Skip to main content
Press Enter
Starting at $31,626.00CAD
Configure
Fluke Calibration

9115A / 9116A Freeze-Point Furnaces

When combined with freeze-point cells, produce the longest plateaus in the industry

Starting at
$31,626.00CAD
Have questions? We're here to help!
Have questions? We're here to help!
Configure
9114
This furnace has a range of 100 °C to 680 °C, which includes the indium, tin, zinc, and aluminum fixed points, all in one furnace.
 
The 9114 furnace has an inlet for use of clean dry air or inert gas to initiate the supercool of a tin cell. Other furnaces require the user to remove the hot and fragile tin cell from the furnace by hand before cooling. In a Fluke Calibration furnace, you simply turn on your gas, monitor your cell during its supercool, and turn the gas off when the freeze begins.
 
The 9114 is a three-zone furnace with the best in digital controller technology. Fluke Calibration designs and builds proprietary controllers that have a reputation of being the best in the business. All of our fixed-point furnaces use them to achieve excellent stability and uniformity.
 
For easy access and visibility, all three zones are controlled from the front of the unit. The primary controller can be set in 0.01 °C increments, and actual temperature is readable to two decimal places.
 
The freezing and melting process can be automated using eight preset, user- programmable temperature settings. The top and bottom zones are slaved to the primary zone using differential thermocouples. A high-temperature PRT acts as the main control sensor for the best accuracy, sensitivity, and repeatability.
 
9115A
The 9115A Sodium Heat Pipe Furnace is specifically designed for maintenance of aluminum and silver freeze-point cells.
 
It has a temperature range of 550 °C to 1000 °C with gradients of less than ± 0.1 °C throughout. The sodium heat-pipe design provides a simple, yet uniform, single heating zone that ensures very uniform changes in states during heating and cooling.
 
Melting, freeze initiation, and plateau control for a variety of freeze-point cells are possible by entering up to eight set-points, ramp rates, and soak times. The controller displays temperature in degrees C or F, and temperature feedback is done via a thermocouple. Freeze-point plateaus of 8 to 10 hours are typical, and 24 hours are possible under controlled conditions.
 
External cooling coils are included for circulation of tap water to reduce chassis temperature and heat load to the lab. Temperature cutouts protect your SPRTs and the furnace from exposure to excessive temperatures.
 
9116A
The 9116A Furnace has a temperature range of 550 °C to 1100 °C and is designed for use in achieving aluminum, silver, or copper freezing point measurements. An advanced high temperature Sodium heat pipe extends usage to more than 1000 hours at 1100 °C and 5000 hours at 982 °C. The heater is embedded in a fiber ceramic insulating block. A hollow section through the center contains the heat pipe.
 
The minimum working temperature of the Sodium heat pipe is about 500 °C. Above that temperature, the sodium circulates throughout the tube providing a uniform temperature zone for freezing point measurements. With uniformity of ± 0.05 °C, zone adjustments are eliminated simplifying installation and increasing throughput.
 
The uniform temperature is maintained over the full length of the metal freeze point cell. A programmable temperature controller simplifies freeze initiation, melting and plateau control. Control stability is ± 0.15 °C, the best in the industry, enabling extension of freezing plateaus of quality fixed point cells for up to 20 hours and longer. For compatibility with automation programs, plateaus may be controlled through standard RS-232 and optional IEEE-488 PC interfaces.

• Designed to extend plateaus
• External cooling coils for circulation of tap water to reduce heat load to the lab
• Each furnace comes with RS-232 ports
• Equilibration blocks are available for comparison calibrations
• High-stability OEM controllers, RS-232 included
• IEEE-488 interface packages are available

Temperature Range9114: 100°C - 680°C
9115A: 550°C - 1000°C
9116A: 550°C to 1100°C
Temperature Stability9114: +/- 0.03°C
9115A: +/- 0.25°C
9116A: +/- 0.5°C
Temperature Uniformity9114: +/- 0.05°C (+/- 0.1°C in the pre-heat well)
9115A: +/- 0.1°C
9116A: +/- 0.05°C
Set-Point Accuracy9114: +/- 0.5 °C
9115A: +/- 3.0°C
9116A: +/- 3.0°C
Set-Point Resolution9114: 0.01°C
9115A: 0.1°C
9116A: 0.1°C
Display Resolution9114: 0.01 °C
9115A: 0.1°C below 1000°C; 1°C above 1000°C
9116A: 0.1°C below 1000°C; 1°C above 1000°C
Thermal Safety Cutout Accuracy9114: +/- 5°C
9115A: +/- 10°C
9116A: +/- 10°C
Heater Power9114: End Zones: 1000 W each (at 230 VAC nominal)
Primary Zone: 1500 W
9115A: 2500 W
9116A: 2500 W
Exterior Dimensions (H x W x D)838 mm x 610 mm x 406 mm (33 in x 24 in x 16 in)
Power Requirements230 VAC (+/- 10 %), 50/60 Hz, 1 Phase, 22 A maximum
Weight9114: 92 kg (203 lb.)
9115A: 82 kg (180 lb.)
9116A: 82 kg (180 lb.)
  • Thumbnail of document Manual - 9115A / 9116A Freeze-Point Furnaces
    Download

    Manual - 9115A / 9116A Freeze-Point Furnaces

    Download
  • Thumbnail of document - 9115A / 9116A Freeze-Point Furnaces
    Download

    - 9115A / 9116A Freeze-Point Furnaces

    Download
  • Thumbnail of document Manual - 9115A / 9116A Freeze-Point Furnaces
    Download

    Manual - 9115A / 9116A Freeze-Point Furnaces

    Download
  • Thumbnail of document Manual - 9115A / 9116A Freeze-Point Furnaces
    Download