Skip to main content
Press Enter
OMEGA

PT-104A 4-Channel RTD Input DAQ Module w/USB or Ethernet Interface

4-Channel RTD Input Data Acquisition Module with USB or Ethernet Interface

Normally ships in
2-3 weeks
Have questions? We're here to help!
Have questions? We're here to help!

Supported Applications

The Omega PT-104A is a four-channel, high-resolution temperature data acquisition module designed for use with Pt100 and Pt1000 RTD sensors. Beyond temperature measurement, the series supports resistance measurements in ranges of 0 to 375 Ω and 0 to 10 kΩ, as well as voltage inputs ranging from 0 to 115 mV or 0 to 2.5 V. The module is compatible with 2-wire, 3-wire, and 4-wire RTD sensor configurations, with 4-wire Pt100 sensors recommended for the highest accuracy.

The PT-104A facilitates data collection via Windows Logging Software or a provided Software Development Kit (SDK) containing drivers and example code for C, C++, Excel, and LabView. Up to 20 units of PT-104A modules can be connected simultaneously in various configurations.

Operating Conditions & Performance

The module operates within an ambient temperature range of 0 to 70°C (32 to 158°F). For stated accuracy, the environmental condition is specified as 20 to 30°C (68 to 86°F) with a relative humidity rating of 20 to 90% RH. The device features Class 'A' DIN Platinum Elements per IEC 751.

Performance specifications include:

• Temperature Range: -200 to 800°C (-328 to 1472°F)
• Resolution: 0.001°C for temperature; 1 μΩ and 0.156 μV for resistance/voltage
• Accuracy @ 25°C: 0.01°C (temperature), 20 ppm (resistance), or 0.2% (voltage)
• Linearity: 20 ppm
• RMS Noise (Using Filter): 0.01°C for temperature; 10 ppm for resistance/voltage
• Conversion Time Per Channel: 720 mS
• Input Impedance: >>1 MΩ
• Overvoltage Protection: ±30V

The response time is less than 75 milliseconds in water flowing at 3 feet per second.

Configuration Options

The PT-104A series offers flexible connectivity and power options through USB or Ethernet interfaces. The module can be powered via USB (USB 1.1: 5V ±10% @ <100 mA; USB 2.0: 5V ±10% @ <200 mA) or Power-Over-Ethernet (PoE). It conforms to IEEE 802.3af PoE standards and is compatible with 10/100/1000Base-T networks.

Key configurable features include:

• Sensor Inputs: Supports Pt100 and Pt1000 RTDs in 2, 3, or 4-wire configurations via a 4-pin mini DIN connector
• Data Interface: USB (conforms to USB 2.0 full-speed at 12 Mbps) or Ethernet (IEEE 802.3 10Base-T)
• Software Support: Includes Windows Logging and Player Software for data visualization in spreadsheet or graph formats, plus an SDK
• Physical Dimensions: 36 H x 135 W x 184 mm D (1.42 x 5.31 x 7.24") with a weight of 500 g (1.1 lb)

The standard package includes the PT-104A module, USB cable, Ethernet patch cable, four mating mini DIN screw terminal connectors, quick start guide, Windows software, and a complete operator's manual on CD.

Key Product Differences

The series is identified by model number PT-104A as the 4-channel RTD input data acquisition module. The device supports multiple operational modes including USB-only connections, USB-powered with Ethernet data transmission, and full Power-over-Ethernet configurations.

Accuracy specifications vary based on measurement type: temperature accuracy is quoted at 0.01°C for options marked in the specification table; resistance accuracy is 20 ppm; voltage accuracy is 0.2% (with a note that for the 115 mV range, accuracy may vary by 2%). The temperature coefficient is specified as 5 ppm/°C for standard ranges and 100 ppm/°C for specific conditions.

• Measures Temperature (RTDs), Resistance and Voltage
• High Accuracy (0.01°C) and Resolution (0.001°C)
• Powered by USB Port or Power-Over-Ethernet (PoE)
• For Use with 2, 3 and 4-Wire Pt100 and Pt1000 RTD Sensors

 

Accuracy± 0.002%
Conversion Rate720 msec
Data InterfaceEthernet, USB 2.0
Data Interface NoteEthernet Cable Length, Max: 100 m;Ethernet Port: Conforms to IEEE 802.3 10Base-T.; Compatible with 10/100/1000Base-T networks.; Conforms to IEEE 802.3af Power-over-Ethernet (PoE);;Supported Operating System: Microsoft Windows®XP SP2 / VISTA/7, Software Development Kit containing drivers and example code for C, C++, Excel and LabView.
Electrical ConnectionDIN 43650 C (solenoid valve)
Height36 mm
Impedance Input>1 MΩ
Input Noise VoltageRMS: RTD: 0.01°C, Resistance: 10 ppm, Voltage:10 ppm
Input TypeRTD, Voltage, Resistance
Input Type NoteRTD: Pt100, Pt1000, 2, 3 and 4-Wire
Length184 mm
LinearityRTD: 20 ppm, Resistance: 20 ppm, Voltage: 20 ppm
Number of Channels, InputTemperature: 4
Operating Temperature, Max70 °C
Operating Temperature, Min0 °C
Relative Humidity20 to 90% RH
Resistance Measurement Range0 to 375 Ω, 0 to 10 kΩ
ResolutionRTD: 0.001°C, Resistance: 1 µΩ, Voltage: 0.156 µV
Series IDPT-104A-DAQ-Module
SoftwareLogging Software
Supply CurrentUp to 200 mA
Supply Power NotePowered by USB or Ethernet: USB 1.1: 5V ±10% @ <100 mA, USB 2.0: 5V ±10% @ <200 mA, Ethernet: 48V ±20% @ <40 mA (<2W)
Supply Voltage5 Vdc
Temperature CoefficientRTD: 5 ppm/°C, Resistance: 5 ppm/°C, Voltage:100 ppm/°C
Voltage Measurement Range0 to 115 mV, 0 to 2.5V
Weight500 g
Width135 mm
  • Download

    Data Sheet - PT-104A 4-Channel RTD Input DAQ Module w/USB or Ethernet Interface

    Download
  • Download

    Manual - PT-104A 4-Channel RTD Input DAQ Module w/USB or Ethernet Interface

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more