Supported Applications
The Omega iServer MicroServer™ Barometric Pressure, Temperature, and Humidity Transmitters (iBTX and iBTHX series) are Ethernet-connected transmitters designed for monitoring barometric pressure, temperature, relative humidity, and dew point. The system functions as a web server to serve active web pages displaying real-time readings, charts, and log data in standard formats compatible with spreadsheets or data acquisition programs.
Documented applications include:
• Clean rooms
• Computer rooms
• HVAC systems
• Pharmaceuticals
• Laboratories
Operating Conditions & Performance
The series supports measurement ranges of 10 to 1100 mbar (1 to 110 kPa) for barometric pressure, 0 to 100% RH for relative humidity, and -40 to 85°C (-40 to 185°F) for temperature using the industrial probe. The wand probe supports a process range of 5 to 45°C (41 to 113°F).
Performance specifications vary by sensor type and range:
• Barometric Pressure Accuracy: ±3.5 mbar at 25°C
• Relative Humidity Accuracy: ±2.75% (10 to 90%), ±3% (5 to 10% and 90 to 95%), ±4% (0 to 5% and 95 to 100%)
• Temperature Accuracy: ±0.5°C for wand probe (5 to 45°C), ±1°C for wand probe (0 to 5°C and 45 to 70°C); ±0.5°C for industrial probe (5 to 45°C), ±1.5°C for industrial probe (-40 to 5°C and 45 to 85°C)
• Resolution: Barometric Pressure: 0.1 mbar; Relative Humidity: 0.1%, 12 bit; Temperature: 0.1°C, 14-bit
• Hysteresis: ±1% RH
• Repeatability: ±0.1% for humidity, ±0.1°C for temperature
• Response Time (tau 63%): 8 seconds for humidity; 5 to 30 seconds for temperature
Ambient operating temperatures are 0 to 70°C (32 to 158°F) for iBT(H)X-W and iBT(H)X-D models, and 0 to 60°C (32 to 140°F) for the iBTX-SD model. Storage temperature is -40 to 85°C (-40 to 185°F).
Configuration Options
The series offers configurable mounting, probe types, and data logging options tailored to specific installation needs.
Housing & Mounting Options
• DIN Rail Mounted Unit (-D): Polycarbonate case with DIN rail mount (iBTX/iBTHX-D). Dimensions: 115 mm (4.53 in) W x 90.2 mm (3.54 in) H x 25.1 mm (0.99 in) D.
• Bench/Wall Mount Unit (-W): Valox® 364 PBT case with wall mount bracket (iBTX/iBTHX-W). Dimensions: 85.9 mm (3.38 in) W x 232.3 mm (9.15 in) H x 66.0 mm (2.60 in) D.
• Bench/Wall Mount with LCD (-SD): Steel metal case with wall bracket mount, includes LCD Display, Flash Memory Card, and Backup Battery (iBTX-SD). Dimensions: 82.6 mm (3.25 in) W x 232.3 mm (9.15 in) H x 61.6 mm (2.42 in) D.
Probe & Sensor Configurations
• iBTX: Barometric Pressure and Temperature only.
• iBTHX: Barometric Pressure, Temperature, Humidity, and Dew Point.
• Wand Probe (Standard): 159 mm (6.25 in) overall length, 19 mm (0.75 in) diameter; cable with DB9 connector or stripped wire leads depending on model.
• Industrial Probe (-5 option): 137 mm (5 in) overall length, 16 mm (0.63 in) diameter stainless steel sheath; includes 3 m (10') cable with DB9 connector or stripped wire leads.
Data Output & Connectivity
• Network: RJ45 Ethernet connector supporting standard TCP/IP packets and Web Server (HTTP, Telnet).
• iBTX-SD Specifics: Includes 2 GB SD flash memory card for data logging in .txt file format; two 1.5 Amp relay outputs.
Power Supply Options
• iBT(H)X-W and iBTX-SD: 9 to 12 Vdc (includes universal 100 to 240 Vac power adapter; includes 9V alkaline battery for backup on SD models).
• iBT(H)X-D: 10 to 32 Vdc.
