 With a 10-year battery and internal memory for one million readings, the SRX6 Precision Thermocouple Logger is built for long deployments and minimal maintenance. Whether you're collecting months of high-resolution data or monitoring hard-to-reach locations, you can trust it to keep recording—without the need for power cables, recharging, downloading, or constant oversight.

 
The SRX6 has been designed for demanding fieldwork, lab testing, and industrial applications. It’s trusted in everything from cryogenic monitoring to concrete maturity and high-heat processing environments. With support for standard thermocouple types and long battery life, it’s a flexible, dependable tool for capturing detailed thermal profiles—even in harsh conditions.

 • Seven Independent Thermocouple Inputs

Flexibly configure each of the seven channels to support Type J, K, T, E, or S thermocouples for multi-point or mixed-type measurements.
• High Resolution 16-Bit Data1
Delivers enhanced accuracy and fine-grained resolution—up to 0.04° C in sensitive temperature ranges
• Calibration Flexibility
Support for multi-point calibration lets you optimize accuracy across your specific measurement range using ACR’s TrendReader TRX Pro software.
• Tough Environmental Performance
Operates reliably from –40° C to +85 °C (logger body), suitable for harsh industrial, refrigerated, or outdoor settings.
• Wide Temperature Range: –200 °C to +1 ,700 °C
The SRX6 supports measurements across a broad temperature range—from cryogenic conditions to high-temperature industrial applications.
• Self-Powered Long Life Battery
Runs for up to 10 years1 on internal batteries, with no need for recharging or external power—well suited for long-term deployments in the field or hard-to-access locations.
• Large Internal Memory - One Million Readings
The ample storage provides flexibility for a range of applications—reducing the need for memory-saving strategies and helping ensure that longer or more detailed logging sessions can run without interruption.
• Integrated Two-Stage Alarm Output
Dual open-drain outputs allow direct integration with alarms, lights, or PLCs for both warning and critical conditions.
 
Number of ChannelsEight - Seven thermocouple inputs plus one cold junction temperature sensor
Thermocouple Types SupportedJ, K, T, E, and S
Internal (CJT) SensorNTC Thermistor - 10,000 Ω @ 25 °C (77° F)
Clock Accuracy±1 second per day
Voltage Inputs
Input Voltage~ -5 to ~70 mV
Input Impedence>2 MΩ
Overvoltage ProtectionInput channels include internal protection against ESD and moderate overvoltage
Signal ProcessingEquation-based conversion (ITS-90 standard)
Voltage Resolution16-bit ADC; ~2.0 µV/bit
Logger Voltage Input Accuracy±120 µV typical within input range
Data Logging
Memory SizeOne million readings
Sampling Interval10 seconds to 24 hours
Sampling MethodsContinuous; Stop when full; Delayed Start; Spot / Averaged
Data RetentionNon-volatile memory preserves logged data and configuration
Power
Battery TypeTwo 3.6V Lithium Thionyl Chloride cells
Total Capacity~4800 mAh (2400 mAh per cell)
Battery LifeUp to 10 years (typical use); turn off channels to save battery
Rechargeable?
No. Contact ACR for battery replacement
Total Lithium Content<2g (IATA exempt)
Alarms
Alarm Outputs2 open-drain outputs
Alarm CapacityUp to 15 VDC, 350 mA per channel
Alarm ModesTwo-stage trigger capability
Led indicatorsGreen: logging; Blue: delayed start; Yellow: alarm warning; Red: critical alarm; Magenta: memory full
Physical
Size111 x 74 x 27 mm (4-3/8 x 2-15/16 x 1-1/16”)
Weight140g (5 oz)
InterfaceUSB (via software)
Enclosure MaterialSabic Cycoloy C2950 thermoplastic (high impact, flame retardant)
Environment
Operating Temperature-40 to +85°C (–40°F to +185°F)
Storage Temperature-20°C to +60°C (–4°F to +140°F)
Humidity Range0–95% RH (non-condensing)
Certifications
CECompliant with applicable EU directives
RoHSCompliant with Directive (2011/65/EU) incl. Category 9
FCC Part 15Complies with FCC Part 15 for Class B digital devices
IATAAir transport compliant (non-rechargeable lithium cells)
WEEERecycle/dispose per WEEE; can return to ACR Systems
  • Thumbnail of document Data Sheet - SRX6 Precision 7-Channel Thermocouple Logger
    Download

    Data Sheet - SRX6 Precision 7-Channel Thermocouple Logger

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more