Features remote sensor to directly measure the formation of condensation for drift-free readings of dew point & relative humidity
It offers a wide measurement range from the equivalent of <0.5 to 100% RH (dew point range: –40 to +120°C) at temperatures up to +120°C. The S8000 Remote features a new sensor design, incorporating a high precision 1/10DIN PT100 to provide ±0.1°C accuracy of dew-point measurement. Combined with the ±0.1°C accuracy temperature measurement sensor, the RH measurement accuracy of the S8000 Remote is better than other products currently available on the market.
• Fundamental, accurate and drift-free measurement
• Remote sensor
• Open design allows remote sensor to be mounted into a sample flow or simply placed in an environment to be monitored
• -40 to +120°C dew-point range with ±0.1°C accuracy
• Data logging via USB, Ethernet or SD card
• 'FAST' guarantees frost formation below 0°C
• Sensor operates in pressures up to 20 barg
Dew-Point Sensor Performance | |
Accuracy | ±0.1°C (±0.18°F) |
Reproducibility | ±0.05°C (±0.90°F) |
Measurement Technology | Chilled Mirror |
Sensor | |
Dew-Point Range 2-Stage | -40°Cdp @ sensor temp of +20°C to +90°Cdp @ sensor temp of +90°C |
Dew-Point Range High Temp PEEK | -40°Cdp @ sensor temp of +20°C to +120°Cdp @ sensor temp of +120°C |
Dew-Point Range Climatic Head | -10°Cdp @ sensor temp of +20°C to +120°Cdp @ sensor temp of +120°C |
Temperature Range 2-Stage | -40 to +90°C |
Temperature Range High Temp PEEK & Climatic Head | -40 to 120°C |
%RH Range 2-Stage & High Temp PEEK | <0.5 to 100% |
%RH Range Climatic Head | 10 to 100% |
Min Measured Dew Point @ 20°C 2-Stage & High Temp PEEK | -40°C |
Min Measured Dew Point @ 20°C Climatic Head | -10°C |
Mirror Material Options | Gold plated copper (standard), gold stud, Platinum stud |
Sensor Boby Material | Acetal (standard), high temperature PEEK, Anodized aluminum |
Response Speed | 1°C/sec plus settling time |
Operating Pressure | 20 barg (290 psig) standard High pressure version: 250 barg (3625 psig) max |
Remote PRT | |
Temperature Measurement | 4 wire Pt100, 1/10 DIN class B |
Accuracy | ±0.1°C (±0.18°F) |
Cable Length | 2m (6.6ft) (250m (820ft) max) |
Optional Remote Pressure Sensor | |
Measurement Range | 0 to 25 bara (0 to 377 psia) |
Accuracy | 0.25% Full Scale |
Measurement Units | psia, bara, KPa or MPag |
Pressure Transducer Thread | 1/8” NPT |
Monitor | |
Resolution | User-selectable to 0.001 dependant on parameter |
Measurement Units | °C and °F for dew point and temperature %RH, g/m3, g/kg, ppmV , ppmW (SF6), for calculated humidities |
Outputs | Analog 3 channels, user selectable 4-20 mA, 0-20 mA or 0-1 V Digital Modbus RTU over USB and optionally Modbus RTU over RS232 or RS485, or Modbus TCP over ethernet Alarm Two volt free changeover contacts, one process alarm, one fault alarm; 1 A @ 30 V DC |
HMI | 5.7" LCD with touchscreen |
Data Logging | SD Card (512Mb supplied) and USB interface SD Card (FAT-16) - 2Gb max. that allows 24million logs or 560 days, logging at 2 second intervals |
Environmental Conditions | -20 to +50°C (-4 to +122°F) |
Power Supply | 85 to 264 VAC, 47/63 Hz |
Power Consumption | 100 VA |
Mechanical Specifications | |
Dimensions (Instrument) | 186.9 x 215 x 215mm (7.36 x 10.12 x 8.4”) h x w x d |
Dimensions (Sensor) | ø45 x 128mm with M36 x 1.5-6g mounting thread |
Weight | 4.2kg (9.26lbs) |
Cable Lengths | 3, 5 or 10m (9.8, 16.4 or 32.8ft) |
General | |
Storage Temperature | -40 to +60°C (-40 to +140°F) |
Detection System | Single optics detection system with auto adjustment |
Calibration | 4-point traceable in-house calibration as standard; UKAS accredited calibrations optional – please consult Michell Instruments |
• Engine test cell monitoring – from commercial vehicle to high performance engines
View All Part Numbers