` PT-MON Explosion proof moisture analyzer

Michell Instruments

PT-MON Explosion proof moisture analyzer

A complete explosion proof moisture analyzer for critical gas process applications

Starting At
$26,430.00 CAD
Normally ships in 6-8 Weeks

Configure   Request Quote

Can We Help You?

CALL US   EMAIL US   CHAT LIVE

Our product experts are here to help you find the right solution for your application!

 Promet EExd Process Moisture Analyzers are hazardous-area rated industrial hygrometer systems for measurement of water vapor content in low and high pressure process gases on natural gas platforms and terminals as well as in petrochemical plants and industrial manufacturing facilities. The Promet EExd is installed directly in the hazardous area, simplifying site installation and retrofits. Promet EExd combines the latest Michell moisture sensing technologies with sample conditioning systems designed to provide a reliable and easy-to-operate, multi-channel, on-line instrument for moisture measurement in flammable gases.

• EExd flameproof certified - IECEx, ATEX, CCSAUS and TC TR Ex
• Complete hazardous area installation
• Single- or dual-channel measurement
• Moisture content, dew point and pressure
• Calibration traceable to NPL (UK) and NIST (USA)
• Remote global access options
• Sour gas compatible
• Field calibration maintenance
• Range -120 to +30°Cdp, 0.001 to 30,000 ppmV

Channel Confiurations Single and two channel
Moisture Parameters Dew point °C and °F
Moisture Content Automatic pressure compensated conversions: ppmV for natural gas and ideal gas, lbs/MMscf and mg/m3 for natural gas.
Dew point at a pressure input by the user for natural gas and ideal gas.
Calculations for natural gas moisture content based on either IS0 18453 or IGT#8 to customer order preference
Analysis Pressure barg, MPa and psig
Sensor Technology Michell Ceramic Moistures Sensor
Measuring Range -120°Cdp to +30°Cdp
1 ppbV to 30,000 ppmV
0 to 250 barg
Calibration Range -100°Cdp to +20°Cdp
10 ppbV to 23,000 ppmV
Accuracy Dew point: ±1°C between -59.9 & +20°Cdp
Moisture content: ±10% of reading
Dew point: ±2°C between -60 & -100°Cdp
Moisture content: ±20% of reading
Analysis Pressure: ±0.25% FS
Measured Resolution 0.1°C: -80 to +20°Cdp
1°C: -100 to -80°Cdp
Display Resolution Dew point: 0.1°C
Moisture content: autoscale, 5 digits
MPa and barg: 0.1, psig: 1.0
Resolution 0.1°C between +20 and -80°Cdp
1°C between -80 and -100°Cdp
Temperature Coefficient Algorithm compensation (-20 to +50°C)
HMI; Keyboard Interface Capacitive touch-screen through glass
Display Vacuum florescent
Data Loggings A rolling maximum of 150 data logs is available
Each log records time, date, moisture and pressure values for each channel
5 minutes minimum and 60 minutes maximum logging intervals can be set by the user
Communications and Output Two non-isolated 4-20 mA per measurement channel. 500 Ω maximum load. Range and parameter settable by user.
Modbus RTU @ 9600 baud-rate
Alarms: two volt free contacts per channel; one process value and one instrument status
Also available via Modbus communications
Main Unit Enclosure Internal temperature control for condensation protection and stable analysis conditions
Type Flameproof EExd
Construction Cast copper-free aluminium
Finish Chromate primer, Polyester P9010 powder coated (black). Meets BS3900
Operating Pressure Max 138 barg (2000 psig)
Sample Connections 1/8” NPT (ATEX / IECEx) or 1/4” NPT (CSA)(female)
Sample Flow Rate 0.5 to 5 I/min
Sample Flow Alarm Optional
Power Supply 90 to 260 V AC, 50/60 Hz, 180 W Main Unit
Operating Environment Indoor/Outdoor -20°C to +60°C
Max 95% RH Shaded location
Certification
Hazardous Area Certification ATEX: II 2G Ex d IIB + H2 Gb T5 (-40°C to +44°C) T4 (-40°C to +60°C)
IECEx: Ex d IIB + H2 Gb T5 (-40°C to +44°C) T4 (-40°C to +60°C)
cSAus: Class I, Division 1, Group B, C, D, T4 Tamb -40°C to +60°C
*TC TR: Ex 1Ex d IIB + H2 T4, T3 Gb (-40°C to +60°C)
Pattern Approval GOST-R, GOST-K


Your Recently Viewed Items...