Skip to main content
Press Enter
Starting at $5,642.00CAD
Configure
Fluke Calibration

5626 Secondary Standard and High-Temp Platinum Resistance Thermometers

High-temperature secondary standards fill the gap between afforadable but temperature-limited PRTs & highly accurate SPRTs

Starting at
$5,642.00CAD
Have questions? We're here to help!
Have questions? We're here to help!
Configure

Fluke Calibration's high-temp secondary standards fill the gap between affordable, but temperature-limited secondary PRTs (platinum resistance thermometers) and more expensive, highly accurate SPRTs (standard platinum resistance thermometers).

If you’re using block calibrators, furnaces, or temperature points above normal PRT (platinum resistance thermometer) temperatures (420°C), then this PRT is for you. The 5626 is nominally 100 Ω. The instrument has a temperature range of -200°C - 661°C. They make great working or check standards for calibration work up to the aluminum point.

Using a regular PRT at temperatures above 500°C exposes the platinum to contamination. If the PRT is used as a reference or calibration standard, contamination is a major problem. SPRTs (standard platinum resistance thermometers), which are more expensive and delicate, can handle the higher temperatures, but with greater risk to the instrument due to shock, contamination, or mishandling. The 5626 is designed to reduce the contamination risk through the use of internal protection while not impairing performance.

In addition to the right measurement performance and durability, a PRT for secondary applications should be priced affordably. Fluke Calibration’s new PRTs are inexpensive and come with an accredited calibration. The calibration comes complete with ITS-90 constants and a resistance-versus-temperature table.

• Range to 661°C
• Meets all ITS-90 requirements for resistance ratios
• Rtp drift < 20 mK after 500 hours at 661°C
• Calibrated accuracy of +/- 0.006°C at 0°C

Temperature Range-200°C - 661°C
Handle Temperature0°C - 80°C
RTPW 5626: 100 Ω (+/- 1 Ω)
5628: 25.5 Ω (+/- 0.5 Ω)
W(Ga)31.11807
Calibrated Accuracy (k = 2) +/- 0.006°C at -200°C
+/- 0.006°C at 0°C
+/- 0.015°C at 420°C
+/- 0.022°C at 661°C
Stability 5626: +/- 0.003°C
5628: +/- 0.002°C
Long-Term Drift (k = 2) 5626: < 0.006°C/100 h at 661°C
5628: < 0.004°C/100 h at 661°C
Immersion At least 12.7 cm (5 in) recommended
Sheath Inconel™ 600
Lead Wires 4-wire Super-Flex PVC, 22 AGW
Termination Gold-plated spade lugs, or specify
Size 6.35 mm dia. x 305 mm, 381 mm, or 508 mm (0.25 in x 12 in / 15 in / 20 in) standard
Calibration Accredited calibrations from Fluke Calibration
Includes calibration and 100 h drift
  • Thumbnail of document Manual - 5626 Secondary Standard and High-Temp Platinum Resistance Thermometers
    Download

    Manual - 5626 Secondary Standard and High-Temp Platinum Resistance Thermometers

    Download

Accessories

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more