OM-CP High Temperature Data Loggers, Dual Remote Temperature Probes
High Temperature Dual Channel Data Loggers With Two Remote Temperature Probes
Supported Applications
The OM-CP-HITEMP140X2-FP Series are High Temperature Dual Channel Data Loggers with two remote temperature probes and a stainless steel body. Their dual probes allow simultaneous temperature monitoring at two points, which suits applications such as oven mapping and other high-temperature profiling where two zones must be compared during the same cycle.
The loggers are available either with two flexible probes or with one flexible probe combined with a stainless steel bendable probe, giving the flexibility to reach and shape a sensor to fixtures, ducts, or product surfaces while a second probe monitors a reference point.
Operating Conditions & Performance
Both channels deliver a resolution of 0.01°C (0.02°F) and a calibrated accuracy of ±0.1°C (±0.18°F). Measurement range depends on the probe: the flexible probe covers -60 to 260°C (-76 to 500°F), while the stainless steel bendable probe covers -200 to 350°C (-328 to 662°F).
• Reading rate: 1 reading every second up to 1 reading every 24 hours.
• Memory: 32,767 readings, or 16,383 readings in trigger settings mode.
• Body (ambient) operating temperature: -40 to 140°C (-40 to 284°F).
• Enclosure rating: IP68.
• Humidity: 0 to 100% RH.
• Power: 3.6V high-temperature lithium battery included; user replaceable.
The logger body measures 48 H x 24.6 mm dia (1.89 x 0.97") in 316 stainless steel, and the flexible probe is 2.5 mm dia (0.1") on a 183 cm (72") length. Wetted materials are 316 stainless steel and PEEK. The series carries CE and RoHS, with optional NIST calibration and optional 5 YEAR WARRANTY, and requires the OM-CP-IFC400 USB docking station or OM-CP-IFC406 multiplexer interface for operation.
Warning: the product is rated for use up to 140°C; the battery may leak or explode if heated above 150°C (302°F).
Configuration Options
All three models share the same dual-channel logger body, accuracy, and resolution; they differ in the pairing of probes.
Two flexible probes
The OM-CP-HITEMP140X2-FP-72 is supplied with two 183 cm (72") flexible probes for monitoring two remote points with identical, highly routable sensors. This model weighs 115 g (4.1 oz).
Flexible plus bendable probe
The OM-CP-HITEMP140X2-FP-72-PT-1 and OM-CP-HITEMP140X2-FP-72-PT-5 pair one 183 cm (72") flexible probe with a 61 cm (24") stainless steel bendable probe, differing in the bendable probe tip: the PT-1 uses a 2.5 cm (1") tip and the PT-5 uses a 12.7 cm (5") tip. These configurations weigh 110 g (3.9 oz) and add the bendable probe's wider -200 to 350°C (-328 to 662°F) range.
Key Product Differences
The model lineup consists of the following dual-probe configurations:
• OM-CP-HITEMP140X2-FP-72 — high temperature data logger with two 183 cm (72") flexible probes.
• OM-CP-HITEMP140X2-FP-72-PT-1 — high temperature data logger with 183 cm (72") flexible probe and 61 cm (24") bendable stainless steel probe with 2.5 cm (1") tip.
• OM-CP-HITEMP140X2-FP-72-PT-5 — high temperature data logger with 183 cm (72") flexible probe and 61 cm (24") bendable stainless steel probe with 12.7 cm (5") tip.
Each configuration is also available with a NIST calibration certificate (-CERT). Required interfaces are the OM-CP-IFC400 and OM-CP-IFC406, and the OM-CP-BAT110 is the replacement battery.
• OM-CP-HITEMP140X2-FP-72 Has Two Flexible Probes
• OM-CP-HITEMP140X2-FP-72-PT-1 and OM-CP-HITEMP140X2-FP-72-PT-5 Have One Flexible Probe and One Stainless Steel Bendable Probe
• Flexible Probe Measures from -60 to 260°C
•(-76 to 500°F)
| Battery life | up to 1 yr |
| Battery Size and Type | 3.6V Li-ion |
| Calibration | Digital calibration through software |
| Data Interface | USB |
| Data Interface Note | OM-CP-IFC400 USB docking station or OM-CP-IFC406 multiplexer interface required; 125,000 baud |
| Diameter | 24.6 mm |
| Enclosure Ratings | IP68 |
| Form factor | Handheld |
| Height | 48 mm |
| Measurement Type | Temperature |
| Memory | 32767 |
| Memory Note | 32,767 readings |
| Number of Measurements | 2 |
| Operating Temperature, Max | 140 °C |
| Operating Temperature, Min | -40 °C |
| Readings in Trigger Settings Mode | 16,383 readings |
| Real Time Recording | May be used with PC to monitor and record data in real time |
| Relative Humidity, Max | 1 |
| Relative Humidity, Min | 0% |
| Samples,Max | 33000 |
| Sampling Rate | 1 Hz (1 update/second) |
| Sampling Rate Note | 1 reading every second to 1 every 24 hours. |
| Series ID | OM-CP-HITEMP140X2-FP-Series |
| Software Compatibility | Windows XP SP3/Vista/7 and 8 (32- and 64-bit) |
| Supply Power Type | Battery |
| Temperature Resolution | 0.01 °C |
| Temperature Sensor Type | RTD |
| Time Accuracy | 1 minute/month @ 25°C (77°F) |
| Weight | 110 g |