Skip to main content
Press Enter
OMEGA

EDAW-N1E1-02T1-HV Electronic Pressure Controller

Electronic pressure switch, range 0-20 psi, 1/4" male NPT, bottom connection, 316L SS, 2 SPDT relays, 4-20 mA output, 120/240 VAC voltage option

Normally ships in
2-3 weeks
Have questions? We're here to help!
Have questions? We're here to help!

Supported Applications

The Dwyer Series EDA Electronic Pressure Controller integrates a pressure gage, two switches, and an optional transmitter into a single unit. Designed for indoor and outdoor use with weatherproof housing (IP65), the series is suitable for process control, compressor control, filter status monitoring, duct or building static pressure, and damper and fan control.

Operating Conditions & Performance

The Series EDA measures compatible liquids and gases across ranges from 0-20 psig to 0-3000 psig. Accuracy is specified as ±1% of FS including linearity, hysteresis, and repeatability for both the indicator and transmitter. Stability is rated at < ±2% of FS per year.

Operating conditions include an ambient temperature range of 20 to 140°F (-6.6 to 60°C) and a process temperature limit of 0 to 176°F (-18 to 80°C). The compensated temperature limits are 32 to 122°F (0 to 50°C), with a thermal effect of ±0.05% of FS/°F. Humidity ratings support 0 to 95% RH up to 104°F (40°C) non-condensing, and 10 to 50% at 140°F (60°C) non-condensing.

Pressure limits are defined as 1.5 x range for ranges up to 6,000 psi; the document also notes an 8,000 psi range with a 10,000 psi limit in specific contexts. The display updates every 600 ms when damping is set to 1.

Configuration Options

The series offers configurable options for housing style, process connections, conduit entries, measurement ranges, and output signals.

• Housing: Weatherproof (W) with a polycarbonate or glass-filled plastic enclosure meeting NEMA 4X / IP65 ratings. Note: The IP65 rating is not evaluated by UL.
• Process Connections: Available in 1/4" male NPT, 1/4" BSPT male, or 7/16" SAE configurations.
• Electrical Entry: Removable terminal blocks with two 1/2" female NPS conduit connections (E1).
• Ranges: Configurable from 0-20 psig through 0-3000 psig.
• Transmitter Outputs: Options include None, 4-20 mA, 1-5 VDC, 0-5 VDC, 1-6 VDC, and 0-10 VDC (direct or reverse output selection). Minimum excitation for transmitter operation is 14 VDC.
• Switching: Two SPDT relays with an electrical rating of 5A @ 120/240 VAC, 1A @ 30 VDC. Set points are adjustable from 0-100% of FS with manual or automatic reset.
• Power Supply: Standard models operate on 12-28 VDC/VAC (50/60 Hz), capable of functioning at 8 VDC for 45 seconds. The T5 option requires 14-30 VDC/VAC, and the -HV option supports 120-240 VAC.
• Wetted Materials: 316 L SS.

Key Product Differences

Variants within the Series EDA differ primarily by measurement range, transmitter output type, and power supply configuration. The model number structure follows the format: EDA - [HOUSING] - [PROCESS CONNECTION] - [CONDUIT CONNECTION] - [RANGE] - [TRANSMITTER OUTPUT] - [OPTIONS].

Housing variations include standard weatherproof enclosures (W). Process connection codes distinguish between 1/4" male NPT (-N1), BSPT, and SAE fittings. Range codes correspond to specific pressure spans: for example, -02 denotes 0-20 psig, while -10 indicates 0-3000 psig.

Transmitter output options are designated by T-codes: T0 (None), T1 (4-20 mA), T2 (1-5 VDC), T3 (0-5 VDC), T4 (1-6 VDC), and T5 (0-10 VDC). Additional option codes include -AT for an aluminum adhesive tag, -HV for 120-240 VAC power, and -NIST for a NIST certificate. Accessories such as the A-EDA-BRK flush mount bracket are available to support surface mounting.

  • Display Units: psi, kg/cm², bar, in Hg, ft w.c., kPa, MPa, %FS, in w.c., mbar, cm w.c., mm Hg, or oz/in²
  • Test Mode: Simulates input over the range without pressuring to easily test switches and transmitter 
  • Fail-safe: For sensor failure, over pressure, high temperature limit, low temperature limit, or keypad
  • Alternation: Selectable alternation of set points between the relays for even wear on duplex pump applications.
Accuracy± 1 %
Accuracy Detail± 1% of FS including linearity, hysteresis, and repeatability (indicator and transmitter)
Compensated Temperature, Max122 °F
Compensated Temperature, Min32 °F
Display Note4-digit backlit LCD (digits: 0.60˝H x 0.33˝ W)
Display TypeLCD
Enclosure RatingsIP66, NEMA Type 4X
MaterialsPolycarbonate
Number of Digits4
Operating Temperature, Max140 °F
Operating Temperature, Min20 °F
Measurement Pressure, Min0 psi
Measurement Pressure TypeGauge
Process Connection Size1/4 in
Process Connection TypeNPT Male
Process Temperature, Max176 °F
Process Temperature, Min0 °F
Repeatability± 1%
Repeatability Note± 1% of FS (switching only)
Requires Factory Media Reviewtrue
Series IDSeries-EDA
Set Point AdjustmentFull Scale Adjustable (0-100%)
Stability<± 2% of FS/year
Switch IndicationExternal LED for each relay on the front panel
Switch Rating5 A @ 120/240Vac, 1 A @ 30 Vdc
Switch ResetManual/Automatic
Switch Type(2) SPDT
Thermal Effects± 0.05% of FS/°F
Warm Up Time<10 s
Weight1.18 lb
Zero Setting Adjustable RangeMenu Scalable within the Range
• Process control
• Compressor control
• Filter status
• Duct or building static pressure
• Damper and fan control
  • Download

    Data Sheet - EDAW-N1E1-02T1-HV Electronic Pressure Controller

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more