Skip to main content
Press Enter
OMEGA

OM-CP Lyophilization Temperature Data Logger

Lyophilization Temperature Data Logger

Normally ships in
3-4 weeks
Have questions? We're here to help!
Have questions? We're here to help!

 Supported Applications

The Omega OM-CP-LYOTEMP is a lyophilization temperature data logger designed for use in ultra-low operating temperatures ranging from -60 to 75°C (-76 to 167°F). The series supports processes such as lyophilization and includes an optional surface temperature probe attachment, the OM-CP-MICRODISC, for monitoring surface conditions.

Operating Conditions & Performance

The device operates within a process and ambient environment of -60 to 75°C (-76 to 167°F) at 0 to 100% RH. It features an accuracy of ±0.5°C with a resolution of 0.1°C. The internal clock maintains time accuracy of ±5 minutes/month at -80°C (-112°F). Data is recorded in date and time-stamped formats supporting °C, °F, K, or °R units.

Power is supplied by an included 3.6V lithium battery (factory replaceable only), providing a typical life of one year at a 10-minute reading rate. The unit carries an IP50 ingress protection rating and weighs 1.4 oz (40 g).

Configuration Options

The series offers configurable options for calibration, software compliance, mounting, and specialized sensing:

• Calibration & Certification: OM-CP-LYOTEMP-CERT includes a NIST calibration certificate.
• Software & Interface: The OM-CP-IFC300 package provides Windows software, a docking station, and USB cable. Software supports XP SP3/Vista/7 and 8 (32-bit and 64-bit) with communication at 38,400 baud.
• Compliance: OM-CP-SVP-SYSTEM offers an FDA 21 CFR Part 11 compliant software validation package including IQ/OQ/PQ workbooks.
• Mounting & Accessories: Options include the OM-CP-WM-300 wall mount station and the OM-CP-MICRODISC surface temperature probe attachment (Brass C46440, silicone rubber insert).

The device features LED status indicators: Green for safe conditions during logging; Yellow when warn limits are exceeded; Red when alarm criteria (limits/delay) are exceeded. Password protection is available to restrict access to configuration options while allowing data readout without the password.

Key Product Differences

The base model OM-CP-LYOTEMP functions as a standard lyophilization temperature logger with manual start/stop and configurable alarms. The OM-CP-LYOTEMP-CERT variant adds NIST calibration documentation to the same hardware platform.

Differentiation is primarily achieved through accessory selection: the OM-CP-MICRODISC changes the sensing application from internal process monitoring to surface temperature measurement, operating in an environment of -60 to 177°C (-75 to 350°F). The OM-CP-SVP-SYSTEM provides a distinct software validation layer for regulated environments without altering the logger's physical specifications.

• Ultra Low Operating Temperature Range from -60 to 75°C (-76 to 167°F) 

• Hermetically Sealed Thermistor Probe 

• LED Status Indicators 

• Manual Start and Stop Options 

Additional Features 1Ultra Low Operating Temperature
Alarm DelayProgrammable time delay
Alarm NoteSoftware programmable high and low alarm limits, high and low warn limits; each set point may be individually enabled or disabled
Battery lifeup to 1 yr
Battery Size and Type3.6V Li-ion
CalibrationDigital calibration through software
Data InterfaceUSB
Data Interface NoteUSB docking station to PC; 38,400 baud
Display TypeLED
EnclosureABS
Enclosure RatingsIP50
LED IndicatorsGreen : Blinks to indicate safe conditions during logging; Yellow : Blinks to indicate warn limits have been exceeded; Red : Blinks to indicate alarm criteria (limits/delay) have been exceeded
Length82.5 mm
Measurement TypeTemperature
Memory NoteMemory Reading Per Channel 32767
Number of Measurements1
Probe Length121.9 cm
Relative Humidity Accuracy± 3% RH
Sampling Rate5 seconds between updates
Sampling Rate Note1 reading every 5 seconds up to 1 reading every 30 minutes
Series IDOM-CP-LYOTEMP-Logger
Software CompatibilityXP SP3/Vista/7 and 8 (32-bit and 64-bit)
Supply Power TypeSupply Power Type
Temperature Accuracy± 0.5 °C
Temperature Measurement, Max75 °C
Temperature Measurement, Min-60 °C
Temperature Resolution± 0.1 °C
Time Accuracy± 5 minute/month
Weight40 g
Width53.3 mm
  • Download

    Data Sheet - OM-CP Lyophilization Temperature Data Logger

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more