Two input channels so you can collect data with two reference sensors, or you can do comparison calibrations of one sensor against a calibrated reference sensor

Buy now!
$6,083.00CAD
Have questions? We're here to help!
Have questions? We're here to help!

The SPRT module reads 25 Ω and 100 Ω four-wire RTDs, PRTs, and SPRTs with very high accuracy. It turns the Stack into a first-rate reference thermometer with an accuracy to +/- 0.005°C.

It has two input channels so you can collect data with two reference sensors, or you can do comparison calibrations of one sensor against a calibrated reference sensor.

Temperature conversion features include direct resistance measurement, ITS-90, W(T90), IPTS-68, Callendar-Van Dusen, or an RTD polynomial conversion. The user-changeable default values for the CVD conversion fit the 100-ohm, 0.00385 ALPHA sensor described by IEC-751.

The SPRT modules can be used one at a time or combined together in any combination for reading up to 16 different reference thermometers. If you stack an SPRT module with a scanner module, you can test multiple sensors against your reference. Unlike other competitive instruments, our PRT Scanner Module operates with or without the two-channel SPRT module. If you can think of a way to use a reference thermometer, you can do it with the Stack.

The 1560 Black Stack is an incredible thermometer. You buy only the modules you need for the work you are doing. If your work changes, simply order the modules with the functions you need and slip them onto the back of the Stack. Your thermometer changes its software, display, and method of operation to match the new functions you’ve added.

Remember, you never have to open the case to add modules. There’s no software to load. It’s all automatic.

Each module stores its own calibration internally, so you can add or change modules without recalibrating the whole stack. Module calibration is digital and is performed manually through the base’s front panel or over the RS-232 link. If your lab has the capability, you can calibrate modules yourself. If not, send them to us with or without your base unit and we’ll recalibrate them.

The LCD screen has multiple methods of displaying data, including a graphical strip chart recorder. The graphical capability of the Black Stack makes testing temperature stability easier than ever. Vertical scaling and graph resolution are automatic.

The Stack has high-accuracy, two-channel capability or multi-channel functionality if you need it. Its memory stores the most recent 1000 readings, or you can send your data to your PC through the RS-232 port. Each data point is time and date stamped. An IEEE-488 port is optional.

With the Black Stack you can read data almost anyway you like—in ohms, millivolts, or temperature, according to your application and preference.

Remember, this thermometer’s calibration is traceable to NIST. Its accuracy is as high as ±0.0013°C, depending on the module and sensor you’re using.
 

Input Channels2
Resistance Range0 Ω - 400 Ω
Basic Resistance Accuracy +/- 20 ppm of reading (0.0005 Ω at 25 Ω, 0.002 Ω at 100 Ω)
Resistance Resolution 0.0001 Ω
Temperature Range–260°C - 962°C
Equivalent Temperature Accuracy+/- 0.005°C at 0°C
+/ -0.007°C at 100°C
Temperature Resolution 0.0001°C
Excitation Current 1.0 mA,
1.4 mA
  • Thumbnail of document Data Sheet - 2560 Standard Platinum Resistance Thermometer Module
    Download

    Data Sheet - 2560 Standard Platinum Resistance Thermometer Module