Fully automated sensor calibration when used with Kaye validation software and traceable temperature standards
The LTR-40/140 temperature reference is specifically designed to fully automate sensor calibration when used with Kaye validation software and traceable temperature standards.
Temperature Range | -40°C to 140°C |
Ambient Operating Range | 5°C to 50°C |
Set Point Accuracy | 0.2°C |
Temperature Stability | 0.02°C |
Transfer Calibration Accuracy* | –40°C to –25°C: ±0.15°C |
IRTD Standard to Thermocouples | –25°C to 80°C: ±0.1°C 80°C to 130°C: ±0.15°C 130°C to 140°C: ±0.18°C |
Typical Heat-Up Time | 25°C to 80°C: 6 minutes 25°C to 140°C: 14 minutes |
Typical Cool-Down Time | 10 minutes: 25°C to –20°C |
Access Opening / Well configuration | Fixed Insert Reference wells (2): 6.7 mm diameter x 155 mm deep Calibration wells (6): 9 mm diameter x 155 mm) deep |
Display | LED w/0.01°C resolution |
Computer Interface | RS232 |
Computer Interface | RS232 |
Dimensions | 343 mm x 198 mm x 318 mm |
Weight | 13.6 kg |
Power | 115 VAC 60 Hz, 3 A or 230 VAC 50 Hz, 1.5 A 350 watts |
Kaye LTR-40/140 Temperature Reference - Manual 779.7 KB |