Designed and manufactured by our primary standards metrologists, the strain-free sensing element in the 5699 meets all ITS-90 requirements for SPRTs and minimizes long-term drift.

After one year of regular usage, drift is less than 0.008°C (< 0.003°C is typical). Even lower drift rates are possible depending on care and handling. A fifth wire for grounding is added to the four-wire sensor to help reduce electrical noise, particularly for AC measurements. Finally, you can get an improved version of an old industry-standard Inconel-sheathed SPRT.

The 5699 is constructed with a 0.219 in diameter Inconel sheath for high durability and fast response times. Inside the sheath, the sensing element is protected by a thin platinum housing that shields the sensor from contamination from free-floating metal ions found within metal environments at high temperatures. Reduced contamination means a low drift rate—even after hours of use in metal-block furnaces at high temperatures.

Affordable working standard SPRT

• Range to aluminum point (660°C )
• Inconel™ and platinum sheaths guard against contamination
• Less than 8 mK/year drift
• Many calibration options available including by fixed point
• Fifth wire provides shield ground

Temperature Range-200°C - 670°C
Nominal RTPW25.5 Ω (+/- 0.5 Ω)
Current1 mA
Resistance RatioW (302.9146 K) ≥ 1.11807
W (234.3156 K) ≤ 0.844235
Sensitivity0.1 Ω/°C
Drift Rate < 0.008°C/year
(< 0.003°C/year typical)
Repeatability< 1 mK
Self-Heating at TPW< 0.001°C under 1 mA current
Reproducibility+/- 0.001°C or better
RTPW Drift After Thermal Cycling< 0.001°C
Diameter of Pt Sensor Wire0.07 mm (0.003 in)
Lead WiresFour sensor wires plus grounding wire
Protective SheathInconel
Diameter5.56 mm +/- 0.13 mm (0.219 in +/- 0.005 in)
Length482 mm (19 in)
Insulation Resistance> 100 MΩ at 661°C
> 1000 MΩ at 20°C
  • Thumbnail of document Manual - 5699 High-Temperature Metal-Sheath Standard Platinum Resistance Thermometer
    Download

    Manual - 5699 High-Temperature Metal-Sheath Standard Platinum Resistance Thermometer

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more