The Omega M12TX series consists of analog PC-programmable 2-wire transmitters with integrated Pt100 RTD sensors. Designed for space-limited areas, these compact mineral-insulated probes feature a molded IP67 connector and are intended for industrial temperature measurement applications where the electronic section must remain within -40 to +80°C.

Operating Conditions & Performance

The M12TX series provides a configurable input temperature range from -50 to 800°C via software, with standard factory defaults typically set for ranges such as 0 to 150°C. The Pt100 sensor (Pt-0.00385) meets Class A accuracy in accordance with IEC751 up to 300°C. Transmitter accuracy is specified at a maximum of ±0.2°C or ±0.2% of span.

Performance characteristics include:

• Response Time (90%): <50 ms for the transmitter electronics.
• Sensor Response Time (63.2%, test in water to IEC751): Less than 3.5 seconds for Ø3 mm probes and less than 13 seconds for Ø6 mm probes.
• Ambient Temperature: -40 to +80°C (for the electronic section).
• Process Temperature Limits: Up to 500°C (-58 to 932°F) is listed for standard operation; higher process temperatures up to 800°C are configurable but require adherence to Tmax./Tenv. graphs to prevent damage from heat transfer.
• Humidity: 0 to 100% relative humidity.

Configuration Options

The series supports a wide envelope of configurations for probe diameter, insertion length, and output parameters. The transmitter is powered by an external supply or via the configuration interface during setup.

• Sensor Element: 3-wire Pt100 (Pt-0.00385) with a sensor current of 0.5 mA; maximum wire resistance per wire is 20 Ω.
• Output Signal: 4 to 20 mA analog output, non-isolated. Permissible load is 700Ω @24Vdc (calculated as R_load = (U - 8.5) / 0.022).
• Power Supply: 8.5 to 32 Vdc with polarity protection.
• Housing & Materials: Thermoplastic body for the transmitter section; probe sheath and thermowell options include AISI 316L stainless steel or compact mineral-insulated (MgO) cable construction.
• Degree of Protection: IP65 and IP67 in accordance with IEC60529. The molded connector is M12x1 male, compliant with VDE0627.
• Error Monitoring: Configurable sensor break monitoring (upscale >21 mA or downscale <3.6 mA) and fixed short-circuit action (downscale <3.6 mA).

Key Product Differences

Variants within the M12TX series are distinguished by probe diameter, insertion length, and specific model families.

• Diameter Options: Probes are available in 3 mm (approx. 1/8") or 6 mm (approx. 1/4").
• Length Configurations: Insertion lengths range from 100 mm to 1000 mm, with specific inch-based variants including 4", 6", 9", 12", 18", and 24".
• Bending Radius: The compact mineral-insulated cable probe allows a minimum bending radius of three times the outer diameter (excluding the sensing tip).
• Configuration Kit: Programming is performed using the M12TX-CONFIG-SET, which includes USB interface hardware and software for Windows 10.0 or higher to adjust range span, units (°C/°F/K), sensor error compensation, and transmitter identification.

• Programmable Transmitter 

• Molded Connector (IP67) 

• M12 Connection 

• 4 to 20 mA Output 

AccuracyClass A
Body MaterialThermoplastic
Downscale3.6 mA
Electrical Output4 to 20 mA
EMCIn accordance to EN 61326
Enclosure RatingsIP67
Enclosure Ratings NoteIP65 and IP67 in accordance to IEC60529
Factory Setting0 to 150°C/sensor break >21 mA (upscale)
Operating Temperature, Max80 °C
Operating Temperature, Min-40 °C
Permissible Load700
Permissible Load Note700 @ 24 Vdc [RLo= (Vsupply-8.5)/0.020]
Process Connection SizeN/A
Process Temperature, Max500 °C
Process Temperature, Min-50 °C
Range Configurations NoteIt is possible set the input temperature range (span) by the M12TX-CONFIG-SETconfiguration kit (PC with OS Windows® required)
Relative Humidity, Max1
Relative Humidity, Min0%
Response Time< 50 ms
RTD Type100 Ω Pt Sensor
Sensor Break Monitoring NoteSelectable: Upscale (>21.0 mA) or downscale (<3.6 mA) action Sensor Short-Circuit: Fixed to downscale (<3.6 mA) action
Sensor Error Compensation NoteOver 2 points (maximum 1% of span)
Series IDM12TX
Span, Min50°C
Span Note50°C (122°F) [if the zero value is set between one of these values: -40°C (-40°F), -20°C (-4°F), 0°C (32°F), 20°C (68°F), 40°C (104°F), the minimum span is 20°C (68°F) rather than 50°C (122°F)]
Supply Power EffectsNegligible
Supply Voltage8.5 to 32 Vdc
Temperature Influence, Max± 0.3%
Temperature Influence NoteMaximum of ± 0.3°C/25°C (77°F) or ± 0.3% of span/25°C (77°F)
Temperature Sensor TypeRTD
Thermocouple TypeN/A
Tmax. Electronic80°C
Upscale21 mA
Zero Setting Adjustable RangeAny value between -50 to 50°C (-58 to 122°F)
  • Download

    Data Sheet - M12TX RTD Probe w/ 4 to 20mA Analog Output & M12 Connector

    Download
  • Download

    Manual - M12TX RTD Probe w/ 4 to 20mA Analog Output & M12 Connector

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more