Skip to main content
Press Enter
OMEGA

OM-EL-USB-CO Carbon Monoxide Data Loggers w/USB Interface

Carbon Monoxide Data Loggers with USB Interface

Normally ships in
2-3 weeks
Have questions? We're here to help!
Have questions? We're here to help!
Configure

Supported Applications

The Omega OM-EL-USB-CO Series consists of professional USB data loggers designed to record Carbon Monoxide (CO) gas levels for later analysis. These standalone instruments measure CO concentration in free air and are intended for non-life-saving analysis projects only. The series is explicitly not an alarm, safety, or warning device; readings are taken at user-set sample intervals, and the LED/sounder indications do not provide adequate warning of dangerous CO levels.

For applications where human or animal safety is a concern, a nationally approved domestic or industrial CO detector/alarm must be installed. The OM-EL-USB-CO Series should not be used as a critical component in life support devices or systems.

Operating Conditions & Performance

The series operates within an ambient temperature range of -10 to 40°C (14 to 104°F) and supports operating pressures from 900 to 1100 mbar. Humidity ratings are specified as 15 to 90% RH (non-condensing). The sensor life is documented at 4 years; units must be replaced once expired.

Performance specifications vary by model:

• OM-EL-USB-CO: Measurement range of 3 to 1000 ppm CO. Accuracy is ±2 ppm min, ±6% typ (Overall Error). Do not expose this model to levels above 1000 ppm.
• OM-EL-USB-CO300: Measurement range of 0 to 300 ppm CO. Accuracy is ±5 ppm or ±4% typ, whichever is greater.

Both models feature an internal resolution of 0.5 ppm CO and a repeatability of ±3%. The response time (to reach 90%) is 1 minute. Long-term drift is specified at 5%/year with a zero offset of ±2 ppm CO.

Configuration Options

The series offers configurable logging parameters via the supplied OM-EL-WIN-USB control software (EasyLog USB), compatible with Windows XP/Vista/7 and 8 (32- and 64-bit). Users can configure the logger name, start date/time, warning threshold, and logging rate.

Key configurable options include:

• Logging Intervals: Selectable from a drop-down list: 10 sec, 30 sec, 1 min, or 5 min.
• Status Indicators: Red and green LEDs indicate logging status (primed, in progress, low battery) and alarm conditions. An audible sounder activates approximately once per second when the warning threshold is exceeded and once every 60 seconds for low battery.
• Latching Option: Latching can be enabled via software so the red LED remains latched if a stored value equals or exceeds the warning threshold.

The devices are powered by a user-replaceable ½ “AA” 3.6V lithium battery (included; replacement PN: OM-EL-BATT). Battery life is approximately 3 months with the supplied battery at a 5-minute logging rate, 25°C ambient temperature, audible warning disabled, and reading <10 ppm CO.

Key Product Differences

The series comprises two distinct models differentiated by measurement range:

• OM-EL-USB-CO: Designed for a 3 to 1000 ppm CO range. This model is not suitable for direct exposure to levels above 1000 ppm (e.g., car or wood burner exhaust).
• OM-EL-USB-CO300: Designed for a lower concentration range of 0 to 300 ppm CO.

Both models share identical physical dimensions: 125 mm (4.9 in) length and 26.4 mm (1.04 in) diameter, with a weight of 43 g (1.5 oz). Both include a protective cap and connect to a PC via USB for setup and data download.

• 0 to 1000 ppm CO (OM-EL-USB-CO) or 0 to 300 ppm CO (OM-EL-USB-CO300) Measurement Ranges
• Stores up to 32,510 Measurements
• User-Programmable Warning Threshold with Bright Visual and Audible Warnings
• USB Interface for Set-Up and Data Download
Accuracy± 6 %
Accuracy Detail±2 ppm min,±6% typ
Alarm NoteThe audible alarm will sound once per second when the warning threshold has been exceeded. To disable the warning, remove the cap from the data logger. The audible warning will sound once every 60 seconds when there is a low battery condition. This indicates that the battery must be replaced.
Battery lifeup to 3 mo
Battery Size and Type3.6V Li-ion
Data InterfaceUSB
Diameter1.04 in
Indicator TypeRed LED, Green LED
Length4.9 in
Long Term Drift5%/year
Measurement TypeCarbon Monoxide
Number of Measurements1
Operating Temperature, Max40 °C
Operating Temperature, Min-10 °C
Relative Humidity15 to 90% RH non-condensing
Repeatability±3%
ResolutionMeasurement: 0.5 ppm CO
Response Time1 min
Sampling Rate10 Hz (10 updates/second)
Sampling Rate Note10 sec, 30 sec, 1 min, 5 min
Sensor TypeCO Detector
Series IDOM-EL-USB-CO-Series
SoftwareWindows XP/Vista/7 and 8
Supply Power TypeBattery
Weight43 g
Working Pressure900 to 1100 mbar
Zero/Balance Error±2 ppm CO
  • Download

    Data Sheet - OM-EL-USB-CO Carbon Monoxide Data Loggers w/USB Interface

    Download
  • Download

    Manual - OM-EL-USB-CO Carbon Monoxide Data Loggers w/USB Interface

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more