Since we introduced the world's first truly handheld dry-well, many have tried to duplicate it—in vain. Despite its small size (2¼ inches high and 5 inches wide) and light weight, the 9100S outperforms every dry-well in its class in the world.
 
It's simple and convenient, too. Anyone can learn to use one in less than 15 minutes. It has a range to 375 °C (707 °F) and is perfect for checking RTDs, thermocouples, and small bimetal thermometers in the field.
 
Plug it in, switch it on, set the temperature with the front-panel buttons, and insert your probe into the properly sized well. Compare the reading of your device to the display temperature or to an external reference, and the difference is the error in your device. With a proprietary Fluke Calibration temperature controller, the 9100S has a display resolution of 0.1 degrees, an accuracy of ±0.5 °C, a stability of ±0.1 °C, and a stabilization time of just five minutes.
 
9102S Dry-Well
 
For work in the temperature range of –10°C to 122°C, 9102S dry-well is another first in the industry, featuring stability to ±0.05°C.
 
This dry-well is only four inches high and six inches wide, achieves temperatures as low as –10°C, includes a NIST-traceable calibration, and is stable to ±0.05°C. The Model 9102S is excellent for dial gauges, digital thermometers, bulb switches, and other sensors that need calibration below ambient.
 
The 9102S has two wells so you can use one for a reference thermometer to increase accuracy. Both wells are 1/2 inch in diameter, and each has inserts available for almost any sensor size. The 9102S also has a battery pack option that gives you approximately four hours of field use when AC power is unavailable.
 

 

9100S:
• Smallest dry-wells in the world
• Ranges from 35 °C to 375 °C
• Accuracy to ±0.25 °C, stability of ±0.25 °C at 50 °C
• RS-232 interface
9102S:
• Smallest dry-wells in the world
• Ranges from –10°C to 122°C
• Accuracy to ±0.25°C, stability of ±0.05°C


 

Range9100S: 35°C - 375°C (95°F - 707°F)
9102S: -10°C - 122°C (14°F - 252°F) at 23°C ambient
Accuracy9100S: +/- 0.25°C at 50°C; +/- 0.25°C at 100°C; +/- 0.5°C at 375°C
9102S: +/- 0.25°C
Stability9100S: +/- 0.07°C at 50°C; +/- 0.1°C at 100°C; +/- 0.3°C at 375°C
9102S: +/- 0.05°C
Well-to-Well Uniformity+/- 0.2°C with sensors of similar size at equal depths within wells
Heating Times9100S: 35°C - 375°C: 9.5 min
9102S: ambient to 100°C: 10 min
Stabilization9100S: 5 min
9102S: 7 min
Cooling Times9100S: 375°C - 100°C: 14 min
9102S: ambient to 0°C: 10 min
Well Depth9100S: 102 mm (4 in);1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep
9102S: 102 mm (4 in)
Removable Inserts9100S: N/A
9102S: Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included]
Power9100S: 115 VAC (+/- 10%), 55 - 65 Hz, 1.5 A or 230 VAC (+/- 10%), 0.8 A, 45 - 55 Hz, 175 W
9102S: 94 - 234 VAC (+/- 10%), 50/60 Hz, 60 W; or 12 VDC
Size (H x W x D)9100S: 57 mm x 125 mm x 150 mm (2.25 in x 4.9 in x 6.1 in)
9102S: 99 mm x 140 mm x 175 mm (3.9 in x 5.5 in x 6.9 in)
Weight9100S: 1 kg (2 lb 3 oz)
9102S: 1.8 kg (4 lb)
Computer InterfaceRS-232 included with free Interface-it software
NIST-Traceable Calibration9100S: Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 375°C
9102S: Data at -10°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 122°C

Specifications

9100S

9102S

Range

35 °C to 375 °C (95 °F to 707 °F

–10 °C to 122 °C (14 °F to 252 °F) at23 °C ambient

Accuracy

± 0.25 °C at 50 °C; ± 0.25 °C at 100 °C; ± 0.5 °C at 375 °C

± 0.25 °C

Stability

± 0.07 °C at 50 °C; ± 0.1 °C at 100 °C; ± 0.3 °C at 375 °C

± 0.05 °C

Well-to-Well Uniformity

± 0.2 °C with sensors of similar size at equal depths within wells

Heating Times

35 °C to 375 °C: 9.5 minutes

ambient to 100 °C: 10 minutes

Stabilization

5 minutes

7 minutes

Cooling Times

375 °C to 100 °C: 14 minutes

ambient to 0 °C: 10 minutes

Well Depth

102 mm (4 in);1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep

102 mm (4 in)

Removable Inserts;

N/A

Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included]

Power

115 V ac (± 10 %), 55–65 Hz, 1.5 A or 230 V ac (± 10 %), 0.8 A, 45–55 Hz, 175 W

94-234 V ac (± 10 %), 50/60 Hz, 60 W;or 12 VDC
 

Size (HxWxD)

57 x 125 x 150 mm (2.25 x 4.9 x 6.1 in)

99 x 140 x 175 mm (3.9 x 5.5 x 6.9 in)

Weight

1 kg (2 lb 3 oz)

1.8 kg (4 lb)

Computer Interface

RS-232 included with free Interface-it software

NIST-Traceable Calibration

Data at 50 °C, 100 °C, 150 °C, 200 °C,250 °C, 300 °C, and 375 °C

Data at –10 °C, 0 °C, 25 °C, 50 °C, 75 °C,100 °C, and 122 °C

Specifications

9100S

9102S

Range

35 °C to 375 °C (95 °F to 707 °F

–10 °C to 122 °C (14 °F to 252 °F) at23 °C ambient

Accuracy

± 0.25 °C at 50 °C; ± 0.25 °C at 100 °C; ± 0.5 °C at 375 °C

± 0.25 °C

Stability

± 0.07 °C at 50 °C; ± 0.1 °C at 100 °C; ± 0.3 °C at 375 °C

± 0.05 °C

Well-to-Well Uniformity

± 0.2 °C with sensors of similar size at equal depths within wells

Heating Times

35 °C to 375 °C: 9.5 minutes

ambient to 100 °C: 10 minutes

Stabilization

5 minutes

7 minutes

  • Thumbnail of document Manual - 9100S/9102S Handheld Dry-Wells
    Download

    Manual - 9100S/9102S Handheld Dry-Wells

    Download
  • Thumbnail of document Manual - 9100S/9102S Handheld Dry-Wells
    Download

    Manual - 9100S/9102S Handheld Dry-Wells

    Download

Accessories

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more