Skip to main content
Press Enter
OMEGA

M12TXC RTD probes w/ 4 to 20mA Analog Output & Mounting Threads

Compact and Programmable M12 RTD Temperature Transmitters

Normally ships in
3-4 weeks
Have questions? We're here to help!
Have questions? We're here to help!
Configure

 Supported Applications

The Omega M12TXC series consists of compact, programmable RTD temperature transmitters designed for space-limited applications. These devices integrate a Pt100 sensor with an analog transmitter to provide direct process monitoring in industrial environments where footprint constraints are critical.

Operating Conditions & Performance

The M12TXC series operates within specific thermal and electrical envelopes defined by the sensor technology and electronic section. The measurement range is documented as -50 to 110°C (-58 to 248°F) for standard configurations, with a separate variant (M12TXSS) covering -50 to +120°C. The ambient temperature limit for the plastic body housing is -40 to 80°C (-40 to 176°F). Accuracy specifications include a transmitter maximum of ±0.2°C or ±0.2% of span, and a Pt100 sensor rated as Class A in accordance with IEC751 up to 300°C.

Response time performance is tested in water per IEC751 (time for reaching 63.2% of instantaneous temperature change). For the M12TXC-PT100 series, response times are <3.5 seconds for a diameter of 3 mm and <13 seconds for a diameter of 6 mm. The transmitter section itself has a response time (90%) of less than 50 ms.

Configuration Options

The series offers configurable dimensions, process connections, and output parameters to accommodate specific installation requirements:

• Sensor Element: RTD Pt100 (Pt - 0.00385) with a 3-wire connection.
• Diameters & Lengths: Available in Ø3 mm and Ø6 mm diameters; insertion lengths include 13 mm, 24 mm, and other options depending on the specific model family (e.g., M12TX-PT100).
• Process Connections: Options include 1/8" NPT male thread or G1/8 thread.
• Housing & Wetted Materials: The body is constructed from stainless steel AISI 316L. Probe types vary between thermowell AISI 316L and compact mineral insulated AISI 316L depending on the model variant (M12TXC vs M12TX-PT100).
• Electrical Output: Provides a 4 to 20 mA analog signal. The permissible load is 700Ω @ 24 Vdc, calculated as R_load (Ω) = (U - 8.5) / 0.022.
• Power Supply: Operates on 8.5 to 32 Vdc with polarity protection.
• Degree of Protection: Rated IP65 and IP67 in accordance with IEC60529.

The series supports PC-based configuration via the M12TX-CONFIG-SET kit, which includes a USB interface and software for Windows 10.0 or higher. This allows users to set the input temperature range (span) between -50°C to 800°C, adjust zero values, and configure sensor break monitoring actions (upscale >21.0 mA or downscale <3.6 mA).

Key Product Differences

The M12TXC series is distinguished from the broader M12TX family by its compact form factor, specifically designed for tight spaces with an overall length of 50 mm (1.96") and a probe tip diameter as small as Ø3 mm (0.11"). While the general M12TX-PT100 series supports ranges up to +800°C using compact mineral insulated probes, the specific M12TXC variants listed are optimized for lower temperature applications (-50 to 110°C) with shorter insertion lengths.

Model variations primarily differ by process connection thread type (NPT vs. G-thread) and probe length (13 mm or 24 mm). The series includes an M12 male connector in accordance with VDE0627 for the output signal, whereas other family members like the M12TX-PT100 may feature separate input/output connectors.

  • Programmable Transmitter
  • M12 Connection
  • 4 to 20 mA Output
  • -50 to 120°C (-58 to 248°F)
AccuracyClass A
Accuracy DetailClass A per IEC60751
Body MaterialStainless Steel AISI 316L
Downscale3.6 mA
Electrical ConnectionM12 4 pin Male
Electrical Output4 to 20 mA
EMCIn accordance to EN 61326
Enclosure RatingsIP67
Enclosure Ratings NoteIP65 and IP67 in accordance to IEC60529
Factory Setting0 to 100°C/ sensor break >21 mA (upscale)
Operating Temperature, Max80 °C
Operating Temperature, Min-40 °C
Permissible Load700
Permissible Load Note700 @ 24 Vdc [RLo= (Vsupply-8.5)/0.020]
Probe Diameter3 mm
Process Temperature, Max120 °C
Process Temperature, Min-50 °C
Range Configurations NoteIt is possible set the input temperature range (span) by the M12TX-CONFIG-SETconfiguration kit (PC with OS Windows® required)
Relative Humidity, Max1
Relative Humidity, Min0%
Response Time< 50 ms
Response Time NoteDiameter 3 mm (0.12") <3.5 seconds, diameter 6 mm (0.24") <13 seconds (test in water to IEC751- time for reaching 63.2% of instantaneous temperature change)
RTD Type100 Ω Pt Sensor
Sensor Break Monitoring NoteSelectable: Upscale (>21.0 mA) or downscale (<3.6 mA) action Sensor Short-Circuit: Fixed to downscale (<3.6 mA) action
Sensor Error Compensation NoteOver 2 points (maximum 1% of span)
Series IDM12TXC
Sheath Material316 Stainless Steel
Span, Min50°C
Span Note50°C (122°F) [if the zero value is set between one of these values: -40°C (-40°F), -20°C (-4°F), 0°C (32°F), 20°C (68°F), 40°C (104°F), the minimum span is 20°C (68°F) rather than 50°C (122°F)]
Supply Power EffectsNegligible
Supply Voltage8.5 to 32 Vdc
Temperature Influence, Max± 0.3%
Temperature Influence NoteMaximum of ± 0.3°C/25°C (77°F) or ± 0.3% of span/25°C (77°F)
Temperature Sensor TypeRTD
Thermocouple TypeN/A
Tmax. Electronic80°C
Upscale21 mA
Zero Setting Adjustable RangeAny value between -50 to 50°C (-58 to 122°F)
  • Download

    Data Sheet - M12TXC RTD probes w/ 4 to 20mA Analog Output & Mounting Threads

    Download
  • Download

    Manual - M12TXC RTD probes w/ 4 to 20mA Analog Output & Mounting Threads

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more