On-line measurement of moisture content in liquids is accurate, fast and reliable with the Liquidew I.S. Moisture in Liquid Analyzer from Michell. A wide variety of non-polar liquids can be measured continuously, on-line, including flammable liquids and hazardous area applications, petrochemicals, power industry and pharmaceutical manufacturing.

• Simple, cost-efficient operation and low maintenance
• Retrofit functionality, easy to integrate into existing sampling or distribution systems
• High integrity moisture in liquid measurement from 0.01 ppmW to saturation point
• Multi-channel with up to four completely independent measurement channels
• Robust design for undisturbed operation in many non-polar liquids over a long time
• Two 4-20 mA configurable outputs and digital Modbus RTU-based communication
• Replaceable sensor element with Michell Calibration Exchange • Service for reliable cost-efficient calibration
• NIST and NPL traceable calibration
• EExd version available

Sensor
Sensor TechnologyMichell Ceramic Moisture Sensor
Enclosure MountingStainless steel wall mounting brackets
Enclosure Ingress ProtectionIP66
Enclosure Temperature ControlHeater/thermostat options for fixed set-point +20°C or adjustable set-point range 0 to control +50°C
Heater Power Supply110/120 or 220/240/255 V AC, 50/60 Hz.
Power consumption 100 W max
Operating EnvironmentShaded position, on or off shore, -20 to +40°C (-40 to +60°C max transient)
Enclosure cooling option recommended for climate ambient > +50°C
Sample Cooling OptionRecommended for process fluid temperatures > +40°C
Plant water supply required ≤ +30°C
Sensor VersionEasidew PRO I.S
Measurement Range0.001 to 1000 ppmw
Calibration Range-100 to +20°Cdp
Accuracy Dew point: ±1°C between -59.9 and +20°Cdp, (±1.8°F between -75.9 and +68°Fdp)
±2°C between –60 & -100°Cdp,(±3.6°F between -76 and -148°Fdp)
Moisture content: ±10% of reading, ±20% of reading
Resolution0.1°C between +20 and -100°Cdp
Temperature MeasurementPt100
Temperature Measurement Range-20 to +70°C
Temperature Measurement Accuracy± 0.2°C
Analysis PressureUp to 5 MPa (50 barg)
Analysis Temperature0 to +50°C
Sample Flow Rate Min 0.01 l/min
Max 10 l/min
0.1 to 0.3 l/min recommended
CalibrationTraceable to British (NPL) and American (NIST) National Humidity Standards
Certification
Hazardous Area CertificationATEX II 1 G Ex ia IIC T4 Ga (–20°C to +70°C)
IECEx Ex ia IIC T4 Ga (–20°C to +70°C)
TC TR 0Ex ia IIC T4 Ga (–20°C to +70°C)
FM Class I, Division 1, Groups A B C D, T4
cCSAus Class I, Division I Groups A B C D, T4
Pattern ApprovalGOST-R, GOST-K
Control Unit
DisplayTwo line 6-digits LED, displaying moisture content / dew point (user toggle) and temperature
Analogue OutputTwo 4-20 mA (max load 500 Ω)
User-configured for parameter, unit and range
Digital OutputRS485 Modbus RTU
Display ModeMoisture content (ppmw)
Dew point (°C or °F)
Temperature (°C or °F)
Display Resolution0.1°Cdp, 0.1°Fdp, 0.001 ppmw, 0.1°C temp
AlarmsFour alarm relays. Control action and set-point are user-programmable
Two Form C contacts rated 10 A, 240 V AC or 8 A, 24 V DC. Non-inductive load
Two Form A contacts rated 5 A, 240 V AC or 4 A, 24 V DC. Non-inductive load
I.S. BarriersGalvanic isolation type, integrated to Control Unit
Power Supply85 to 265 V AC, 50/60Hz or 10 to 72 V DC
10 W max power consumption
Enclosure19" sub rack unit
Dimensions 132 x 483 x 375mm (h x w x d)
100mm minimum clearance depth for cables and vents
Operating EnvironmentIndoor, safe area, 0 to +50°C, < 90% RH
Premium Sampling System
Enclosure304 stainless steel (EN1.4301) enclosure
Option for complete enclosure in 316 stainless steel (EN1.4401)
All fixtures stainless steel
Galvanised steel internal mounting plate
Open panel version available for indoor installation
Dimensions 800 x 600 x 300mm (h x w x d)
  • Thumbnail of document Manual - SAM-LDI Liquidew I.S. Moisture in Liquid Analyzer
    Download

    Manual - SAM-LDI Liquidew I.S. Moisture in Liquid Analyzer

    Download
  • Thumbnail of document Data Sheet - SAM-LDI Liquidew I.S. Moisture in Liquid Analyzer
    Download

    Data Sheet - SAM-LDI Liquidew I.S. Moisture in Liquid Analyzer

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more