Corrosion resistant PPS body, accurate measurement of flow rates with zero drift, 0 - 5 VDC output, no required maintenance
The Series TFP-GV Gas Flowmeters utilize a turbine wheel and electro-optical detection to convert flow rates into a linear 0 - 5 VDC output signal for recording and data logging. A power adapter or mating cable assembly is required for operation.
• Corrosion resistant PPS body
• High repeatability with patented microturbine design
• Design accurately measures flow rates with no zero drift and no required maintenance
Service | Clean dry gases compatible with wetted materials |
Wetted Materials | PPS, acetal, sapphire, glass, epoxy, and fluoroelastomer |
Accuracy | ± 3% of FS |
Linearity | ± 3% of FS |
Repeatability | ± 0.5% of FS |
Temperature Limits | 41°F - 131°F (5°C - 55°C); Storage: 32°F - 158°F (0°C - 70°C); Sensitivity: ± 0.2% of FS per °C |
Pressure Limits | 40 PSIG (2.8 Bar) |
Process Connection | Compression fitting, see model table on Catalog page |
Power Requirements | 11.5 VDC - 15 VDC |
Power Consumption | 35 mA @ 12 VDC |
Output Signal | 0 VDC - 5 VDC; minimum 2.5 kΩ load |
Electrical Connections | Four-pin power and signal connector. A power adapter or mating cable required for operation. See accessories table |
Enclosure Rating | IP10 (NEMA 1) |
Weight | 0.16 lb (75 g) |
Agency Approvals | CE |
• Industrial flow monitoring
• Commercial systems
• Laboratory equipment
View All Part Numbers