The SAH is designed for applications where a more rugged sensor and cable combination is required. The completely waterproof design easily handles the harsh, repeated steam/vacuum cycling of the autoclave process and incorporates an EDPM protective jacket to minimize sensor/cable separation and failure during rough handling or accidental stretching/extension. The continuous one-piece, low-profile design with a “ruggedized” shock-resistant sensor is an excellent choice and provides an extra measure of protection when used in larger “walk-in” chambers where the movement of carts or repeated handling and accidental “abuse” may occur.

Nominal Ice Point Resistance (RO)100.00 Ω
RO InterchangeabilityRO ± 0.10 Ω
Typical Alpha Value0.0038500 Ω/Ω/°C
Temperature Rating-50°C to 200°C
Element ConfigurationSingle or dual
Long-Term Stability± 0.26°C (0.10 Ω) maximum shift at 0°C after 1000 hours at 200°C
Long-Term Repeatability± 0.10°C (0.04 Ω) maximum shift at 0°C after 10 cycles between -50°C and 200°C
Short-Term Repeatability and Hysteresis± 0.025°C (0.01 Ω) maximum change at 0°C over any 5 consecutive thermal cycles from 0°C to 135°C
Rated Pressure1 psia to 35 psia
Transition Fitting and Cable Limits-50°C to 200°C continuous exposure
Insulation Resistance100 MΩ minimum at 100 VDC at room termperature
  • Thumbnail of document Data Sheet - SAH Sanitary Autoclave RTD
    Download

    Data Sheet - SAH Sanitary Autoclave RTD

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more