Key Product Differences
The primary distinction between the iBTX and iBTHX series is sensor capability: the iBTHX includes humidity and dew point sensing, while the iBTX measures only barometric pressure and temperature. The difference lies solely in the probe/sensor configuration.
• Model Suffixes: Models ending in -D are DIN rail mounted; models ending in -W are bench/wall mount units with Valox® 364 PBT cases; models ending in -SD include an LCD display, flash memory card, and backup battery housed in a steel metal case.
• Probe Types: Standard wand probes (152 mm/6" cable) are included unless the industrial probe option (-5) is specified. Industrial probes feature stripped wire leads or DB9 connectors depending on the base model configuration.
| AC Adaptor Operating Temperature | 0 to 40 °C |
| Accuracy Detail | Accuracy/Range at 25°C: Non-Condensing: 10 to 90%: 2.75% RH, 5 to 10% and 90 to 95%: ± 3% , 0 to 5% and 95 to 100%: ± 4% |
| Approvals | CE, FCC-B |
| Battery Operating Temperature | -18 to 55 °C |
| Configuration | Internal Web UI |
| Data Interface | Ethernet |
| Data Interface Note | Ethernet (RJ45): 10BASE-T Protocols :TCP, UDP, ARP, ICMP, DHCP, DNS, HTTP, and Telnet |
| Hysteresis | ± 1% RH |
| Input Type | External |
| Measurement Type | Barometric Pressure, Dew Point, Relative Humidity, Temperature |
| Non-Linearity | ± 3% |
| Number of Measurements | 3 |
| Operating Temperature, Max | 70 °C |
| Operating Temperature, Min | 0 °C |
| Pressure Accuracy | ± 3.5 mbar |
| Measurement Pressure, Max | 110 kPa |
| Measurement Pressure, Min | 1 kPa |
| Pressure Resolution | 0.1 mbar |
| Relative Humidity Accuracy | ± 2.75% for 10 to 90%; ± 3% for 5 to 10% and 90 to 95%; ± 4% for 0 to 5% and 95 to 100% |
| Repeatability | ± 0.1% |
| Response Time | Temperature: 5 to 30 sec, RH: 8 sec |
| Response Time Note | Temperature: 5 to 30 seconds, tau 63%, RH: 8 seconds, tau 63% |
| RH Resolution | 0.1%, 12 bit |
| Samples,Max | 85k Samples, Interval 1 sec |
| Series ID | ibtx-ibthx |
| Software | OPC Server; macro for data logging in Excel program; compatible with Windows® operating systems |
| Storage Temperature, Max | 85 °C |
| Storage Temperature, Min | 40 °C |
| Supply Power Note | Safety-Qualified AC Power Adaptor, Nominal Output: 9 Vdc @ 0.5 A , Input (iBT(H)X-W, iBTX-SD): 100 to 240 Vac, 50/60 Hz (included), Switching Power Supply (iBT(H)X-D): Sold separately (iDRN-PS-1000), Back-up Battery (iBTX-SD): 9V alkaline (included) |
| Temperature Accuracy | ± 0.5 °C |
| Temperature Repeatability | ± 0.1 °C |
| Temperature Resolution | 0.1 °C |
OMEGAIND-KIT Replacement Mounting Kit
Spare duct mount kit for HX92B Duct Mount and Remote Probe models
Call Today!
800-567-8686Normally ships in 2-3 weeks
OMEGACT485B-CAL-KIT Calibration Kit (Saturated salts)
Calibration kit for 110V or 220V units
Call Today!
800-567-8686Normally ships in 2-3 weeks
OMEGAIBTP-W-6 Wand probe, cable with DB9
Wand probe, cable 152 mm (6") with DB9 connector (barometric pressure/temperature)
Call Today!
800-567-8686Normally ships in 2-3 weeks
OMEGAIBTP-5 iServer MicroServer
Industrial probe 137 mm (5"), cable 3m (10') w/ stripped wire leads
Call Today!
800-567-8686Normally ships in 2-3 weeks
OMEGAIBTHP-W-6 iServer MicroServer
Wand probe, cable 152 mm (6") with DB9 connector (barometric pressure/temperature/humidity)
Call Today!
800-567-8686Normally ships in 2-3 weeks