Pressure: 800psig (55barg)
Environmental: Dust and Water Resistant
Media: Most Gases (Excellent Chemical Resistance)
GENERAL:
Speed of Response: 6 -12 sec (Flow dependent)
Life Expectancy: In excess of 5 years - Duty dependent
Storage: -20°C to +50°C (-4°F to 122°F)
Cable Length: 1m (3ft) Supplied as standard. Additional supplied on request.
Sensor Tip Filter: Sintered SS 100m. Easily removable for cleaning or field replacement
Wetted Material: 316 SS & Sensor components
Enclosure: 316 Stainless Steel
Supplied Pipe Fittings: ½” Male NPT (standard) or ¾” inch male NPT. Insertion length adjustable.
POWER & OUTPUTS:
Supply: 24V DC ±10% Unregulated, 50mA max
Connection: 22AWG wire leads, within single shielded cable
Analog Outputs: 0 to 10 VDC
Analog Output Cables: 1 Common, 4 Each + Signal
Serial Output: RS-232 serial output. Reporting function every 10s, space delimited

• Air Separators
• Atmospheric Sciences
• Automotive
• Compressed Air
• Food and Beverage
• Furnace
• Government Agencies
• HVAC
• Medical Gas
• Metrology
• Pharmaceuticals
• Plastics
• Power and Energy
• Semiconductor
• General Purpose
• R&D Test & Measurement
• Dynamometers
• Engine Test Cells



Your Recently Viewed Items...