Skip to main content
Press Enter
Starting at $7,649.59CAD
Configure
Fluke

718Ex Intrinsically Safe Pressure Calibrator

Self-contained pressure calibrator provides a total pressure calibration solution for transmitters, gauges, & switches

Starting at
$7,649.59CAD
Have questions? We're here to help!
Have questions? We're here to help!
Configure
Pump Up the Pressure
 
The Fluke 718Ex Intrinsically Safe Pressure Calibrator is a powerful intrinsically safe, self contained pressure calibrator for usage in explosion endangered areas.

• ATEX II 1G Ex ia IIC T4 Compliant.
• Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
• 30, 100, and 300 psi ranges (2, 7, and 20 bar)
• Pressure measurement to 0.05% of full span, using an internal pressure sensor.
• Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
• Wide range of selectable measurement units for pressure.
• Current measurement with 0.02% accuracy and 0.001 mA resolution.
• Pressure switch test function.
• Min-Max Hold functions.
• Compact size and weight.
• Simple, push-button user interface.
• Rugged and reliable, ideal for field use.
 


Measurement Accuracy

 

718Ex 30G

718Ex 100G

718Ex 300G

Pressure

Range

-830 mbar - 8 bar (internal sensor)

-830 mbar - 7 bar (internal sensor)

-830 mbar - 20 bar (internal sensor)

Over Pressure

3x FS

2x FS

375 PSI, 25 bar

Resolution

0.0001 bar

0.001 bar

0.0001 bar

Accuracy

0.05% FS

0.05% FS

0.05% FS

Functions

Zero, Min, Max, Hold, Damp*

Zero, Min, Max, Hold, Damp*

Zero, Min, Max, Hold, Damp*

Source Pressure via Built-In Pump

Range

-820 mbar to full scale

Supported Pressure Units

PSI, in. H2O (4 °C), in. H2O (20 °C), kPa, cm H2O (4 °C), cm H2O (20 °C), bar, mbar, kg/cm², mmHg, in Hg

Technical Data

Measure Accuracy

Range

0 mA - 24 mA

Resolution

0.001 mA

Accuracy

0.02% RDG +/- 2 counts

Display

LCD, 5 digit pressure and current simultaneous

Environmental Specifications

Temperature

Operating

-10°C - 55°C

 

Storage

-40°C - 60°C

Humidity (Without Condensation)

95% (10°C - 30°C)

75% (30°C - 40°C)

45% (40°C - 50°C)

  • Thumbnail of document Data Sheet - 718Ex Intrinsically Safe Pressure Calibrator
    Download

    Data Sheet - 718Ex Intrinsically Safe Pressure Calibrator

    Download
  • Thumbnail of document Manual - 718Ex Intrinsically Safe Pressure Calibrator
    Download

    Manual - 718Ex Intrinsically Safe Pressure Calibrator

    Download

Accessories

Pressure Calibrators

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more